Farhad Mazoochian, Christoph Pellengahr, Armin Huber, Jörn Kircher, Hans Jürgen Refior, Volkmar Jansson
{"title":"使用caspar系统植入THR的低准确性:10髋的前倾测量。","authors":"Farhad Mazoochian, Christoph Pellengahr, Armin Huber, Jörn Kircher, Hans Jürgen Refior, Volkmar Jansson","doi":"10.1080/00016470410001178","DOIUrl":null,"url":null,"abstract":"<p><p>We studied 10 patients by means of pre- and postoperative CT who underwent computer-assisted total hip replacement using the CASPAR-system (OrthoMaquet GmbH, Rastatt, Germany). The anteversion angles of the stem and translational deviations measured after surgery were compared to the preoperatively planned ones as a quality control. We found an average difference of 7.8 degrees (SD 6.3 degrees; 95% CI 3.3 degrees-12.3 degrees) in the angles and an average difference of 1.8 (SD 1.7; 95% CI 0.6-3.0) mm and 1.2 (SD 1.4: 95% CI 0.25-2.2), respectively, in the medial and lateral deviation. In conclusion, we could not confirm the same accuracy of implant position as that claimed by the manufacturer.</p>","PeriodicalId":75403,"journal":{"name":"Acta orthopaedica Scandinavica","volume":"75 3","pages":"261-4"},"PeriodicalIF":0.0000,"publicationDate":"2004-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1080/00016470410001178","citationCount":"14","resultStr":"{\"title\":\"Low accuracy of stem implantation in THR using the CASPAR-system: anteversion measurements in 10 hips.\",\"authors\":\"Farhad Mazoochian, Christoph Pellengahr, Armin Huber, Jörn Kircher, Hans Jürgen Refior, Volkmar Jansson\",\"doi\":\"10.1080/00016470410001178\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>We studied 10 patients by means of pre- and postoperative CT who underwent computer-assisted total hip replacement using the CASPAR-system (OrthoMaquet GmbH, Rastatt, Germany). The anteversion angles of the stem and translational deviations measured after surgery were compared to the preoperatively planned ones as a quality control. We found an average difference of 7.8 degrees (SD 6.3 degrees; 95% CI 3.3 degrees-12.3 degrees) in the angles and an average difference of 1.8 (SD 1.7; 95% CI 0.6-3.0) mm and 1.2 (SD 1.4: 95% CI 0.25-2.2), respectively, in the medial and lateral deviation. In conclusion, we could not confirm the same accuracy of implant position as that claimed by the manufacturer.</p>\",\"PeriodicalId\":75403,\"journal\":{\"name\":\"Acta orthopaedica Scandinavica\",\"volume\":\"75 3\",\"pages\":\"261-4\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2004-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1080/00016470410001178\",\"citationCount\":\"14\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Acta orthopaedica Scandinavica\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1080/00016470410001178\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta orthopaedica Scandinavica","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1080/00016470410001178","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 14
摘要
我们研究了10例使用caspar系统(OrthoMaquet GmbH, Rastatt, Germany)行计算机辅助全髋关节置换术的患者,通过术前和术后CT检查。将手术后测量的椎弓根前倾角和平移偏差与术前计划的进行比较,作为质量控制。我们发现平均差异为7.8度(标准差6.3度;95% CI 3.3度-12.3度),平均差异为1.8 (SD 1.7;95% CI 0.6-3.0) mm和1.2 (SD 1.4: 95% CI 0.25-2.2),分别为内侧和外侧偏。总之,我们无法证实种植体位置的准确性与制造商声称的相同。
Low accuracy of stem implantation in THR using the CASPAR-system: anteversion measurements in 10 hips.
We studied 10 patients by means of pre- and postoperative CT who underwent computer-assisted total hip replacement using the CASPAR-system (OrthoMaquet GmbH, Rastatt, Germany). The anteversion angles of the stem and translational deviations measured after surgery were compared to the preoperatively planned ones as a quality control. We found an average difference of 7.8 degrees (SD 6.3 degrees; 95% CI 3.3 degrees-12.3 degrees) in the angles and an average difference of 1.8 (SD 1.7; 95% CI 0.6-3.0) mm and 1.2 (SD 1.4: 95% CI 0.25-2.2), respectively, in the medial and lateral deviation. In conclusion, we could not confirm the same accuracy of implant position as that claimed by the manufacturer.