不同液体对7102微浴温度均匀性和稳定性的影响

Waslina Rangkuti, Kerista Tarigan, S. Humaidi, M. Situmorang, E. Frida, Yahya Darmawan
{"title":"不同液体对7102微浴温度均匀性和稳定性的影响","authors":"Waslina Rangkuti, Kerista Tarigan, S. Humaidi, M. Situmorang, E. Frida, Yahya Darmawan","doi":"10.33394/j-ps.v11i2.7842","DOIUrl":null,"url":null,"abstract":"Microbath Fluke Type 7102 is used for thermometer calibration. In the calibration process, Microbath uses liquid media as heat conductor. Liquid media in Microbath during the calibration process there is a value of uniformity and temperature stability. The value of temperature uniformity and stability is an influential component in determining the value of measurement uncertainty (U95). The smaller the U95 value, the better the calibration results. This is a factor in this study to analyse the uniformity and temperature stability of liquid types of Water, Methanol and Glycol. The uniformity test method is carried out using 5 (five) point measurements, where the reference point is in the middle. The stability test method uses the measurement of one reference point. Uniformity and stability values are connected to determine the uncertainty of measurement value using the GUM (Guide to the expression of Uncertainty in Measurement) method. The analysis showed that Methanol is more homogeneous than Glycol and Water, with values of 0.0855 ºC < 0.0942 ºC < 0.1030 ºC. Water is more stable than Methanol and Glycol, with values of 0.0021 ºC < 0.0027 ºC < 0.0028 ºC. The time to stabilise Methanol is better than Water and Glycol. Methanol can be stabilised with ± 35 - 40 minutes, Water needs ± 38 - 40 minutes and Glycol needs ± 48 - 50 minutes. The relationship between uniformity and temperature stability is that the smaller the uniformity and stability values, the smaller the U95 of a calibration result. The U95 value of Methanol 0.11 ºC, Glycol 0.12 ºC and Water is 0.13 ºC.","PeriodicalId":33562,"journal":{"name":"Prisma Sains Jurnal Pengkajian Ilmu dan Pembelajaran Matematika dan IPA IKIP Mataram","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2023-04-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"The Effect of Different Liquid on Temperature Uniformity and Stability in Microbath 7102\",\"authors\":\"Waslina Rangkuti, Kerista Tarigan, S. Humaidi, M. Situmorang, E. Frida, Yahya Darmawan\",\"doi\":\"10.33394/j-ps.v11i2.7842\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Microbath Fluke Type 7102 is used for thermometer calibration. In the calibration process, Microbath uses liquid media as heat conductor. Liquid media in Microbath during the calibration process there is a value of uniformity and temperature stability. The value of temperature uniformity and stability is an influential component in determining the value of measurement uncertainty (U95). The smaller the U95 value, the better the calibration results. This is a factor in this study to analyse the uniformity and temperature stability of liquid types of Water, Methanol and Glycol. The uniformity test method is carried out using 5 (five) point measurements, where the reference point is in the middle. The stability test method uses the measurement of one reference point. Uniformity and stability values are connected to determine the uncertainty of measurement value using the GUM (Guide to the expression of Uncertainty in Measurement) method. The analysis showed that Methanol is more homogeneous than Glycol and Water, with values of 0.0855 ºC < 0.0942 ºC < 0.1030 ºC. Water is more stable than Methanol and Glycol, with values of 0.0021 ºC < 0.0027 ºC < 0.0028 ºC. The time to stabilise Methanol is better than Water and Glycol. Methanol can be stabilised with ± 35 - 40 minutes, Water needs ± 38 - 40 minutes and Glycol needs ± 48 - 50 minutes. The relationship between uniformity and temperature stability is that the smaller the uniformity and stability values, the smaller the U95 of a calibration result. The U95 value of Methanol 0.11 ºC, Glycol 0.12 ºC and Water is 0.13 ºC.\",\"PeriodicalId\":33562,\"journal\":{\"name\":\"Prisma Sains Jurnal Pengkajian Ilmu dan Pembelajaran Matematika dan IPA IKIP Mataram\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-04-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Prisma Sains Jurnal Pengkajian Ilmu dan Pembelajaran Matematika dan IPA IKIP Mataram\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.33394/j-ps.v11i2.7842\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Prisma Sains Jurnal Pengkajian Ilmu dan Pembelajaran Matematika dan IPA IKIP Mataram","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.33394/j-ps.v11i2.7842","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

7102型Micropath Fluke用于温度计校准。在校准过程中,Micropath使用液体介质作为导热体。微通道中的液体介质在校准过程中存在一定的均匀性和温度稳定性。温度均匀性和稳定性的值是确定测量不确定度(U95)值的一个有影响的组成部分。U95值越小,校准结果越好。这是本研究中分析水、甲醇和乙二醇液体类型的均匀性和温度稳定性的一个因素。均匀性测试方法使用5(五)个点测量进行,其中参考点位于在中间。稳定性测试方法使用一个参考点的测量。使用GUM(测量不确定度表达指南)方法连接均匀性和稳定性值以确定测量值的不确定度。分析表明,甲醇比乙二醇和水更均匀,其值为0.0855ºC<0.0942ºC<0.1030ºC。水比甲醇和乙二醇更稳定,其值为0.0021ºC<0.0027ºC<0.0028ºC。稳定甲醇的时间比水和乙二醇好。甲醇需要±35-40分钟才能稳定,水需要±38-40分钟,乙二醇需要±48-50分钟。均匀性和温度稳定性之间的关系是,均匀性和稳定性值越小,校准结果的U95就越小。甲醇0.11ºC、乙二醇0.12ºC和水的U95值为0.13ºC。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
The Effect of Different Liquid on Temperature Uniformity and Stability in Microbath 7102
Microbath Fluke Type 7102 is used for thermometer calibration. In the calibration process, Microbath uses liquid media as heat conductor. Liquid media in Microbath during the calibration process there is a value of uniformity and temperature stability. The value of temperature uniformity and stability is an influential component in determining the value of measurement uncertainty (U95). The smaller the U95 value, the better the calibration results. This is a factor in this study to analyse the uniformity and temperature stability of liquid types of Water, Methanol and Glycol. The uniformity test method is carried out using 5 (five) point measurements, where the reference point is in the middle. The stability test method uses the measurement of one reference point. Uniformity and stability values are connected to determine the uncertainty of measurement value using the GUM (Guide to the expression of Uncertainty in Measurement) method. The analysis showed that Methanol is more homogeneous than Glycol and Water, with values of 0.0855 ºC < 0.0942 ºC < 0.1030 ºC. Water is more stable than Methanol and Glycol, with values of 0.0021 ºC < 0.0027 ºC < 0.0028 ºC. The time to stabilise Methanol is better than Water and Glycol. Methanol can be stabilised with ± 35 - 40 minutes, Water needs ± 38 - 40 minutes and Glycol needs ± 48 - 50 minutes. The relationship between uniformity and temperature stability is that the smaller the uniformity and stability values, the smaller the U95 of a calibration result. The U95 value of Methanol 0.11 ºC, Glycol 0.12 ºC and Water is 0.13 ºC.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
79
审稿时长
24 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信