{"title":"使用Icare®Home进行家庭眼压监测及其日模式。","authors":"Mari Sakamoto,&nbsp;Akiyasu Kanamori,&nbsp;Sotaro Mori,&nbsp;Kaori Ueda,&nbsp;Yukako Inoue,&nbsp;Takuji Kurimoto,&nbsp;Yuko Yamada,&nbsp;Makoto Nakamura","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Purpose: To evaluate Icare® HOME for home monitoring of intraocular pressure (IOP) and to observe diurnal IOP patterns for a short term.\nPatients and methods: Twenty six eyes of 16 patients with open angle glaucoma were included. After instruction and practice in using Icare® HOME, patients were asked to measure their IOP by themselves at home or in a sick room. Patients measured their IOP four times a day with a four hour interval between each measurement for three days. Patients repeated measurements until three successful measurements were obtained at each time. Intra-rater reliability was assessed calculating intraclass correlation coefficient (ICC) of the three measurements. Diurnal IOP patterns were assessed for three days.\nResults: No adverse events occurred. ICC of three measurements was 0.76 (95% confidence coefficient; 0.71 to 0.81). The median of the difference between the highest and the lowest IOP during a day was 4.26 mmHg (95% CI; 4.06 to 4.67). Repeatable patterns were not found in diurnal IOP for three days. Three out of ten patients who answered the questionnaire after measurements found difficulties in handling the device.\nConclusions: Although Icare® HOME is safe and could be used for home monitoring of IOP, some patients had difficulties in using the instrument. Diurnal IOP patterns did not show repeatability during a short term.</p>","PeriodicalId":19670,"journal":{"name":"Nippon Ganka Gakkai zasshi","volume":"121 4","pages":"366-72"},"PeriodicalIF":0.0000,"publicationDate":"2017-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Home Monitoring of Intraocular Pressure and it's Diurnal Patterns Using Icare® HOME.\",\"authors\":\"Mari Sakamoto,&nbsp;Akiyasu Kanamori,&nbsp;Sotaro Mori,&nbsp;Kaori Ueda,&nbsp;Yukako Inoue,&nbsp;Takuji Kurimoto,&nbsp;Yuko Yamada,&nbsp;Makoto Nakamura\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Purpose: To evaluate Icare® HOME for home monitoring of intraocular pressure (IOP) and to observe diurnal IOP patterns for a short term.\\nPatients and methods: Twenty six eyes of 16 patients with open angle glaucoma were included. After instruction and practice in using Icare® HOME, patients were asked to measure their IOP by themselves at home or in a sick room. Patients measured their IOP four times a day with a four hour interval between each measurement for three days. Patients repeated measurements until three successful measurements were obtained at each time. Intra-rater reliability was assessed calculating intraclass correlation coefficient (ICC) of the three measurements. Diurnal IOP patterns were assessed for three days.\\nResults: No adverse events occurred. ICC of three measurements was 0.76 (95% confidence coefficient; 0.71 to 0.81). The median of the difference between the highest and the lowest IOP during a day was 4.26 mmHg (95% CI; 4.06 to 4.67). Repeatable patterns were not found in diurnal IOP for three days. Three out of ten patients who answered the questionnaire after measurements found difficulties in handling the device.\\nConclusions: Although Icare® HOME is safe and could be used for home monitoring of IOP, some patients had difficulties in using the instrument. Diurnal IOP patterns did not show repeatability during a short term.</p>\",\"PeriodicalId\":19670,\"journal\":{\"name\":\"Nippon Ganka Gakkai zasshi\",\"volume\":\"121 4\",\"pages\":\"366-72\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2017-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Nippon Ganka Gakkai zasshi\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nippon Ganka Gakkai zasshi","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:评估Icare®HOME在家庭监测眼压(IOP)中的应用,并观察短期内的每日眼压模式。患者与方法:选取16例开角型青光眼患者的26只眼。在使用Icare®HOME的指导和实践之后,患者被要求在家中或病房自行测量他们的IOP。患者每天测量四次IOP,每次测量间隔四小时,持续三天。患者重复测量,直到每次获得三次成功测量。通过计算三个测量值的类内相关系数(ICC)来评估组内信度。三天内评估每日眼压模式。结果:无不良事件发生。三个测量的ICC为0.76(95%置信系数;0.71 - 0.81)。一天内最高和最低眼压差的中位数为4.26 mmHg (95% CI;4.06 ~ 4.67)。连续3天未发现可重复的IOP模式。在测量后回答问卷的患者中,有十分之三的人在使用该设备时感到困难。结论:尽管Icare®HOME是安全的,可用于家庭IOP监测,但一些患者在使用该仪器时存在困难。在短期内,每日IOP模式不具有可重复性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Home Monitoring of Intraocular Pressure and it's Diurnal Patterns Using Icare® HOME.

Purpose: To evaluate Icare® HOME for home monitoring of intraocular pressure (IOP) and to observe diurnal IOP patterns for a short term. Patients and methods: Twenty six eyes of 16 patients with open angle glaucoma were included. After instruction and practice in using Icare® HOME, patients were asked to measure their IOP by themselves at home or in a sick room. Patients measured their IOP four times a day with a four hour interval between each measurement for three days. Patients repeated measurements until three successful measurements were obtained at each time. Intra-rater reliability was assessed calculating intraclass correlation coefficient (ICC) of the three measurements. Diurnal IOP patterns were assessed for three days. Results: No adverse events occurred. ICC of three measurements was 0.76 (95% confidence coefficient; 0.71 to 0.81). The median of the difference between the highest and the lowest IOP during a day was 4.26 mmHg (95% CI; 4.06 to 4.67). Repeatable patterns were not found in diurnal IOP for three days. Three out of ten patients who answered the questionnaire after measurements found difficulties in handling the device. Conclusions: Although Icare® HOME is safe and could be used for home monitoring of IOP, some patients had difficulties in using the instrument. Diurnal IOP patterns did not show repeatability during a short term.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信