{"title":"基于视觉配准的机器人脑深部刺激电极植入精度研究","authors":"Huan-Guang Liu, A. Yang, De-Feng Liu","doi":"10.3760/CMA.J.ISSN.1001-2346.2019.10.003","DOIUrl":null,"url":null,"abstract":"Objective \nTo evaluate the accuracy and safety of electrode implantation of deep brain stimulation (DBS) with visual registration-based robot assistance. \n \n \nMethods \nThe clinical data of 15 patients with Parkinson′s disease (PD) admitted to Department of Neurosurgery, Beijing Tiantan Hospital, Capital Medical University from October 2018 to June 2019 were retrospectively analyzed. All those patients underwent DBS assisted with neurosurgical robot (Remebot system). The distance deviations of implanted electrodes in the lateral direction (x-axis) and longitudinal direction (y-axis) were documented according to postoperative CT fused with preoperative imaging of surgical plan. Postoperative complications were documented as well. \n \n \nResults \nA total of 30 DBS electrodes were successfully implanted in 15 patients. Twenty-six electrodes were implanted into the subthalamic nucleus and 4 into the globus pallidus internus. The x-axis distance deviation between the tip of electrode and planned target ranged from 0.10 mm to 1.34 mm, and the average distance deviation was 0.60±0.33 mm. The y-axis distance deviation ranged from 0.02 mm to 0.83 mm, and the average distance deviation was 0.45±0.24 mm. There were no electrode-related bleeding events, intracranial infection or surgical incision non-healing. \n \n \nConclusion \nDBS with assistance of neurosurgical robot seems to be associated with relatively little deviation in electrode implantation and no surgery-related complications. \n \n \nKey words: \nDeep brain stimulation; Parkinson disease; Robotics; Accuracy","PeriodicalId":10100,"journal":{"name":"中华神经外科杂志","volume":"35 1","pages":"981-984"},"PeriodicalIF":0.0000,"publicationDate":"2019-10-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Accuracy study of electrode implantation of robot-assisted deep brain stimulation based on visual registration\",\"authors\":\"Huan-Guang Liu, A. Yang, De-Feng Liu\",\"doi\":\"10.3760/CMA.J.ISSN.1001-2346.2019.10.003\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective \\nTo evaluate the accuracy and safety of electrode implantation of deep brain stimulation (DBS) with visual registration-based robot assistance. \\n \\n \\nMethods \\nThe clinical data of 15 patients with Parkinson′s disease (PD) admitted to Department of Neurosurgery, Beijing Tiantan Hospital, Capital Medical University from October 2018 to June 2019 were retrospectively analyzed. All those patients underwent DBS assisted with neurosurgical robot (Remebot system). The distance deviations of implanted electrodes in the lateral direction (x-axis) and longitudinal direction (y-axis) were documented according to postoperative CT fused with preoperative imaging of surgical plan. Postoperative complications were documented as well. \\n \\n \\nResults \\nA total of 30 DBS electrodes were successfully implanted in 15 patients. Twenty-six electrodes were implanted into the subthalamic nucleus and 4 into the globus pallidus internus. The x-axis distance deviation between the tip of electrode and planned target ranged from 0.10 mm to 1.34 mm, and the average distance deviation was 0.60±0.33 mm. The y-axis distance deviation ranged from 0.02 mm to 0.83 mm, and the average distance deviation was 0.45±0.24 mm. There were no electrode-related bleeding events, intracranial infection or surgical incision non-healing. \\n \\n \\nConclusion \\nDBS with assistance of neurosurgical robot seems to be associated with relatively little deviation in electrode implantation and no surgery-related complications. \\n \\n \\nKey words: \\nDeep brain stimulation; Parkinson disease; Robotics; Accuracy\",\"PeriodicalId\":10100,\"journal\":{\"name\":\"中华神经外科杂志\",\"volume\":\"35 1\",\"pages\":\"981-984\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2019-10-28\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"中华神经外科杂志\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.3760/CMA.J.ISSN.1001-2346.2019.10.003\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"中华神经外科杂志","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3760/CMA.J.ISSN.1001-2346.2019.10.003","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
摘要
目的评价基于视觉配准的机器人辅助深部脑刺激电极植入的准确性和安全性。方法回顾性分析首都医科大学附属北京天坛医院神经外科2018年10月至2019年6月收治的15例帕金森病(PD)患者的临床资料。所有患者均在神经外科机器人(Remebot系统)的辅助下接受DBS治疗。根据术后CT与术前手术方案影像融合记录植入电极在横向(x轴)和纵向(y轴)的距离偏差。术后并发症也有记录。结果15例患者共成功植入30个DBS电极。26个电极植入丘脑下核,4个电极植入内苍白球。电极尖端与计划靶的x轴距离偏差范围为0.10 mm ~ 1.34 mm,平均距离偏差为0.60±0.33 mm。y轴距离偏差范围为0.02 mm ~ 0.83 mm,平均距离偏差为0.45±0.24 mm。无电极相关出血事件、颅内感染或手术切口未愈合。结论神经外科机器人辅助DBS电极植入偏差较小,无手术相关并发症。关键词:脑深部电刺激;帕金森病;机器人技术;精度
Accuracy study of electrode implantation of robot-assisted deep brain stimulation based on visual registration
Objective
To evaluate the accuracy and safety of electrode implantation of deep brain stimulation (DBS) with visual registration-based robot assistance.
Methods
The clinical data of 15 patients with Parkinson′s disease (PD) admitted to Department of Neurosurgery, Beijing Tiantan Hospital, Capital Medical University from October 2018 to June 2019 were retrospectively analyzed. All those patients underwent DBS assisted with neurosurgical robot (Remebot system). The distance deviations of implanted electrodes in the lateral direction (x-axis) and longitudinal direction (y-axis) were documented according to postoperative CT fused with preoperative imaging of surgical plan. Postoperative complications were documented as well.
Results
A total of 30 DBS electrodes were successfully implanted in 15 patients. Twenty-six electrodes were implanted into the subthalamic nucleus and 4 into the globus pallidus internus. The x-axis distance deviation between the tip of electrode and planned target ranged from 0.10 mm to 1.34 mm, and the average distance deviation was 0.60±0.33 mm. The y-axis distance deviation ranged from 0.02 mm to 0.83 mm, and the average distance deviation was 0.45±0.24 mm. There were no electrode-related bleeding events, intracranial infection or surgical incision non-healing.
Conclusion
DBS with assistance of neurosurgical robot seems to be associated with relatively little deviation in electrode implantation and no surgery-related complications.
Key words:
Deep brain stimulation; Parkinson disease; Robotics; Accuracy
期刊介绍:
Chinese Journal of Neurosurgery is one of the series of journals organized by the Chinese Medical Association under the supervision of the China Association for Science and Technology. The journal is aimed at neurosurgeons and related researchers, and reports on the leading scientific research results and clinical experience in the field of neurosurgery, as well as the basic theoretical research closely related to neurosurgery.Chinese Journal of Neurosurgery has been included in many famous domestic search organizations, such as China Knowledge Resources Database, China Biomedical Journal Citation Database, Chinese Biomedical Journal Literature Database, China Science Citation Database, China Biomedical Literature Database, China Science and Technology Paper Citation Statistical Analysis Database, and China Science and Technology Journal Full Text Database, Wanfang Data Database of Medical Journals, etc.