{"title":"起搏器与皮下植入式心脏除颤器之间的不良器械相互作用。","authors":"Nicholas Abbott, Aron Bender, Charles Henrikson, Jared Miller, Babak Nazer, Seshadri Balaji","doi":"10.1111/pace.14325","DOIUrl":null,"url":null,"abstract":"<p><p>A 26-year-old patient with prior surgery for Ebstein's anomaly and a pacemaker (placed for post-surgical heart block and poor underlying rhythm) underwent SICD was placement. During defibrillation testing, device-device interaction led to undersensing of ventricular fibrillation with failure to shock. Increasing the pacemaker sensitivity resolved the problem but post shock pacing was unable to capture the heart after both shocks. The patient underwent removal of both the pacemaker and the SICD and placement of a transvenous ICD. Complex device-device interactions can occur in patients who are pacemaker dependent and undergo placement of a SICD.</p>","PeriodicalId":520740,"journal":{"name":"Pacing and clinical electrophysiology : PACE","volume":" ","pages":"1944-1948"},"PeriodicalIF":1.3000,"publicationDate":"2021-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1111/pace.14325","citationCount":"3","resultStr":"{\"title\":\"Adverse device-device interaction between pacemaker and subcutaneous implantable cardiac defibrillator.\",\"authors\":\"Nicholas Abbott, Aron Bender, Charles Henrikson, Jared Miller, Babak Nazer, Seshadri Balaji\",\"doi\":\"10.1111/pace.14325\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>A 26-year-old patient with prior surgery for Ebstein's anomaly and a pacemaker (placed for post-surgical heart block and poor underlying rhythm) underwent SICD was placement. During defibrillation testing, device-device interaction led to undersensing of ventricular fibrillation with failure to shock. Increasing the pacemaker sensitivity resolved the problem but post shock pacing was unable to capture the heart after both shocks. The patient underwent removal of both the pacemaker and the SICD and placement of a transvenous ICD. Complex device-device interactions can occur in patients who are pacemaker dependent and undergo placement of a SICD.</p>\",\"PeriodicalId\":520740,\"journal\":{\"name\":\"Pacing and clinical electrophysiology : PACE\",\"volume\":\" \",\"pages\":\"1944-1948\"},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2021-11-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1111/pace.14325\",\"citationCount\":\"3\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Pacing and clinical electrophysiology : PACE\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1111/pace.14325\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2021/7/31 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pacing and clinical electrophysiology : PACE","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1111/pace.14325","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2021/7/31 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
Adverse device-device interaction between pacemaker and subcutaneous implantable cardiac defibrillator.
A 26-year-old patient with prior surgery for Ebstein's anomaly and a pacemaker (placed for post-surgical heart block and poor underlying rhythm) underwent SICD was placement. During defibrillation testing, device-device interaction led to undersensing of ventricular fibrillation with failure to shock. Increasing the pacemaker sensitivity resolved the problem but post shock pacing was unable to capture the heart after both shocks. The patient underwent removal of both the pacemaker and the SICD and placement of a transvenous ICD. Complex device-device interactions can occur in patients who are pacemaker dependent and undergo placement of a SICD.