{"title":"一氧化氮对体外生命支持电路的修改是一种安全的治疗方法。","authors":"Carlisle O'Meara,&nbsp;Joseph Timpa,&nbsp;Giles Peek,&nbsp;Melissa Sindelar,&nbsp;Jenny Ross,&nbsp;Justin Raper,&nbsp;Jonathan W Byrnes","doi":"10.1182/ject-142-147","DOIUrl":null,"url":null,"abstract":"<p><p>Nitric oxide (NO) incorporation into the sweep gas of the extracorporeal life support (ECLS) circuit has been proposed as a strategy to ameliorate the insults caused by the systemic inflammatory response. This technical study describes circuit modifications allowing nitric oxide to be incorporated into the circuit and describing and validating the oxygenator sweep flow rates necessary to achieve consistent safe delivery of the therapy. For patients requiring sweep rates less than 2 L/min, a simplified setup, incorporating a pressure relief valve/low flow meter in the gas delivery line, was placed in line between the blender/NO injector module and the NO sampling port/oxygenator. This setup allows titration of sweep to low flows without the need to blend in CO<sub>2</sub> while maintaining the manufacturer recommendation of a minimum 2 L/min of sweep gas to safely deliver NO without nitric dioxide (NO<sub>2</sub>) buildup. This setup was tested three times at three different FiO<sub>2</sub> rates and eleven different desired low sweep flows to test for reproducibility and safety to build an easy-to-follow chart for making gas flow changes. For patients requiring oxygenator sweep rates greater than 2 L/min, the pressure relief valve/low flow meter apparatus is not needed. Maintaining consistent sweep rate and nitric oxide delivery is required in order to utilize this therapy in ECLS. We demonstrated gas delivery across all flow rates. There were no issues delivering 20 parts per million of NO and negligible NO<sub>2</sub> detection. The results from testing this setup were used to provide the specialist a chart at which to set the low flow meter to produce the desired flow rate at which the patient needs. This has been used clinically on 15 ECLS patients with success.</p>","PeriodicalId":39644,"journal":{"name":"Journal of Extra-Corporeal Technology","volume":" ","pages":"142-147"},"PeriodicalIF":0.0000,"publicationDate":"2022-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9302404/pdf/ject-142-147.pdf","citationCount":"0","resultStr":"{\"title\":\"Nitric Oxide on Extracorporeal Life Support-Circuit Modifications for a Safe Therapy.\",\"authors\":\"Carlisle O'Meara,&nbsp;Joseph Timpa,&nbsp;Giles Peek,&nbsp;Melissa Sindelar,&nbsp;Jenny Ross,&nbsp;Justin Raper,&nbsp;Jonathan W Byrnes\",\"doi\":\"10.1182/ject-142-147\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Nitric oxide (NO) incorporation into the sweep gas of the extracorporeal life support (ECLS) circuit has been proposed as a strategy to ameliorate the insults caused by the systemic inflammatory response. This technical study describes circuit modifications allowing nitric oxide to be incorporated into the circuit and describing and validating the oxygenator sweep flow rates necessary to achieve consistent safe delivery of the therapy. For patients requiring sweep rates less than 2 L/min, a simplified setup, incorporating a pressure relief valve/low flow meter in the gas delivery line, was placed in line between the blender/NO injector module and the NO sampling port/oxygenator. This setup allows titration of sweep to low flows without the need to blend in CO<sub>2</sub> while maintaining the manufacturer recommendation of a minimum 2 L/min of sweep gas to safely deliver NO without nitric dioxide (NO<sub>2</sub>) buildup. This setup was tested three times at three different FiO<sub>2</sub> rates and eleven different desired low sweep flows to test for reproducibility and safety to build an easy-to-follow chart for making gas flow changes. For patients requiring oxygenator sweep rates greater than 2 L/min, the pressure relief valve/low flow meter apparatus is not needed. Maintaining consistent sweep rate and nitric oxide delivery is required in order to utilize this therapy in ECLS. We demonstrated gas delivery across all flow rates. There were no issues delivering 20 parts per million of NO and negligible NO<sub>2</sub> detection. The results from testing this setup were used to provide the specialist a chart at which to set the low flow meter to produce the desired flow rate at which the patient needs. This has been used clinically on 15 ECLS patients with success.</p>\",\"PeriodicalId\":39644,\"journal\":{\"name\":\"Journal of Extra-Corporeal Technology\",\"volume\":\" \",\"pages\":\"142-147\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9302404/pdf/ject-142-147.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Extra-Corporeal Technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1182/ject-142-147\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"Health Professions\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Extra-Corporeal Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1182/ject-142-147","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"Health Professions","Score":null,"Total":0}
引用次数: 0

摘要

一氧化氮(NO)加入到体外生命支持(ECLS)回路的扫气中已被提出作为一种改善全身炎症反应引起的损伤的策略。这项技术研究描述了电路的修改,允许一氧化氮被纳入电路,并描述和验证了氧合器扫描流速,以实现一致的安全输送治疗。对于要求扫描速率低于2 L/min的患者,在气体输送管道中加入一个减压阀/低流量计,将其放置在搅拌器/NO注射器模块和NO采样端口/充氧器之间的线路上。这种设置允许滴定扫气到低流量,而不需要混合二氧化碳,同时保持制造商建议的至少2升/分钟的扫气,以安全输送NO,而不会产生一氧化氮(NO2)。该装置在三种不同的FiO2速率和十一种不同的低扫频流量下测试了三次,以测试再现性和安全性,并建立了易于遵循的气体流量变化图表。对于需要氧合器扫描速率大于2l /min的患者,不需要使用减压阀/低流量仪表。为了在ECLS中使用这种疗法,需要保持一致的扫描速率和一氧化氮的递送。我们演示了所有流量下的天然气输送。输送百万分之20的一氧化氮和可以忽略不计的二氧化氮检测是没有问题的。测试该装置的结果用于为专科医生提供一个图表,在该图表上设置低流量仪表以产生患者所需的所需流量。该方法已成功应用于15例ECLS患者。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

Nitric Oxide on Extracorporeal Life Support-Circuit Modifications for a Safe Therapy.

Nitric Oxide on Extracorporeal Life Support-Circuit Modifications for a Safe Therapy.

Nitric oxide (NO) incorporation into the sweep gas of the extracorporeal life support (ECLS) circuit has been proposed as a strategy to ameliorate the insults caused by the systemic inflammatory response. This technical study describes circuit modifications allowing nitric oxide to be incorporated into the circuit and describing and validating the oxygenator sweep flow rates necessary to achieve consistent safe delivery of the therapy. For patients requiring sweep rates less than 2 L/min, a simplified setup, incorporating a pressure relief valve/low flow meter in the gas delivery line, was placed in line between the blender/NO injector module and the NO sampling port/oxygenator. This setup allows titration of sweep to low flows without the need to blend in CO2 while maintaining the manufacturer recommendation of a minimum 2 L/min of sweep gas to safely deliver NO without nitric dioxide (NO2) buildup. This setup was tested three times at three different FiO2 rates and eleven different desired low sweep flows to test for reproducibility and safety to build an easy-to-follow chart for making gas flow changes. For patients requiring oxygenator sweep rates greater than 2 L/min, the pressure relief valve/low flow meter apparatus is not needed. Maintaining consistent sweep rate and nitric oxide delivery is required in order to utilize this therapy in ECLS. We demonstrated gas delivery across all flow rates. There were no issues delivering 20 parts per million of NO and negligible NO2 detection. The results from testing this setup were used to provide the specialist a chart at which to set the low flow meter to produce the desired flow rate at which the patient needs. This has been used clinically on 15 ECLS patients with success.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Extra-Corporeal Technology
Journal of Extra-Corporeal Technology Medicine-Medicine (all)
CiteScore
1.90
自引率
0.00%
发文量
12
期刊介绍: The Journal of Extracorporeal Technology is dedicated to the study and practice of Basic Science and Clinical issues related to extracorporeal circulation. Areas emphasized in the Journal include: •Cardiopulmonary Bypass •Cardiac Surgery •Cardiovascular Anesthesia •Hematology •Blood Management •Physiology •Fluid Dynamics •Laboratory Science •Coagulation and Hematology •Transfusion •Business Practices •Pediatric Perfusion •Total Quality Management • Evidence-Based Practices
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信