消失的威胁:封闭吸入系统的隐形故障

Akshaya Das, Faisal Qureshi, Surinder Kumar, Nikhil Kothari
{"title":"消失的威胁:封闭吸入系统的隐形故障","authors":"Akshaya Das, Faisal Qureshi, Surinder Kumar, Nikhil Kothari","doi":"10.53097/jmv.10085","DOIUrl":null,"url":null,"abstract":"This case report presents a rare case of a mechanical ventilation leak that was initially missed by clinicians in a patient with a myxoedema coma. Despite all efforts to investigate the causes of the leak, including a bedside lung ultrasound and chest radiograph, the leak persisted. It was eventually discovered that the rhythmic inflation and deflation of the polythene covering the closed suction system was causing the leak. The closed suction system was replaced with a new one, and the tidal volume was restored, resulting in the resolution of the leak alarm. The volume leak alarm and low volume alarm are important indicators of potential problems during mechanical ventilation, and close suction system malfunction is a potential cause of volume leak that should be considered in mechanically ventilated patients. Regular monitoring and appropriate management of these alarms and potential causes can help prevent complications and optimize patient care. Keywords: Mechanical ventilation, volume leak alarm, closed suction system, and myxoedema coma.","PeriodicalId":73813,"journal":{"name":"Journal of mechanical ventilation","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2023-09-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"The vanishing threat: Stealthy malfunction of closed suction system\",\"authors\":\"Akshaya Das, Faisal Qureshi, Surinder Kumar, Nikhil Kothari\",\"doi\":\"10.53097/jmv.10085\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"This case report presents a rare case of a mechanical ventilation leak that was initially missed by clinicians in a patient with a myxoedema coma. Despite all efforts to investigate the causes of the leak, including a bedside lung ultrasound and chest radiograph, the leak persisted. It was eventually discovered that the rhythmic inflation and deflation of the polythene covering the closed suction system was causing the leak. The closed suction system was replaced with a new one, and the tidal volume was restored, resulting in the resolution of the leak alarm. The volume leak alarm and low volume alarm are important indicators of potential problems during mechanical ventilation, and close suction system malfunction is a potential cause of volume leak that should be considered in mechanically ventilated patients. Regular monitoring and appropriate management of these alarms and potential causes can help prevent complications and optimize patient care. Keywords: Mechanical ventilation, volume leak alarm, closed suction system, and myxoedema coma.\",\"PeriodicalId\":73813,\"journal\":{\"name\":\"Journal of mechanical ventilation\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-09-15\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of mechanical ventilation\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.53097/jmv.10085\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of mechanical ventilation","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.53097/jmv.10085","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

本病例报告介绍了一例罕见的机械通气泄漏病例,临床医生最初在一名黏液水肿昏迷患者中遗漏了该病例。尽管尽了一切努力调查泄漏的原因,包括床边肺部超声和胸部x线片,但泄漏仍然存在。最终发现,覆盖在封闭抽吸系统上的聚乙烯有节奏的膨胀和放气导致了泄漏。将封闭式吸入系统更换为新系统,并恢复了潮气量,从而解决了泄漏警报。容量泄漏警报和低容量警报是机械通气过程中潜在问题的重要指标,闭式抽吸系统故障是机械通气患者应考虑的容量泄漏的潜在原因。对这些警报和潜在原因进行定期监测和适当管理有助于预防并发症并优化患者护理。关键词:机械通气,容量泄漏报警,闭式抽吸系统,黏液水肿昏迷。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
The vanishing threat: Stealthy malfunction of closed suction system
This case report presents a rare case of a mechanical ventilation leak that was initially missed by clinicians in a patient with a myxoedema coma. Despite all efforts to investigate the causes of the leak, including a bedside lung ultrasound and chest radiograph, the leak persisted. It was eventually discovered that the rhythmic inflation and deflation of the polythene covering the closed suction system was causing the leak. The closed suction system was replaced with a new one, and the tidal volume was restored, resulting in the resolution of the leak alarm. The volume leak alarm and low volume alarm are important indicators of potential problems during mechanical ventilation, and close suction system malfunction is a potential cause of volume leak that should be considered in mechanically ventilated patients. Regular monitoring and appropriate management of these alarms and potential causes can help prevent complications and optimize patient care. Keywords: Mechanical ventilation, volume leak alarm, closed suction system, and myxoedema coma.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信