V. Femery , F. Potdevin , A. Thevenon , P. Moretto
{"title":"开发和验证一种新的足底压力监测装置:应用于足部排出","authors":"V. Femery , F. Potdevin , A. Thevenon , P. Moretto","doi":"10.1016/j.annrmp.2008.01.007","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><p>To validate a new plantar pressure control device able to detect excessive plantar pressure on-line and to improve the subject's awareness of the gait pattern in order to correct and optimize the load distribution patterns. The aim of this study was to verify the feasibility of a deliberate and partial unload of the first metatarsal head.</p></div><div><h3>Material and method</h3><p>Eight healthy subjects were equipped with the plantar pressure sensors integrated in customized insoles. They were told to unload the first metatarsal head of the right foot by 5%. During the trials, the auditory and visual signals inform the subject of an excessive, insufficient or correct unload.</p></div><div><h3>Results</h3><p>Five subjects over eight succeeding in modifying significantly their gait pattern and in unloading by 5% the first metatarsal head. The unload is effective in 70.7% of the trials. However, the subjects were spontaneously inclined to reduce peak pressure under the first metatarsal head beyond the critical peak plantar pressure threshold (48.2% of the trials).</p></div><div><h3>Discussion – conclusion</h3><p>The results showed the feasibility of a partial and deliberate foot unloading in using biofeedback device but emphasized the subject's difficulty to control a discriminated and accurate unload of the first metatarsal head. That points out the necessity of learning period. The baropodometric biofeedback rehabilitation focuses on the subject's capacities to modify its own locomotor pattern. It would be used as primary and secondary prevention means of diabetic foot ulceration.</p></div>","PeriodicalId":72206,"journal":{"name":"Annales de readaptation et de medecine physique : revue scientifique de la Societe francaise de reeducation fonctionnelle de readaptation et de medecine physique","volume":"51 4","pages":"Pages 231-237"},"PeriodicalIF":0.0000,"publicationDate":"2008-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/j.annrmp.2008.01.007","citationCount":"12","resultStr":"{\"title\":\"Élaboration et validation d’un nouveau dispositif de surveillance des pressions plantaires : application à la décharge du pied\",\"authors\":\"V. Femery , F. Potdevin , A. Thevenon , P. Moretto\",\"doi\":\"10.1016/j.annrmp.2008.01.007\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Objective</h3><p>To validate a new plantar pressure control device able to detect excessive plantar pressure on-line and to improve the subject's awareness of the gait pattern in order to correct and optimize the load distribution patterns. The aim of this study was to verify the feasibility of a deliberate and partial unload of the first metatarsal head.</p></div><div><h3>Material and method</h3><p>Eight healthy subjects were equipped with the plantar pressure sensors integrated in customized insoles. They were told to unload the first metatarsal head of the right foot by 5%. During the trials, the auditory and visual signals inform the subject of an excessive, insufficient or correct unload.</p></div><div><h3>Results</h3><p>Five subjects over eight succeeding in modifying significantly their gait pattern and in unloading by 5% the first metatarsal head. The unload is effective in 70.7% of the trials. However, the subjects were spontaneously inclined to reduce peak pressure under the first metatarsal head beyond the critical peak plantar pressure threshold (48.2% of the trials).</p></div><div><h3>Discussion – conclusion</h3><p>The results showed the feasibility of a partial and deliberate foot unloading in using biofeedback device but emphasized the subject's difficulty to control a discriminated and accurate unload of the first metatarsal head. That points out the necessity of learning period. The baropodometric biofeedback rehabilitation focuses on the subject's capacities to modify its own locomotor pattern. It would be used as primary and secondary prevention means of diabetic foot ulceration.</p></div>\",\"PeriodicalId\":72206,\"journal\":{\"name\":\"Annales de readaptation et de medecine physique : revue scientifique de la Societe francaise de reeducation fonctionnelle de readaptation et de medecine physique\",\"volume\":\"51 4\",\"pages\":\"Pages 231-237\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2008-05-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1016/j.annrmp.2008.01.007\",\"citationCount\":\"12\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Annales de readaptation et de medecine physique : revue scientifique de la Societe francaise de reeducation fonctionnelle de readaptation et de medecine physique\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0168605408000135\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annales de readaptation et de medecine physique : revue scientifique de la Societe francaise de reeducation fonctionnelle de readaptation et de medecine physique","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0168605408000135","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Élaboration et validation d’un nouveau dispositif de surveillance des pressions plantaires : application à la décharge du pied
Objective
To validate a new plantar pressure control device able to detect excessive plantar pressure on-line and to improve the subject's awareness of the gait pattern in order to correct and optimize the load distribution patterns. The aim of this study was to verify the feasibility of a deliberate and partial unload of the first metatarsal head.
Material and method
Eight healthy subjects were equipped with the plantar pressure sensors integrated in customized insoles. They were told to unload the first metatarsal head of the right foot by 5%. During the trials, the auditory and visual signals inform the subject of an excessive, insufficient or correct unload.
Results
Five subjects over eight succeeding in modifying significantly their gait pattern and in unloading by 5% the first metatarsal head. The unload is effective in 70.7% of the trials. However, the subjects were spontaneously inclined to reduce peak pressure under the first metatarsal head beyond the critical peak plantar pressure threshold (48.2% of the trials).
Discussion – conclusion
The results showed the feasibility of a partial and deliberate foot unloading in using biofeedback device but emphasized the subject's difficulty to control a discriminated and accurate unload of the first metatarsal head. That points out the necessity of learning period. The baropodometric biofeedback rehabilitation focuses on the subject's capacities to modify its own locomotor pattern. It would be used as primary and secondary prevention means of diabetic foot ulceration.