{"title":"宫腔镜下绝育:基于新型线圈导管输送系统的结果","authors":"Dr. John F. Kerin M.D. ,&nbsp;Dr. David N. Munday M.D. ,&nbsp;Dr. Martin G. Ritossa M.D. ,&nbsp;Dr. Andrew Pesce M.D. ,&nbsp;Dr. David Rosen M.D.","doi":"10.1016/S1074-3804(05)60056-4","DOIUrl":null,"url":null,"abstract":"<div><h3>Study Objective</h3><p>To assess the safety and placement effectiveness of a new delivery catheter for the Essure micro-insert hysteroscopic sterilization system.</p></div><div><h3>Design</h3><p>Prospective, multicenter, single-arm clinical study. (Canadian Task Force classification II-3).</p></div><div><h3>Setting</h3><p>Hospital-based clinical research centers.</p></div><div><h3>Patients</h3><p>One hundred two women of reproductive age and proven fertility.</p></div><div><h3>Intervention</h3><p>A new coil catheter delivery system was used for micro-insert placement.</p></div><div><h3>Measurements and Main Results</h3><p>The bilateral placement of micro-inserts using the coil catheter delivery system occurred in 100 women out of 102 attempts (98%). The two women who did not have successful micro-insert placement were subsequently shown to have proximal tubal stenotic disease.</p></div><div><h3>Conclusion</h3><p>These findings indicate that the introduction of the coil catheter delivery system for Essure hysteroscopic sterilization improves the micro-insert bilateral placement rate when compared with previous studies. It is hypothesized that the coil catheter has superior navigational properties, particularly within tortuous or narrowed tubal lumens.</p></div>","PeriodicalId":79466,"journal":{"name":"The Journal of the American Association of Gynecologic Laparoscopists","volume":"11 3","pages":"Pages 388-393"},"PeriodicalIF":0.0000,"publicationDate":"2004-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/S1074-3804(05)60056-4","citationCount":"75","resultStr":"{\"title\":\"Essure Hysteroscopic Sterilization: Results Based on Utilizing a New Coil Catheter Delivery System\",\"authors\":\"Dr. John F. Kerin M.D. ,&nbsp;Dr. David N. Munday M.D. ,&nbsp;Dr. Martin G. Ritossa M.D. ,&nbsp;Dr. Andrew Pesce M.D. ,&nbsp;Dr. David Rosen M.D.\",\"doi\":\"10.1016/S1074-3804(05)60056-4\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Study Objective</h3><p>To assess the safety and placement effectiveness of a new delivery catheter for the Essure micro-insert hysteroscopic sterilization system.</p></div><div><h3>Design</h3><p>Prospective, multicenter, single-arm clinical study. (Canadian Task Force classification II-3).</p></div><div><h3>Setting</h3><p>Hospital-based clinical research centers.</p></div><div><h3>Patients</h3><p>One hundred two women of reproductive age and proven fertility.</p></div><div><h3>Intervention</h3><p>A new coil catheter delivery system was used for micro-insert placement.</p></div><div><h3>Measurements and Main Results</h3><p>The bilateral placement of micro-inserts using the coil catheter delivery system occurred in 100 women out of 102 attempts (98%). The two women who did not have successful micro-insert placement were subsequently shown to have proximal tubal stenotic disease.</p></div><div><h3>Conclusion</h3><p>These findings indicate that the introduction of the coil catheter delivery system for Essure hysteroscopic sterilization improves the micro-insert bilateral placement rate when compared with previous studies. It is hypothesized that the coil catheter has superior navigational properties, particularly within tortuous or narrowed tubal lumens.</p></div>\",\"PeriodicalId\":79466,\"journal\":{\"name\":\"The Journal of the American Association of Gynecologic Laparoscopists\",\"volume\":\"11 3\",\"pages\":\"Pages 388-393\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2004-08-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1016/S1074-3804(05)60056-4\",\"citationCount\":\"75\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"The Journal of the American Association of Gynecologic Laparoscopists\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1074380405600564\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"The Journal of the American Association of Gynecologic Laparoscopists","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1074380405600564","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 75

摘要

目的评价一种新型置入导管用于微创宫腔镜灭菌系统的安全性和放置效果。前瞻性、多中心、单臂临床研究。(加拿大工作队分类II-3)。建立以医院为基础的临床研究中心。病人:102名育龄妇女,经证实有生育能力。介入:采用新型线圈导管输送系统进行微植入。测量结果和主要结果102例患者中有100例(98%)采用线圈导管输送系统双侧放置微植入物。两名没有成功植入微植入物的妇女随后被证明患有近端输卵管狭窄疾病。结论与以往研究相比,采用螺旋管输尿管系统进行Essure宫腔镜灭菌可提高微植入物双侧置入率。假设线圈导管具有优越的导航性能,特别是在弯曲或狭窄的管腔内。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Essure Hysteroscopic Sterilization: Results Based on Utilizing a New Coil Catheter Delivery System

Study Objective

To assess the safety and placement effectiveness of a new delivery catheter for the Essure micro-insert hysteroscopic sterilization system.

Design

Prospective, multicenter, single-arm clinical study. (Canadian Task Force classification II-3).

Setting

Hospital-based clinical research centers.

Patients

One hundred two women of reproductive age and proven fertility.

Intervention

A new coil catheter delivery system was used for micro-insert placement.

Measurements and Main Results

The bilateral placement of micro-inserts using the coil catheter delivery system occurred in 100 women out of 102 attempts (98%). The two women who did not have successful micro-insert placement were subsequently shown to have proximal tubal stenotic disease.

Conclusion

These findings indicate that the introduction of the coil catheter delivery system for Essure hysteroscopic sterilization improves the micro-insert bilateral placement rate when compared with previous studies. It is hypothesized that the coil catheter has superior navigational properties, particularly within tortuous or narrowed tubal lumens.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信