两种紫杉醇涂层支架在实验环境中的比较评价。

IF 1.7 2区 医学 Q3 PERIPHERAL VASCULAR DISEASE
Journal of Endovascular Therapy Pub Date : 2025-06-01 Epub Date: 2023-09-20 DOI:10.1177/15266028231198033
Alexandre Pouhin, Raphaël Coscas, Valentin Crespy, Olivia Poupardin, Jean-Paul Pais-De-Barros, Olivier Bouchot, Alain Bernard, Eric Steinmetz
{"title":"两种紫杉醇涂层支架在实验环境中的比较评价。","authors":"Alexandre Pouhin, Raphaël Coscas, Valentin Crespy, Olivia Poupardin, Jean-Paul Pais-De-Barros, Olivier Bouchot, Alain Bernard, Eric Steinmetz","doi":"10.1177/15266028231198033","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Unlike paclitaxel-coated balloons, pre-clinical data comparing different paclitaxel-coated stents (PCSs) are weak. The study objective was to compare the features of the 2 main PCSs: Eluvia® (Boston Scientific, Marlborough, MA) versus ZilverPTX® (Cook Medical, Bloomington, IN).</p><p><strong>Method: </strong>Analysis was carried out on 12 pigs divided into 2 groups: Eluvia® (n=6) and ZilverPTX® (n=6). The pigs received the PCS corresponding to their group in each external iliac artery and were paired one by one, to examine 6 different post-implantation timepoints: after 30 minutes, 6 hours, 24 hours, 3 days, 7 days, and 14 days. The paclitaxel concentration measurements and the histological analysis were carried out under blind testing on the plasma, arterial, lymph node, and muscle samples. A linear regression model and Wilcoxon Mann-Whitney test were used to study the variables.</p><p><strong>Results: </strong>The plasma paclitaxel rate decrease over 24 hours after PCS implantation was significantly different between the two groups, expressed by the correlation coefficient 0.19 (0.14-0.23; p<0.001) with an undetectable concentration at the 10th hour for Eluvia® versus 3 days for ZilverPTX®. Significantly higher paclitaxel concentrations with ZilverPTX® PCS were observed in muscle samples at each timepoint: extensor digitorum brevis 3.2 (1.17-5.23; p=0.005), biceps femoris 4.27 (2.27-6.26; p<0.001), semi-tendinosus 3.79 (1.85-5.73; p=0.001), tibialis anterior 3.0 (1.37-4.64; p=0.001), and in the femoral nodes 2.27±1.74 ng/g versus 0.14±0.13 ng/g (p<0.001). Histological analysis revealed a trend for more marked intimal inflammation in the arteries stented with ZilverPTX® (p=0.063), especially after the 7th and 14th days.</p><p><strong>Conclusion: </strong>Such a difference in the concentration of paclitaxel in the plasma, muscles, and lymph nodes between the two stents was higher than expected based on differences in device design. The clinical consequences of these results remain to be elucidated, particularly regarding the concerning presence of paclitaxel in muscles and adjacent lymph nodes.Clinical ImpactThis experimental study compares 2 paclitaxel-coated stents. It demonstrates that differences in stent designs and drug features (coatings and concentrations) translate into differences in terms of concentrations of paclitaxel in the plasma, muscles, and lymph nodes. Our results favor the Eluvia® stent over the ZilverPTX® stent, although more studies are required to confirm this conclusion.</p>","PeriodicalId":50210,"journal":{"name":"Journal of Endovascular Therapy","volume":" ","pages":"869-877"},"PeriodicalIF":1.7000,"publicationDate":"2025-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Comparative Evaluation of Two Paclitaxel-Coated Stents in an Experimental Setting.\",\"authors\":\"Alexandre Pouhin, Raphaël Coscas, Valentin Crespy, Olivia Poupardin, Jean-Paul Pais-De-Barros, Olivier Bouchot, Alain Bernard, Eric Steinmetz\",\"doi\":\"10.1177/15266028231198033\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>Unlike paclitaxel-coated balloons, pre-clinical data comparing different paclitaxel-coated stents (PCSs) are weak. The study objective was to compare the features of the 2 main PCSs: Eluvia® (Boston Scientific, Marlborough, MA) versus ZilverPTX® (Cook Medical, Bloomington, IN).</p><p><strong>Method: </strong>Analysis was carried out on 12 pigs divided into 2 groups: Eluvia® (n=6) and ZilverPTX® (n=6). The pigs received the PCS corresponding to their group in each external iliac artery and were paired one by one, to examine 6 different post-implantation timepoints: after 30 minutes, 6 hours, 24 hours, 3 days, 7 days, and 14 days. The paclitaxel concentration measurements and the histological analysis were carried out under blind testing on the plasma, arterial, lymph node, and muscle samples. A linear regression model and Wilcoxon Mann-Whitney test were used to study the variables.</p><p><strong>Results: </strong>The plasma paclitaxel rate decrease over 24 hours after PCS implantation was significantly different between the two groups, expressed by the correlation coefficient 0.19 (0.14-0.23; p<0.001) with an undetectable concentration at the 10th hour for Eluvia® versus 3 days for ZilverPTX®. Significantly higher paclitaxel concentrations with ZilverPTX® PCS were observed in muscle samples at each timepoint: extensor digitorum brevis 3.2 (1.17-5.23; p=0.005), biceps femoris 4.27 (2.27-6.26; p<0.001), semi-tendinosus 3.79 (1.85-5.73; p=0.001), tibialis anterior 3.0 (1.37-4.64; p=0.001), and in the femoral nodes 2.27±1.74 ng/g versus 0.14±0.13 ng/g (p<0.001). Histological analysis revealed a trend for more marked intimal inflammation in the arteries stented with ZilverPTX® (p=0.063), especially after the 7th and 14th days.</p><p><strong>Conclusion: </strong>Such a difference in the concentration of paclitaxel in the plasma, muscles, and lymph nodes between the two stents was higher than expected based on differences in device design. The clinical consequences of these results remain to be elucidated, particularly regarding the concerning presence of paclitaxel in muscles and adjacent lymph nodes.Clinical ImpactThis experimental study compares 2 paclitaxel-coated stents. It demonstrates that differences in stent designs and drug features (coatings and concentrations) translate into differences in terms of concentrations of paclitaxel in the plasma, muscles, and lymph nodes. Our results favor the Eluvia® stent over the ZilverPTX® stent, although more studies are required to confirm this conclusion.</p>\",\"PeriodicalId\":50210,\"journal\":{\"name\":\"Journal of Endovascular Therapy\",\"volume\":\" \",\"pages\":\"869-877\"},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2025-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Endovascular Therapy\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1177/15266028231198033\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/9/20 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q3\",\"JCRName\":\"PERIPHERAL VASCULAR DISEASE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Endovascular Therapy","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/15266028231198033","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/9/20 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

摘要

引言:与紫杉醇涂层球囊不同,比较不同紫杉醇涂层支架(PCS)的临床前数据较弱。本研究的目的是比较两种主要PCS的特征:Eluvia®(马萨诸塞州马尔伯勒波士顿科学公司)与ZilverPTX®(印第安纳州布鲁明顿库克医疗公司)。方法:对12头猪进行分析,分为两组:Eluvia®(n=6)和ZilverPTX®(n=六)。猪在每个髂外动脉接受与其组对应的PCS,并逐一配对,以检查6个不同的植入后时间点:30分钟、6小时、24小时、3天、7天和14天。紫杉醇浓度测量和组织学分析是在对血浆、动脉、淋巴结和肌肉样本进行盲测的情况下进行的。使用线性回归模型和Wilcoxon-Mann-Whitney检验来研究变量。结果:两组患者植入PCS后24小时血浆紫杉醇下降率差异有统计学意义,用相关系数0.19表示(0.14-0.23;p结论:由于器械设计的差异,两个支架之间血浆、肌肉和淋巴结中紫杉醇浓度的差异高于预期。这些结果的临床后果仍有待阐明,特别是关于肌肉和邻近淋巴结中存在紫杉醇的问题ental研究比较了2个紫杉醇涂层支架。研究表明,支架设计和药物特征(涂层和浓度)的差异转化为血浆、肌肉和淋巴结中紫杉醇浓度的差异。我们的结果支持Eluvia®支架而不是ZilverPTX®支架,尽管还需要更多的研究来证实这一结论。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Comparative Evaluation of Two Paclitaxel-Coated Stents in an Experimental Setting.

Introduction: Unlike paclitaxel-coated balloons, pre-clinical data comparing different paclitaxel-coated stents (PCSs) are weak. The study objective was to compare the features of the 2 main PCSs: Eluvia® (Boston Scientific, Marlborough, MA) versus ZilverPTX® (Cook Medical, Bloomington, IN).

Method: Analysis was carried out on 12 pigs divided into 2 groups: Eluvia® (n=6) and ZilverPTX® (n=6). The pigs received the PCS corresponding to their group in each external iliac artery and were paired one by one, to examine 6 different post-implantation timepoints: after 30 minutes, 6 hours, 24 hours, 3 days, 7 days, and 14 days. The paclitaxel concentration measurements and the histological analysis were carried out under blind testing on the plasma, arterial, lymph node, and muscle samples. A linear regression model and Wilcoxon Mann-Whitney test were used to study the variables.

Results: The plasma paclitaxel rate decrease over 24 hours after PCS implantation was significantly different between the two groups, expressed by the correlation coefficient 0.19 (0.14-0.23; p<0.001) with an undetectable concentration at the 10th hour for Eluvia® versus 3 days for ZilverPTX®. Significantly higher paclitaxel concentrations with ZilverPTX® PCS were observed in muscle samples at each timepoint: extensor digitorum brevis 3.2 (1.17-5.23; p=0.005), biceps femoris 4.27 (2.27-6.26; p<0.001), semi-tendinosus 3.79 (1.85-5.73; p=0.001), tibialis anterior 3.0 (1.37-4.64; p=0.001), and in the femoral nodes 2.27±1.74 ng/g versus 0.14±0.13 ng/g (p<0.001). Histological analysis revealed a trend for more marked intimal inflammation in the arteries stented with ZilverPTX® (p=0.063), especially after the 7th and 14th days.

Conclusion: Such a difference in the concentration of paclitaxel in the plasma, muscles, and lymph nodes between the two stents was higher than expected based on differences in device design. The clinical consequences of these results remain to be elucidated, particularly regarding the concerning presence of paclitaxel in muscles and adjacent lymph nodes.Clinical ImpactThis experimental study compares 2 paclitaxel-coated stents. It demonstrates that differences in stent designs and drug features (coatings and concentrations) translate into differences in terms of concentrations of paclitaxel in the plasma, muscles, and lymph nodes. Our results favor the Eluvia® stent over the ZilverPTX® stent, although more studies are required to confirm this conclusion.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
5.30
自引率
15.40%
发文量
203
审稿时长
6-12 weeks
期刊介绍: The Journal of Endovascular Therapy (formerly the Journal of Endovascular Surgery) was established in 1994 as a forum for all physicians, scientists, and allied healthcare professionals who are engaged or interested in peripheral endovascular techniques and technology. An official publication of the International Society of Endovascular Specialists (ISEVS), the Journal of Endovascular Therapy publishes peer-reviewed articles of interest to clinicians and researchers in the field of peripheral endovascular interventions.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信