{"title":"celltac自动血液学分析仪血小板计数准确性的验证。","authors":"Yutaka Nagai, Hiroshi Kondo, Noriyuki Tatsumi","doi":"10.1155/JAMMC.2005.235","DOIUrl":null,"url":null,"abstract":"<p><p>Rapid and accurate analysis of platelet count plays an important role in evaluating hemorrhagic status. Therefore, we evaluated platelet counting performance of a hematology analyzer, Celltac F (MEK-8222, Nihon Kohden Corporation, Tokyo, Japan), that features easy use with low reagent consumption and high throughput while occupying minimal space in the clinical laboratory. All blood samples were anticoagulated with dipotassium ethylenediaminetetraacetic acid (EDTA-2K). The samples were stored at room temperature (18(;)C-22(;)C) and tested within 4 hours of phlebotomy. We evaluated the counting ability of the Celltac F hematology analyzer by comparing it with the platelet counts obtained by the flow cytometry method that ISLH and ICSH recommended, and also the manual visual method by Unopette (Becton Dickinson Vacutainer Systems). The ICSH/ISLH reference method is based on the fact that platelets can be stained with monoclonal antibodies to CD41 and/or CD61. The dilution ratio was optimized after the precision, coincidence events, and debris counts were confirmed by the reference method. Good correlation of platelet count between the Celltac F and the ICSH/ISLH reference method (r = 0.99, and the manual visual method (r= 0.93) were obtained. The regressions were y = 0.90 x+9.0 and y=1.11x+8.4, respectively. We conclude that the Celltac F hematology analyzer for platelet counting was well suited to the ICSH/ISLH reference method for rapidness and reliability.</p>","PeriodicalId":15248,"journal":{"name":"Journal of Automated Methods & Management in Chemistry","volume":"2005 ","pages":"235-9"},"PeriodicalIF":0.0000,"publicationDate":"2005-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1155/JAMMC.2005.235","citationCount":"5","resultStr":"{\"title\":\"Validation of platelet counting accuracy with the celltac f automated hematology analyzer.\",\"authors\":\"Yutaka Nagai, Hiroshi Kondo, Noriyuki Tatsumi\",\"doi\":\"10.1155/JAMMC.2005.235\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Rapid and accurate analysis of platelet count plays an important role in evaluating hemorrhagic status. Therefore, we evaluated platelet counting performance of a hematology analyzer, Celltac F (MEK-8222, Nihon Kohden Corporation, Tokyo, Japan), that features easy use with low reagent consumption and high throughput while occupying minimal space in the clinical laboratory. All blood samples were anticoagulated with dipotassium ethylenediaminetetraacetic acid (EDTA-2K). The samples were stored at room temperature (18(;)C-22(;)C) and tested within 4 hours of phlebotomy. We evaluated the counting ability of the Celltac F hematology analyzer by comparing it with the platelet counts obtained by the flow cytometry method that ISLH and ICSH recommended, and also the manual visual method by Unopette (Becton Dickinson Vacutainer Systems). The ICSH/ISLH reference method is based on the fact that platelets can be stained with monoclonal antibodies to CD41 and/or CD61. The dilution ratio was optimized after the precision, coincidence events, and debris counts were confirmed by the reference method. Good correlation of platelet count between the Celltac F and the ICSH/ISLH reference method (r = 0.99, and the manual visual method (r= 0.93) were obtained. The regressions were y = 0.90 x+9.0 and y=1.11x+8.4, respectively. We conclude that the Celltac F hematology analyzer for platelet counting was well suited to the ICSH/ISLH reference method for rapidness and reliability.</p>\",\"PeriodicalId\":15248,\"journal\":{\"name\":\"Journal of Automated Methods & Management in Chemistry\",\"volume\":\"2005 \",\"pages\":\"235-9\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2005-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1155/JAMMC.2005.235\",\"citationCount\":\"5\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Automated Methods & Management in Chemistry\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1155/JAMMC.2005.235\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Automated Methods & Management in Chemistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1155/JAMMC.2005.235","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 5
摘要
快速准确地分析血小板计数在评估出血状态中起着重要作用。因此,我们评估了血液学分析仪Celltac F (MEK-8222, Nihon Kohden Corporation, Tokyo, Japan)的血小板计数性能,该分析仪易于使用,试剂消耗少,通量高,同时占用临床实验室最小的空间。所有血样均用乙二胺四乙酸二钾(EDTA-2K)抗凝。样品在室温(18(;)C-22(;)C)下保存,并在采血后4小时内进行检测。我们通过将Celltac F血液学分析仪与ISLH和ICSH推荐的流式细胞术方法获得的血小板计数以及Unopette (Becton Dickinson Vacutainer Systems)的手动视觉方法进行比较,来评估其计数能力。ICSH/ISLH参考方法是基于血小板可以用CD41和/或CD61单克隆抗体染色的事实。通过参考方法确定精密度、符合事件和碎片计数,优化稀释比。Celltac F血小板计数与ICSH/ISLH参考法(r= 0.99)和人工目测法(r= 0.93)具有良好的相关性。回归分别为y= 0.90 x+9.0和y=1.11x+8.4。我们得出结论,Celltac F血液学分析仪用于血小板计数在快速和可靠性方面非常适合ICSH/ISLH参考方法。
Validation of platelet counting accuracy with the celltac f automated hematology analyzer.
Rapid and accurate analysis of platelet count plays an important role in evaluating hemorrhagic status. Therefore, we evaluated platelet counting performance of a hematology analyzer, Celltac F (MEK-8222, Nihon Kohden Corporation, Tokyo, Japan), that features easy use with low reagent consumption and high throughput while occupying minimal space in the clinical laboratory. All blood samples were anticoagulated with dipotassium ethylenediaminetetraacetic acid (EDTA-2K). The samples were stored at room temperature (18(;)C-22(;)C) and tested within 4 hours of phlebotomy. We evaluated the counting ability of the Celltac F hematology analyzer by comparing it with the platelet counts obtained by the flow cytometry method that ISLH and ICSH recommended, and also the manual visual method by Unopette (Becton Dickinson Vacutainer Systems). The ICSH/ISLH reference method is based on the fact that platelets can be stained with monoclonal antibodies to CD41 and/or CD61. The dilution ratio was optimized after the precision, coincidence events, and debris counts were confirmed by the reference method. Good correlation of platelet count between the Celltac F and the ICSH/ISLH reference method (r = 0.99, and the manual visual method (r= 0.93) were obtained. The regressions were y = 0.90 x+9.0 and y=1.11x+8.4, respectively. We conclude that the Celltac F hematology analyzer for platelet counting was well suited to the ICSH/ISLH reference method for rapidness and reliability.