{"title":"开角型青光眼静脉注射小梁微搭桥支架植入术联合白内障摘出的早期临床体会。","authors":"Tanner J Ferguson, Zachary Dockter, Adam Bleeker, Kayla L Karpuk, Justin Schweitzer, Mitch J Ibach, John P Berdahl","doi":"10.1186/s40662-020-00194-3","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Retrospective, consecutive case series to evaluate the implantation of two second-generation trabecular microbypass stents in combination with cataract surgery in a real-world, clinical setting.</p><p><strong>Methods: </strong>The series included 56 eyes implanted with the iStent inject device with phacoemulsification. The series consisted of eyes with primary open-angle glaucoma (<i>n</i> = 52) and pseudoexfoliative glaucoma (<i>n</i> = 4). Primary outcome measures included intraocular pressure (IOP) and number of glaucoma medications. Safety outcomes included the need for secondary surgical intervention and the incidence of IOP spikes ≥10 mmHg and ≥ 15 mmHg.</p><p><strong>Results: </strong>IOP was reduced by 21% to 14.7 ± 2.9 mmHg (<i>p</i> < 0.01) at 6 months postoperative from 18.7 ± 5.8 mmHg at baseline. Preoperatively, the mean number of glaucoma medications was 1.5 ± 0.9 and reduced by 39% to 0.9 ± 1.2 (<i>p</i> < 0.01) at 6 months. At 6 months, 68% of eyes had an IOP ≤15 mmHg, increased from 30% at baseline. 55% of eyes were medication-free at 6 months, up from 18% at baseline. There were no severe postoperative complications. No eyes underwent an additional glaucoma procedure.</p><p><strong>Conclusions: </strong>Implantation of the iStent inject device with concomitant cataract surgery effectively provides a sustained reduction in IOP with a markedly improved medication burden out to 6 months postoperative. The safety profile is excellent.</p>","PeriodicalId":520624,"journal":{"name":"Eye and vision (London, England)","volume":" ","pages":"28"},"PeriodicalIF":0.0000,"publicationDate":"2020-05-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1186/s40662-020-00194-3","citationCount":"24","resultStr":"{\"title\":\"iStent inject trabecular microbypass stent implantation with cataract extraction in open-angle glaucoma: early clinical experience.\",\"authors\":\"Tanner J Ferguson, Zachary Dockter, Adam Bleeker, Kayla L Karpuk, Justin Schweitzer, Mitch J Ibach, John P Berdahl\",\"doi\":\"10.1186/s40662-020-00194-3\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Retrospective, consecutive case series to evaluate the implantation of two second-generation trabecular microbypass stents in combination with cataract surgery in a real-world, clinical setting.</p><p><strong>Methods: </strong>The series included 56 eyes implanted with the iStent inject device with phacoemulsification. The series consisted of eyes with primary open-angle glaucoma (<i>n</i> = 52) and pseudoexfoliative glaucoma (<i>n</i> = 4). Primary outcome measures included intraocular pressure (IOP) and number of glaucoma medications. Safety outcomes included the need for secondary surgical intervention and the incidence of IOP spikes ≥10 mmHg and ≥ 15 mmHg.</p><p><strong>Results: </strong>IOP was reduced by 21% to 14.7 ± 2.9 mmHg (<i>p</i> < 0.01) at 6 months postoperative from 18.7 ± 5.8 mmHg at baseline. Preoperatively, the mean number of glaucoma medications was 1.5 ± 0.9 and reduced by 39% to 0.9 ± 1.2 (<i>p</i> < 0.01) at 6 months. At 6 months, 68% of eyes had an IOP ≤15 mmHg, increased from 30% at baseline. 55% of eyes were medication-free at 6 months, up from 18% at baseline. There were no severe postoperative complications. No eyes underwent an additional glaucoma procedure.</p><p><strong>Conclusions: </strong>Implantation of the iStent inject device with concomitant cataract surgery effectively provides a sustained reduction in IOP with a markedly improved medication burden out to 6 months postoperative. The safety profile is excellent.</p>\",\"PeriodicalId\":520624,\"journal\":{\"name\":\"Eye and vision (London, England)\",\"volume\":\" \",\"pages\":\"28\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2020-05-20\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1186/s40662-020-00194-3\",\"citationCount\":\"24\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Eye and vision (London, England)\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1186/s40662-020-00194-3\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2020/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Eye and vision (London, England)","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s40662-020-00194-3","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2020/1/1 0:00:00","PubModel":"eCollection","JCR":"","JCRName":"","Score":null,"Total":0}
iStent inject trabecular microbypass stent implantation with cataract extraction in open-angle glaucoma: early clinical experience.
Background: Retrospective, consecutive case series to evaluate the implantation of two second-generation trabecular microbypass stents in combination with cataract surgery in a real-world, clinical setting.
Methods: The series included 56 eyes implanted with the iStent inject device with phacoemulsification. The series consisted of eyes with primary open-angle glaucoma (n = 52) and pseudoexfoliative glaucoma (n = 4). Primary outcome measures included intraocular pressure (IOP) and number of glaucoma medications. Safety outcomes included the need for secondary surgical intervention and the incidence of IOP spikes ≥10 mmHg and ≥ 15 mmHg.
Results: IOP was reduced by 21% to 14.7 ± 2.9 mmHg (p < 0.01) at 6 months postoperative from 18.7 ± 5.8 mmHg at baseline. Preoperatively, the mean number of glaucoma medications was 1.5 ± 0.9 and reduced by 39% to 0.9 ± 1.2 (p < 0.01) at 6 months. At 6 months, 68% of eyes had an IOP ≤15 mmHg, increased from 30% at baseline. 55% of eyes were medication-free at 6 months, up from 18% at baseline. There were no severe postoperative complications. No eyes underwent an additional glaucoma procedure.
Conclusions: Implantation of the iStent inject device with concomitant cataract surgery effectively provides a sustained reduction in IOP with a markedly improved medication burden out to 6 months postoperative. The safety profile is excellent.