{"title":"[HIFU下血流停止对静脉损伤的影响]。","authors":"N N Petrishchev, D U Semenov, A U Tsibin, G U Yukina, A E Berkovich, A A Bursian, A B Belyankin, K U Senchik","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The purpose. In the study we investigated the impact of the partial blood flow shutdown on structural changes in the rabbit vena cava posterior wall after exposure to high-intensity focused ultrasound (HIFU). Methods. Ultrasound Exposure: frequency of 1.65 MHz, the ultrasound intensity in the focus of 13.6 kW/cm2, the area of the focal spot 1 mm2, continuous ultrasound, exposure for 3 seconds. Results. Immediately after HIFU exposure all layers of the vein wall showed characteristic signs of thermal damage. A week after exposure structural changes in the intima, media and adventitia was minimal in the part of vessel with preserved blood flow, and after 4 weeks the changes were not revealed. A week after HIFU exposure partial endothelium destruction, destruction of myocytes, disorganization and consolidation of collagen fibers of the adventitia were observed in an isolated segment of the vessel, and in 4 weeks endothelium restored and signs of damage in media and adventitia persisted, but were less obvious than in a week after exposure. Conclusion. The shutdown of blood flow after exposure to HIFU promotes persistent changes in the vein wall. Vein compression appears to be necessary for the obliteration of the vessel, when using HIFU-technology.</p>","PeriodicalId":19857,"journal":{"name":"Patologicheskaia fiziologiia i eksperimental'naia terapiia","volume":"61 2","pages":"46-50"},"PeriodicalIF":0.0000,"publicationDate":"2017-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Effekt of blood flow shutdown on the vein damage when exposed to HIFU].\",\"authors\":\"N N Petrishchev, D U Semenov, A U Tsibin, G U Yukina, A E Berkovich, A A Bursian, A B Belyankin, K U Senchik\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The purpose. In the study we investigated the impact of the partial blood flow shutdown on structural changes in the rabbit vena cava posterior wall after exposure to high-intensity focused ultrasound (HIFU). Methods. Ultrasound Exposure: frequency of 1.65 MHz, the ultrasound intensity in the focus of 13.6 kW/cm2, the area of the focal spot 1 mm2, continuous ultrasound, exposure for 3 seconds. Results. Immediately after HIFU exposure all layers of the vein wall showed characteristic signs of thermal damage. A week after exposure structural changes in the intima, media and adventitia was minimal in the part of vessel with preserved blood flow, and after 4 weeks the changes were not revealed. A week after HIFU exposure partial endothelium destruction, destruction of myocytes, disorganization and consolidation of collagen fibers of the adventitia were observed in an isolated segment of the vessel, and in 4 weeks endothelium restored and signs of damage in media and adventitia persisted, but were less obvious than in a week after exposure. Conclusion. The shutdown of blood flow after exposure to HIFU promotes persistent changes in the vein wall. Vein compression appears to be necessary for the obliteration of the vessel, when using HIFU-technology.</p>\",\"PeriodicalId\":19857,\"journal\":{\"name\":\"Patologicheskaia fiziologiia i eksperimental'naia terapiia\",\"volume\":\"61 2\",\"pages\":\"46-50\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2017-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Patologicheskaia fiziologiia i eksperimental'naia terapiia\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Patologicheskaia fiziologiia i eksperimental'naia terapiia","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
[Effekt of blood flow shutdown on the vein damage when exposed to HIFU].
The purpose. In the study we investigated the impact of the partial blood flow shutdown on structural changes in the rabbit vena cava posterior wall after exposure to high-intensity focused ultrasound (HIFU). Methods. Ultrasound Exposure: frequency of 1.65 MHz, the ultrasound intensity in the focus of 13.6 kW/cm2, the area of the focal spot 1 mm2, continuous ultrasound, exposure for 3 seconds. Results. Immediately after HIFU exposure all layers of the vein wall showed characteristic signs of thermal damage. A week after exposure structural changes in the intima, media and adventitia was minimal in the part of vessel with preserved blood flow, and after 4 weeks the changes were not revealed. A week after HIFU exposure partial endothelium destruction, destruction of myocytes, disorganization and consolidation of collagen fibers of the adventitia were observed in an isolated segment of the vessel, and in 4 weeks endothelium restored and signs of damage in media and adventitia persisted, but were less obvious than in a week after exposure. Conclusion. The shutdown of blood flow after exposure to HIFU promotes persistent changes in the vein wall. Vein compression appears to be necessary for the obliteration of the vessel, when using HIFU-technology.