{"title":"[注射泵作为执行器——反馈控制系统的实际应用问题]。","authors":"B Pohl,&nbsp;O Simanski,&nbsp;R Hofmockel","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Based on a computer simulation programme, the accuracy of the Graseby 3400 syringe pump was tested for its use in a feedback control system of the neuromuscular block. Firstly, a calculation of errors to determine the difference between the target and actual flow rates was carried out. Next, the characteristic curves of the syringe pump were determined under different flow and sampling rates to correct the application software of the feedback controller with a view to obtaining the correct flow rates online. It was discovered that, particularly with short 12 s sampling rates, dose-dependent errors of almost 100% were provable. Therefore, the application of a robust controller and integration of the characteristic curves at the outlet of the controller software are required. For the adoption of syringe pumps in medical feedback control systems, definite determination of the actual infusion quantity using an exact calculation of errors is required. Especially in cases of short sampling rates in combination with low infusion quantities, the syringe pump comes almost to a standstill, which results in extreme differences between target and actual flow rates.</p>","PeriodicalId":76993,"journal":{"name":"Anaesthesiologie und Reanimation","volume":"28 6","pages":"152-5"},"PeriodicalIF":0.0000,"publicationDate":"2003-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[The syringe pump as actuator--a practical application problem of feedback control systems].\",\"authors\":\"B Pohl,&nbsp;O Simanski,&nbsp;R Hofmockel\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Based on a computer simulation programme, the accuracy of the Graseby 3400 syringe pump was tested for its use in a feedback control system of the neuromuscular block. Firstly, a calculation of errors to determine the difference between the target and actual flow rates was carried out. Next, the characteristic curves of the syringe pump were determined under different flow and sampling rates to correct the application software of the feedback controller with a view to obtaining the correct flow rates online. It was discovered that, particularly with short 12 s sampling rates, dose-dependent errors of almost 100% were provable. Therefore, the application of a robust controller and integration of the characteristic curves at the outlet of the controller software are required. For the adoption of syringe pumps in medical feedback control systems, definite determination of the actual infusion quantity using an exact calculation of errors is required. Especially in cases of short sampling rates in combination with low infusion quantities, the syringe pump comes almost to a standstill, which results in extreme differences between target and actual flow rates.</p>\",\"PeriodicalId\":76993,\"journal\":{\"name\":\"Anaesthesiologie und Reanimation\",\"volume\":\"28 6\",\"pages\":\"152-5\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2003-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Anaesthesiologie und Reanimation\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Anaesthesiologie und Reanimation","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

基于计算机仿真程序,测试了grasby 3400注射泵在神经肌肉阻滞反馈控制系统中的准确性。首先进行误差计算,确定目标流量与实际流量的差值;其次,确定注射泵在不同流量和采样率下的特性曲线,对反馈控制器的应用软件进行校正,以期在线获得正确的流量。人们发现,特别是在短12秒的采样率下,几乎100%的剂量依赖性误差是可以证明的。因此,需要应用鲁棒控制器并在控制器软件出口集成特征曲线。为了在医疗反馈控制系统中采用注射泵,需要通过精确计算误差来确定实际注射量。特别是在采样率短且注射量低的情况下,注射泵几乎处于停顿状态,这导致目标流量与实际流量之间存在极大差异。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
[The syringe pump as actuator--a practical application problem of feedback control systems].

Based on a computer simulation programme, the accuracy of the Graseby 3400 syringe pump was tested for its use in a feedback control system of the neuromuscular block. Firstly, a calculation of errors to determine the difference between the target and actual flow rates was carried out. Next, the characteristic curves of the syringe pump were determined under different flow and sampling rates to correct the application software of the feedback controller with a view to obtaining the correct flow rates online. It was discovered that, particularly with short 12 s sampling rates, dose-dependent errors of almost 100% were provable. Therefore, the application of a robust controller and integration of the characteristic curves at the outlet of the controller software are required. For the adoption of syringe pumps in medical feedback control systems, definite determination of the actual infusion quantity using an exact calculation of errors is required. Especially in cases of short sampling rates in combination with low infusion quantities, the syringe pump comes almost to a standstill, which results in extreme differences between target and actual flow rates.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信