Leonardus Lewo Hayon, A. Alimuddin, A. S. Panggabean
{"title":"加里曼丹PUPUK质量控制实验室火焰原子吸收分光光度法测定NPK肥料中Cd (II)的验证方法","authors":"Leonardus Lewo Hayon, A. Alimuddin, A. S. Panggabean","doi":"10.30872/jkm.v19i1.898","DOIUrl":null,"url":null,"abstract":"The research about validation method of determination of Cd (II) in NPK fertilizer by Flame Atomic Absorption Spectrophotometry at the Quality Control Laboratory, PT. Pupuk Kaltim has been done. Determination of Cd (II) was done by referring to the Indonesian National Standard (SNI) 2803:2012. Based on the research results obtained by the correlation coefficient (r) was 0.9991; reproducibility with a relative standard deviation (RSD) value was 4.502 % less than 2/3 CV Horwitz was 9.757 and repeatability with a relative standard deviation (RSD) value was 5.225% less than 2/3 CV Horwitz was 9.695; % recovery was in the range of 80-110 % with an average was 98.56 %; limit of detection was 0.056 mg/kg and limit of quantitation was 0,066 mg/kg; uncertainty estimation of measurement with uncertainty components of concentration, volume, mass, precision and purity of the standard solution, obtained the expanded uncertainty value with a 95 % confidence level was ± 0.197 mg/kg and uncertainty percentage value was 10.43 %.","PeriodicalId":31725,"journal":{"name":"Jurnal Kimia Mulawarman","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2021-11-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"VALIDATION METHOD OF DETERMINATION OF Cd (II) IN NPK FERTILIZER BY FLAME ATOMIC ABSORPTION SPECTROPHOTOMETRY AT THE QUALITY CONTROL LABORATORY, PT. PUPUK KALIMANTAN TIMUR\",\"authors\":\"Leonardus Lewo Hayon, A. Alimuddin, A. S. Panggabean\",\"doi\":\"10.30872/jkm.v19i1.898\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The research about validation method of determination of Cd (II) in NPK fertilizer by Flame Atomic Absorption Spectrophotometry at the Quality Control Laboratory, PT. Pupuk Kaltim has been done. Determination of Cd (II) was done by referring to the Indonesian National Standard (SNI) 2803:2012. Based on the research results obtained by the correlation coefficient (r) was 0.9991; reproducibility with a relative standard deviation (RSD) value was 4.502 % less than 2/3 CV Horwitz was 9.757 and repeatability with a relative standard deviation (RSD) value was 5.225% less than 2/3 CV Horwitz was 9.695; % recovery was in the range of 80-110 % with an average was 98.56 %; limit of detection was 0.056 mg/kg and limit of quantitation was 0,066 mg/kg; uncertainty estimation of measurement with uncertainty components of concentration, volume, mass, precision and purity of the standard solution, obtained the expanded uncertainty value with a 95 % confidence level was ± 0.197 mg/kg and uncertainty percentage value was 10.43 %.\",\"PeriodicalId\":31725,\"journal\":{\"name\":\"Jurnal Kimia Mulawarman\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2021-11-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Jurnal Kimia Mulawarman\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.30872/jkm.v19i1.898\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Jurnal Kimia Mulawarman","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.30872/jkm.v19i1.898","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
VALIDATION METHOD OF DETERMINATION OF Cd (II) IN NPK FERTILIZER BY FLAME ATOMIC ABSORPTION SPECTROPHOTOMETRY AT THE QUALITY CONTROL LABORATORY, PT. PUPUK KALIMANTAN TIMUR
The research about validation method of determination of Cd (II) in NPK fertilizer by Flame Atomic Absorption Spectrophotometry at the Quality Control Laboratory, PT. Pupuk Kaltim has been done. Determination of Cd (II) was done by referring to the Indonesian National Standard (SNI) 2803:2012. Based on the research results obtained by the correlation coefficient (r) was 0.9991; reproducibility with a relative standard deviation (RSD) value was 4.502 % less than 2/3 CV Horwitz was 9.757 and repeatability with a relative standard deviation (RSD) value was 5.225% less than 2/3 CV Horwitz was 9.695; % recovery was in the range of 80-110 % with an average was 98.56 %; limit of detection was 0.056 mg/kg and limit of quantitation was 0,066 mg/kg; uncertainty estimation of measurement with uncertainty components of concentration, volume, mass, precision and purity of the standard solution, obtained the expanded uncertainty value with a 95 % confidence level was ± 0.197 mg/kg and uncertainty percentage value was 10.43 %.