[低浓度二硫苏糖醇对输血相容性检测中单克隆抗cd38干扰阴性作用的评价]。

Jie Song, Ying-Hui Chen Li, Ai-Ping Liu, Shu-Ya Wang, Xian-Ping Lyu
{"title":"[低浓度二硫苏糖醇对输血相容性检测中单克隆抗cd38干扰阴性作用的评价]。","authors":"Jie Song,&nbsp;Ying-Hui Chen Li,&nbsp;Ai-Ping Liu,&nbsp;Shu-Ya Wang,&nbsp;Xian-Ping Lyu","doi":"10.19746/j.cnki.issn.1009-2137.2022.04.035","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To investigate the effectiveness and safety of low concentration dithiothreitol (DTT) in removing the interference of monoclonal anti-CD38 on transfusion compatibility testing, and develop a reasonable clinical transfusion strategy.</p><p><strong>Methods: </strong>The blood type, direct antiglobulin testing (DAT) and antibody screening were tested according to standard methods. Antibody screening cells and donor's red blood cells were treated by DTT 0.2, 0.1, 0.05, 0.02, 0.01 and 0.005 mol/L, and antibody screening and cross-matching of serums after monoclonal anti-CD38 treatment were performed by anti-human globulin card.</p><p><strong>Results: </strong>The 0.01 mol/L DTT at 37℃ for 30 minutes could remove the effect of monoclonal anti-CD38 on antibody screening and cross-matching, meanwhile retain their effectiveness in detecting anti-K, anti-LW, anti-JMH, anti-Lu<sup>b</sup>, anti-e, anti-Di<sup>a</sup> and anti-Jk<sup>a</sup> alloantibodies. All the 10 patients had no acute or delayed haemolytic transfusion reactions and their routine blood tests showed that the red blood cells transfusion was effective.</p><p><strong>Conclusion: </strong>The 0.01 mol/L DTT is a safe and effective method for removing the interference of monoclonal anti-CD38 with transfusion compatibility testing, while retaining the ability to detect most alloantibodies.</p>","PeriodicalId":519535,"journal":{"name":"Zhongguo shi yan xue ye xue za zhi","volume":" ","pages":"1198-1202"},"PeriodicalIF":0.0000,"publicationDate":"2022-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Evaluation of Low Concentration Dithiothreitol for Negating the Monoclonal Anti-CD38 Interference with Transfusion Compatibility Testing].\",\"authors\":\"Jie Song,&nbsp;Ying-Hui Chen Li,&nbsp;Ai-Ping Liu,&nbsp;Shu-Ya Wang,&nbsp;Xian-Ping Lyu\",\"doi\":\"10.19746/j.cnki.issn.1009-2137.2022.04.035\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>To investigate the effectiveness and safety of low concentration dithiothreitol (DTT) in removing the interference of monoclonal anti-CD38 on transfusion compatibility testing, and develop a reasonable clinical transfusion strategy.</p><p><strong>Methods: </strong>The blood type, direct antiglobulin testing (DAT) and antibody screening were tested according to standard methods. Antibody screening cells and donor's red blood cells were treated by DTT 0.2, 0.1, 0.05, 0.02, 0.01 and 0.005 mol/L, and antibody screening and cross-matching of serums after monoclonal anti-CD38 treatment were performed by anti-human globulin card.</p><p><strong>Results: </strong>The 0.01 mol/L DTT at 37℃ for 30 minutes could remove the effect of monoclonal anti-CD38 on antibody screening and cross-matching, meanwhile retain their effectiveness in detecting anti-K, anti-LW, anti-JMH, anti-Lu<sup>b</sup>, anti-e, anti-Di<sup>a</sup> and anti-Jk<sup>a</sup> alloantibodies. All the 10 patients had no acute or delayed haemolytic transfusion reactions and their routine blood tests showed that the red blood cells transfusion was effective.</p><p><strong>Conclusion: </strong>The 0.01 mol/L DTT is a safe and effective method for removing the interference of monoclonal anti-CD38 with transfusion compatibility testing, while retaining the ability to detect most alloantibodies.</p>\",\"PeriodicalId\":519535,\"journal\":{\"name\":\"Zhongguo shi yan xue ye xue za zhi\",\"volume\":\" \",\"pages\":\"1198-1202\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-08-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Zhongguo shi yan xue ye xue za zhi\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.19746/j.cnki.issn.1009-2137.2022.04.035\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Zhongguo shi yan xue ye xue za zhi","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.19746/j.cnki.issn.1009-2137.2022.04.035","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:探讨低浓度二硫苏糖醇(DTT)消除单克隆抗cd38对输血相容性检测干扰的有效性和安全性,制定合理的临床输血策略。方法:按标准方法进行血型、直接抗球蛋白试验(DAT)和抗体筛查。抗体筛选细胞和供者红细胞分别用DTT 0.2、0.1、0.05、0.02、0.01和0.005 mol/L处理,单克隆抗cd38处理后的血清用抗人球蛋白卡进行抗体筛选和交叉配型。结果:37℃、30 min、0.01 mol/L DTT可消除单克隆抗cd38对抗体筛选和交叉配型的影响,同时保留其检测抗k、抗lw、抗jmh、抗lub、抗e、抗dia和抗jka同种异体抗体的有效性。10例患者均无急性或迟发性溶血输血反应,血常规检查显示红细胞输注有效。结论:0.01 mol/L DTT是一种安全有效的方法,可消除单克隆抗cd38对输血相容性检测的干扰,同时保留对大多数同种异体抗体的检测能力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
[Evaluation of Low Concentration Dithiothreitol for Negating the Monoclonal Anti-CD38 Interference with Transfusion Compatibility Testing].

Objective: To investigate the effectiveness and safety of low concentration dithiothreitol (DTT) in removing the interference of monoclonal anti-CD38 on transfusion compatibility testing, and develop a reasonable clinical transfusion strategy.

Methods: The blood type, direct antiglobulin testing (DAT) and antibody screening were tested according to standard methods. Antibody screening cells and donor's red blood cells were treated by DTT 0.2, 0.1, 0.05, 0.02, 0.01 and 0.005 mol/L, and antibody screening and cross-matching of serums after monoclonal anti-CD38 treatment were performed by anti-human globulin card.

Results: The 0.01 mol/L DTT at 37℃ for 30 minutes could remove the effect of monoclonal anti-CD38 on antibody screening and cross-matching, meanwhile retain their effectiveness in detecting anti-K, anti-LW, anti-JMH, anti-Lub, anti-e, anti-Dia and anti-Jka alloantibodies. All the 10 patients had no acute or delayed haemolytic transfusion reactions and their routine blood tests showed that the red blood cells transfusion was effective.

Conclusion: The 0.01 mol/L DTT is a safe and effective method for removing the interference of monoclonal anti-CD38 with transfusion compatibility testing, while retaining the ability to detect most alloantibodies.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信