Intox EC/IR II对异丙醇和异丙醇/乙醇混合物的响应

IF 0.2 Q4 MEDICINE, LEGAL
Karen C. W. Chan
{"title":"Intox EC/IR II对异丙醇和异丙醇/乙醇混合物的响应","authors":"Karen C. W. Chan","doi":"10.1080/00085030.2018.1491691","DOIUrl":null,"url":null,"abstract":"Abstract The Intox EC/IR II was tested with isopropyl alcohol (IPA) solutions, and IPA and ethanol (EtOH) mixtures, at various concentrations to determine whether IPA contributes to, or produces, an apparent ethanol concentration (AEC). Two test modes were used – Subject Test, which has the interfering substance (IS) detection mechanism enabled and reports truncated results; and Quick Test, which reports untruncated results and does not detect IS. When testing IPA solutions alone, concentrations of 40 mg/dL (mg%) and above produced the IS status message while concentrations below 40 mg% produced numerical results in the Subject Test mode. With IPA alone, the IPA contribution to AEC was approximately half of the target IPA concentration. When IPA + EtOH mixtures were tested, all mixtures except one generated the IS status message. The one IPA + EtOH mixture, with 10 mg% IPA and 150 mg% target apparent EtOH, consistently produced numerical results in the Subject Test mode; however, the 10 mg% IPA in the mixture did not contribute to the AEC. This study showed that with the IS detection mechanism activated on the Intox EC/IR II, Subject Tests were aborted in instances where IPA would have contributed to the AEC and the AEC was 20 mg% or above.","PeriodicalId":44383,"journal":{"name":"Canadian Society of Forensic Science Journal","volume":"51 1","pages":"67 - 82"},"PeriodicalIF":0.2000,"publicationDate":"2018-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1080/00085030.2018.1491691","citationCount":"0","resultStr":"{\"title\":\"The response of the Intox EC/IR II to isopropanol and isopropanol/ethanol mixtures\",\"authors\":\"Karen C. W. Chan\",\"doi\":\"10.1080/00085030.2018.1491691\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Abstract The Intox EC/IR II was tested with isopropyl alcohol (IPA) solutions, and IPA and ethanol (EtOH) mixtures, at various concentrations to determine whether IPA contributes to, or produces, an apparent ethanol concentration (AEC). Two test modes were used – Subject Test, which has the interfering substance (IS) detection mechanism enabled and reports truncated results; and Quick Test, which reports untruncated results and does not detect IS. When testing IPA solutions alone, concentrations of 40 mg/dL (mg%) and above produced the IS status message while concentrations below 40 mg% produced numerical results in the Subject Test mode. With IPA alone, the IPA contribution to AEC was approximately half of the target IPA concentration. When IPA + EtOH mixtures were tested, all mixtures except one generated the IS status message. The one IPA + EtOH mixture, with 10 mg% IPA and 150 mg% target apparent EtOH, consistently produced numerical results in the Subject Test mode; however, the 10 mg% IPA in the mixture did not contribute to the AEC. This study showed that with the IS detection mechanism activated on the Intox EC/IR II, Subject Tests were aborted in instances where IPA would have contributed to the AEC and the AEC was 20 mg% or above.\",\"PeriodicalId\":44383,\"journal\":{\"name\":\"Canadian Society of Forensic Science Journal\",\"volume\":\"51 1\",\"pages\":\"67 - 82\"},\"PeriodicalIF\":0.2000,\"publicationDate\":\"2018-10-02\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1080/00085030.2018.1491691\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Canadian Society of Forensic Science Journal\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1080/00085030.2018.1491691\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"MEDICINE, LEGAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Canadian Society of Forensic Science Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1080/00085030.2018.1491691","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MEDICINE, LEGAL","Score":null,"Total":0}
引用次数: 0

摘要

用不同浓度的异丙醇(IPA)溶液和IPA与乙醇(EtOH)混合物对Intox EC/IR II进行了测试,以确定IPA是否有助于或产生表观乙醇浓度(AEC)。采用两种测试模式——受试者测试,启用干扰物质(IS)检测机制并报告截断结果;快速测试,它报告未截断的结果,不检测IS。当单独测试IPA溶液时,浓度为40 mg/dL (mg%)及以上会产生IS状态信息,而浓度低于40 mg%会在Subject Test模式下产生数值结果。单独使用IPA时,IPA对AEC的贡献约为目标IPA浓度的一半。当测试IPA + EtOH混合物时,除一种混合物外,所有混合物都产生IS状态信息。一种IPA + EtOH混合物,IPA为10 mg%, EtOH为150 mg%,在受试者测试模式下始终产生数值结果;然而,混合物中10 mg%的IPA对AEC没有贡献。本研究表明,在Intox EC/IR II上激活IS检测机制后,在IPA可能导致AEC且AEC为20mg %或以上的情况下,受试者测试被中止。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
The response of the Intox EC/IR II to isopropanol and isopropanol/ethanol mixtures
Abstract The Intox EC/IR II was tested with isopropyl alcohol (IPA) solutions, and IPA and ethanol (EtOH) mixtures, at various concentrations to determine whether IPA contributes to, or produces, an apparent ethanol concentration (AEC). Two test modes were used – Subject Test, which has the interfering substance (IS) detection mechanism enabled and reports truncated results; and Quick Test, which reports untruncated results and does not detect IS. When testing IPA solutions alone, concentrations of 40 mg/dL (mg%) and above produced the IS status message while concentrations below 40 mg% produced numerical results in the Subject Test mode. With IPA alone, the IPA contribution to AEC was approximately half of the target IPA concentration. When IPA + EtOH mixtures were tested, all mixtures except one generated the IS status message. The one IPA + EtOH mixture, with 10 mg% IPA and 150 mg% target apparent EtOH, consistently produced numerical results in the Subject Test mode; however, the 10 mg% IPA in the mixture did not contribute to the AEC. This study showed that with the IS detection mechanism activated on the Intox EC/IR II, Subject Tests were aborted in instances where IPA would have contributed to the AEC and the AEC was 20 mg% or above.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
1.00
自引率
0.00%
发文量
21
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信