{"title":"一种评估外科手套作为病毒屏障的全手套方法。","authors":"J R Nelson, T A Roming, J K Bennett","doi":"10.1053/AJCD01000183","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Today, because of the wide variety of infectious agents encountered in the health care environment, clinicians must be particularly concerned about the potential for small-sized virus penetration through glove defects.</p><p><strong>Objective: </strong>To describe a method for testing gloves that evaluates the entire glove and allows for detection of low levels of virus penetration. Ten sets of 10 different gloves from 4 manufacturers were evaluated using this method.</p><p><strong>Methods: </strong>Barrier properties were evaluated using the bacteriophage, phiX174. Gloves were filled with surfactant solution placed in flasks containing 10(6) viruses per mL. Flasks were agitated at 37 degrees C +/- 2 degrees C and assayed for 180 minutes.</p><p><strong>Results: </strong>Virus penetration was detected in 8% of the 100 gloves tested using the quantitative assay. The qualitative assay determined that 14% of the gloves tested allowed penetration.</p>","PeriodicalId":7612,"journal":{"name":"American journal of contact dermatitis : official journal of the American Contact Dermatitis Society","volume":"10 4","pages":"183-9"},"PeriodicalIF":0.0000,"publicationDate":"1999-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A whole-glove method for the evaluation of surgical gloves as barriers to viruses.\",\"authors\":\"J R Nelson, T A Roming, J K Bennett\",\"doi\":\"10.1053/AJCD01000183\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Today, because of the wide variety of infectious agents encountered in the health care environment, clinicians must be particularly concerned about the potential for small-sized virus penetration through glove defects.</p><p><strong>Objective: </strong>To describe a method for testing gloves that evaluates the entire glove and allows for detection of low levels of virus penetration. Ten sets of 10 different gloves from 4 manufacturers were evaluated using this method.</p><p><strong>Methods: </strong>Barrier properties were evaluated using the bacteriophage, phiX174. Gloves were filled with surfactant solution placed in flasks containing 10(6) viruses per mL. Flasks were agitated at 37 degrees C +/- 2 degrees C and assayed for 180 minutes.</p><p><strong>Results: </strong>Virus penetration was detected in 8% of the 100 gloves tested using the quantitative assay. The qualitative assay determined that 14% of the gloves tested allowed penetration.</p>\",\"PeriodicalId\":7612,\"journal\":{\"name\":\"American journal of contact dermatitis : official journal of the American Contact Dermatitis Society\",\"volume\":\"10 4\",\"pages\":\"183-9\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1999-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"American journal of contact dermatitis : official journal of the American Contact Dermatitis Society\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1053/AJCD01000183\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"American journal of contact dermatitis : official journal of the American Contact Dermatitis Society","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1053/AJCD01000183","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
A whole-glove method for the evaluation of surgical gloves as barriers to viruses.
Background: Today, because of the wide variety of infectious agents encountered in the health care environment, clinicians must be particularly concerned about the potential for small-sized virus penetration through glove defects.
Objective: To describe a method for testing gloves that evaluates the entire glove and allows for detection of low levels of virus penetration. Ten sets of 10 different gloves from 4 manufacturers were evaluated using this method.
Methods: Barrier properties were evaluated using the bacteriophage, phiX174. Gloves were filled with surfactant solution placed in flasks containing 10(6) viruses per mL. Flasks were agitated at 37 degrees C +/- 2 degrees C and assayed for 180 minutes.
Results: Virus penetration was detected in 8% of the 100 gloves tested using the quantitative assay. The qualitative assay determined that 14% of the gloves tested allowed penetration.