{"title":"干涉螺钉直径对软组织移植固定的影响。","authors":"Suk Namkoong,&nbsp;Christian S Heywood,&nbsp;Jonathan T Bravman,&nbsp;Kazuho Ieyasa,&nbsp;Frederick J Kummer,&nbsp;Robert J Meislin","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Tibial fixation of soft-tissue grafts is a weak link in anterior cruciate ligament reconstruction. Previous studies have examined varying interference screw lengths, screw types and tunnel sizes as means to improve graft fixation. We hypothesized that increasing interference screw diameter would significantly increase the maximum load to failure of the graft and decrease the graft's initial slippage. Seventy tibialis anterior and tibialis posterior tendons were divided, looped, trimmed, and sutured to simulate 4-strand hamstring grafts. These grafts were then inserted into composite bone blocks having pre-drilled 8 mm holes and fixed with 8 mm, 9 mm, 10 mm, 11 mm, or 12 mm interference screws. Fourteen grafts were tested for each screw size. The graft was first cyclically loaded from 50 N to 250 N at 0.3 Hz for 100 cycles to measure graft slippage. The graft was then tested to failure at 0.5 mm/sec to determine the maximum load to failure and mode of failure. Graft slippage was not affected by screw diameter. Maximum load to failure increased with increasing screw diameter up to 11 mm; 11 mm screw fixation was 20% stronger than 8 mm screw fixation. In this model, no increase in graft fixation was seen in by increasing interference screw diameter beyond 3 mm of the tunnel diameter.</p>","PeriodicalId":77050,"journal":{"name":"Bulletin (Hospital for Joint Diseases (New York, N.Y.))","volume":"63 3-4","pages":"153-5"},"PeriodicalIF":0.0000,"publicationDate":"2006-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"The effect of interference screw diameter on soft tissue graft fixation.\",\"authors\":\"Suk Namkoong,&nbsp;Christian S Heywood,&nbsp;Jonathan T Bravman,&nbsp;Kazuho Ieyasa,&nbsp;Frederick J Kummer,&nbsp;Robert J Meislin\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Tibial fixation of soft-tissue grafts is a weak link in anterior cruciate ligament reconstruction. Previous studies have examined varying interference screw lengths, screw types and tunnel sizes as means to improve graft fixation. We hypothesized that increasing interference screw diameter would significantly increase the maximum load to failure of the graft and decrease the graft's initial slippage. Seventy tibialis anterior and tibialis posterior tendons were divided, looped, trimmed, and sutured to simulate 4-strand hamstring grafts. These grafts were then inserted into composite bone blocks having pre-drilled 8 mm holes and fixed with 8 mm, 9 mm, 10 mm, 11 mm, or 12 mm interference screws. Fourteen grafts were tested for each screw size. The graft was first cyclically loaded from 50 N to 250 N at 0.3 Hz for 100 cycles to measure graft slippage. The graft was then tested to failure at 0.5 mm/sec to determine the maximum load to failure and mode of failure. Graft slippage was not affected by screw diameter. Maximum load to failure increased with increasing screw diameter up to 11 mm; 11 mm screw fixation was 20% stronger than 8 mm screw fixation. In this model, no increase in graft fixation was seen in by increasing interference screw diameter beyond 3 mm of the tunnel diameter.</p>\",\"PeriodicalId\":77050,\"journal\":{\"name\":\"Bulletin (Hospital for Joint Diseases (New York, N.Y.))\",\"volume\":\"63 3-4\",\"pages\":\"153-5\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2006-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Bulletin (Hospital for Joint Diseases (New York, N.Y.))\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Bulletin (Hospital for Joint Diseases (New York, N.Y.))","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

胫骨软组织移植固定是前交叉韧带重建的薄弱环节。以前的研究已经研究了不同的干涉螺钉长度、螺钉类型和隧道大小作为改善移植物固定的手段。我们假设增加干涉螺钉直径会显著增加接枝破坏的最大载荷,并降低接枝的初始滑移。将70根胫骨前、后肌腱分离、环化、修剪、缝合,模拟4股腘绳肌移植。然后将这些移植物插入预钻8mm孔的复合骨块中,并用8mm、9mm、10mm、11mm或12mm干涉螺钉固定。每种螺钉大小分别检测14个移植物。接枝剂首先在0.3 Hz下从50 N到250 N循环加载100次,以测量接枝剂的滑移。接枝然后以0.5 mm/sec的速度进行失效测试,以确定失效的最大载荷和失效模式。移植物滑移不受螺钉直径的影响。随着螺杆直径的增大,最大失效载荷增大至11 mm;11mm螺钉固定强度比8mm螺钉固定强度高20%。在本模型中,增加干涉螺钉直径超过隧道直径3mm后,植骨固定未见增加。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
The effect of interference screw diameter on soft tissue graft fixation.

Tibial fixation of soft-tissue grafts is a weak link in anterior cruciate ligament reconstruction. Previous studies have examined varying interference screw lengths, screw types and tunnel sizes as means to improve graft fixation. We hypothesized that increasing interference screw diameter would significantly increase the maximum load to failure of the graft and decrease the graft's initial slippage. Seventy tibialis anterior and tibialis posterior tendons were divided, looped, trimmed, and sutured to simulate 4-strand hamstring grafts. These grafts were then inserted into composite bone blocks having pre-drilled 8 mm holes and fixed with 8 mm, 9 mm, 10 mm, 11 mm, or 12 mm interference screws. Fourteen grafts were tested for each screw size. The graft was first cyclically loaded from 50 N to 250 N at 0.3 Hz for 100 cycles to measure graft slippage. The graft was then tested to failure at 0.5 mm/sec to determine the maximum load to failure and mode of failure. Graft slippage was not affected by screw diameter. Maximum load to failure increased with increasing screw diameter up to 11 mm; 11 mm screw fixation was 20% stronger than 8 mm screw fixation. In this model, no increase in graft fixation was seen in by increasing interference screw diameter beyond 3 mm of the tunnel diameter.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信