{"title":"子宫内Fincoid宫内节育器内铜线的腐蚀和重量损失。","authors":"S Kaivola,&nbsp;E Hirvonen","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>A total of 190 Fincoid IUDs were removed after varying periods of use in utero. Of 84 wires removed after 4 years of use, the 1st 12 wires were SEM photographed. In 1 of these wires, which was fragmented, a SEM photograph revealed deep corrosion pits. In the 11 unbroken wires, SEM photographs disclosed no clear pitting corrosion. Among 84 copper wires removed after 4 years of use, there were 4 fragmented ones. The risk of fragmentation after 4 years was thus 4.8%. Among the 106 copper wires removed for different reasons before the lapse of 4 years, there was not a single fragmented one. Annually, the mean daily release of copper ranged from 40.2 mcg to 43.3 mcg without any markedly higher initial release, 80% of the original weight still existing in the copper wires after 4 years in utero. The low incidence of pitting corrosion and fragmentation is most probably attributable to the higher positioning of the copper wire on the IUD. It is concluded that the effective life of the Fincoid IUD is more than 4 years.</p>","PeriodicalId":84493,"journal":{"name":"Contraceptive delivery systems","volume":"5 2","pages":"105-8"},"PeriodicalIF":0.0000,"publicationDate":"1984-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Corrosion and weight loss from copper wire in the Fincoid IUD in utero.\",\"authors\":\"S Kaivola,&nbsp;E Hirvonen\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>A total of 190 Fincoid IUDs were removed after varying periods of use in utero. Of 84 wires removed after 4 years of use, the 1st 12 wires were SEM photographed. In 1 of these wires, which was fragmented, a SEM photograph revealed deep corrosion pits. In the 11 unbroken wires, SEM photographs disclosed no clear pitting corrosion. Among 84 copper wires removed after 4 years of use, there were 4 fragmented ones. The risk of fragmentation after 4 years was thus 4.8%. Among the 106 copper wires removed for different reasons before the lapse of 4 years, there was not a single fragmented one. Annually, the mean daily release of copper ranged from 40.2 mcg to 43.3 mcg without any markedly higher initial release, 80% of the original weight still existing in the copper wires after 4 years in utero. The low incidence of pitting corrosion and fragmentation is most probably attributable to the higher positioning of the copper wire on the IUD. It is concluded that the effective life of the Fincoid IUD is more than 4 years.</p>\",\"PeriodicalId\":84493,\"journal\":{\"name\":\"Contraceptive delivery systems\",\"volume\":\"5 2\",\"pages\":\"105-8\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1984-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Contraceptive delivery systems\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Contraceptive delivery systems","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

在子宫内使用不同时间后,总共取出190个Fincoid宫内节育器。在使用4年后取出的84根电线中,对前12根电线进行了扫描电镜拍照。在其中一根断裂的电线上,扫描电镜照片显示出深深的腐蚀坑。在11根未断裂的电线中,扫描电镜照片显示没有明显的点蚀。使用4年拆除的84根铜线中,有4根断裂。4年后的碎片化风险为4.8%。4年前因各种原因拆除的106根铜线中,没有一根断裂的铜线。年平均日铜释放量在40.2 ~ 43.3 mcg之间,未见明显升高的初始释放量,4年后铜丝中仍有原来重量的80%。点蚀和碎裂发生率低,很可能是由于宫内节育器上铜线的位置较高。结论:Fincoid宫内节育器有效寿命在4年以上。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Corrosion and weight loss from copper wire in the Fincoid IUD in utero.

A total of 190 Fincoid IUDs were removed after varying periods of use in utero. Of 84 wires removed after 4 years of use, the 1st 12 wires were SEM photographed. In 1 of these wires, which was fragmented, a SEM photograph revealed deep corrosion pits. In the 11 unbroken wires, SEM photographs disclosed no clear pitting corrosion. Among 84 copper wires removed after 4 years of use, there were 4 fragmented ones. The risk of fragmentation after 4 years was thus 4.8%. Among the 106 copper wires removed for different reasons before the lapse of 4 years, there was not a single fragmented one. Annually, the mean daily release of copper ranged from 40.2 mcg to 43.3 mcg without any markedly higher initial release, 80% of the original weight still existing in the copper wires after 4 years in utero. The low incidence of pitting corrosion and fragmentation is most probably attributable to the higher positioning of the copper wire on the IUD. It is concluded that the effective life of the Fincoid IUD is more than 4 years.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信