{"title":"不同实验室原子吸收分光光度计测定铅金属离子结果的分析性能比较","authors":"Arifin Syam Surya, Bohari Yusuf, A. S. Panggabean","doi":"10.30872/jkm.v19i1.909","DOIUrl":null,"url":null,"abstract":"The study about the comparison of analytical performance of the determination of Pb metal ions using Atomic Absorption Spectrophotometer in various laboratories in samarinda has been conducted. The validated methods are linearity, detection limits, quantity and precision limits. The results showed, a standardized regression test obtained by laboratory A, a value of R2 ≥ 0.4103, laboratory B a value of R2 ≥ 0.983 and laboratory C, a value of R2 ≥ 0.000. The detection limit found at laboratory A was 0.6728 µg/L, laboratory B was 0.1989 µg/L and laboratory C was 0.0001 µg/L. The quantization limit obtained by laboratory A was 0.6817 µg/L, laboratory B was 0.6063 µg/L and laboratory C was 0.0001 µg/L. The precision test of laboratory A was 7.2208, laboratory B was 5.4557 and laboratory C was 0,000. The result of determining Pb of gas station soil samples in laboratory A is 0.737 µg/L and laboratory B is 0.391 µg/L, so that means they have exceeded the quality standards set by the Indonesian National Standard (INS).","PeriodicalId":31725,"journal":{"name":"Jurnal Kimia Mulawarman","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2021-11-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"COMPARISON OF ANALYTIC PERFORMANCE OF RESULTS OF DETERMINATION OF Pb METAL IONS USING ATOMIC ABSORPTION SPECTROPHOTOMETER of VARIOUS in SAMARINDA LABORATORIES\",\"authors\":\"Arifin Syam Surya, Bohari Yusuf, A. S. Panggabean\",\"doi\":\"10.30872/jkm.v19i1.909\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The study about the comparison of analytical performance of the determination of Pb metal ions using Atomic Absorption Spectrophotometer in various laboratories in samarinda has been conducted. The validated methods are linearity, detection limits, quantity and precision limits. The results showed, a standardized regression test obtained by laboratory A, a value of R2 ≥ 0.4103, laboratory B a value of R2 ≥ 0.983 and laboratory C, a value of R2 ≥ 0.000. The detection limit found at laboratory A was 0.6728 µg/L, laboratory B was 0.1989 µg/L and laboratory C was 0.0001 µg/L. The quantization limit obtained by laboratory A was 0.6817 µg/L, laboratory B was 0.6063 µg/L and laboratory C was 0.0001 µg/L. The precision test of laboratory A was 7.2208, laboratory B was 5.4557 and laboratory C was 0,000. The result of determining Pb of gas station soil samples in laboratory A is 0.737 µg/L and laboratory B is 0.391 µg/L, so that means they have exceeded the quality standards set by the Indonesian National Standard (INS).\",\"PeriodicalId\":31725,\"journal\":{\"name\":\"Jurnal Kimia Mulawarman\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2021-11-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Jurnal Kimia Mulawarman\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.30872/jkm.v19i1.909\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Jurnal Kimia Mulawarman","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.30872/jkm.v19i1.909","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
COMPARISON OF ANALYTIC PERFORMANCE OF RESULTS OF DETERMINATION OF Pb METAL IONS USING ATOMIC ABSORPTION SPECTROPHOTOMETER of VARIOUS in SAMARINDA LABORATORIES
The study about the comparison of analytical performance of the determination of Pb metal ions using Atomic Absorption Spectrophotometer in various laboratories in samarinda has been conducted. The validated methods are linearity, detection limits, quantity and precision limits. The results showed, a standardized regression test obtained by laboratory A, a value of R2 ≥ 0.4103, laboratory B a value of R2 ≥ 0.983 and laboratory C, a value of R2 ≥ 0.000. The detection limit found at laboratory A was 0.6728 µg/L, laboratory B was 0.1989 µg/L and laboratory C was 0.0001 µg/L. The quantization limit obtained by laboratory A was 0.6817 µg/L, laboratory B was 0.6063 µg/L and laboratory C was 0.0001 µg/L. The precision test of laboratory A was 7.2208, laboratory B was 5.4557 and laboratory C was 0,000. The result of determining Pb of gas station soil samples in laboratory A is 0.737 µg/L and laboratory B is 0.391 µg/L, so that means they have exceeded the quality standards set by the Indonesian National Standard (INS).