R I Rückert, P Romaniuk, P Rogalla, T Umscheid, W J Stelter, J M Müller
{"title":"一种调整错位分叉主动脉内移植物的方法。","authors":"R I Rückert, P Romaniuk, P Rogalla, T Umscheid, W J Stelter, J M Müller","doi":"10.1583/1074-6218(1998)005<0261:AMFAAM>2.0.CO;2","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To report the successful application of a method to adjust a malpositioned bifurcated stent-graft after endovascular aortic aneurysm repair.</p><p><strong>Method and results: </strong>A 62-year-old male patient underwent endovascular repair of a 5.1-cm abdominal aortic aneurysm (AAA) with a Vanguard bifurcated stent-graft. After complete deployment of the stent-graft, the intraoperative completion angiogram disclosed unexpected occlusion of the left renal artery. Intra-aortic adjustment of the bifurcated graft was possible with a crossover guidewire, which was pulled caudally. The method worked perfectly to restore blood flow to the left renal artery. The patient is well 16 months postoperatively without any evidence of endoleak or graft migration; the left renal artery remains open.</p><p><strong>Conclusions: </strong>A technique is demonstrated for intra-aortic repositioning of a bifurcated stentgraft to correct insufficient deployment. If required, this technique should be attempted before conversion to an open procedure.</p>","PeriodicalId":79443,"journal":{"name":"Journal of endovascular surgery : the official journal of the International Society for Endovascular Surgery","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"1998-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"5","resultStr":"{\"title\":\"A method for adjusting a malpositioned bifurcated aortic endograft.\",\"authors\":\"R I Rückert, P Romaniuk, P Rogalla, T Umscheid, W J Stelter, J M Müller\",\"doi\":\"10.1583/1074-6218(1998)005<0261:AMFAAM>2.0.CO;2\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To report the successful application of a method to adjust a malpositioned bifurcated stent-graft after endovascular aortic aneurysm repair.</p><p><strong>Method and results: </strong>A 62-year-old male patient underwent endovascular repair of a 5.1-cm abdominal aortic aneurysm (AAA) with a Vanguard bifurcated stent-graft. After complete deployment of the stent-graft, the intraoperative completion angiogram disclosed unexpected occlusion of the left renal artery. Intra-aortic adjustment of the bifurcated graft was possible with a crossover guidewire, which was pulled caudally. The method worked perfectly to restore blood flow to the left renal artery. The patient is well 16 months postoperatively without any evidence of endoleak or graft migration; the left renal artery remains open.</p><p><strong>Conclusions: </strong>A technique is demonstrated for intra-aortic repositioning of a bifurcated stentgraft to correct insufficient deployment. If required, this technique should be attempted before conversion to an open procedure.</p>\",\"PeriodicalId\":79443,\"journal\":{\"name\":\"Journal of endovascular surgery : the official journal of the International Society for Endovascular Surgery\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1998-08-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"5\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of endovascular surgery : the official journal of the International Society for Endovascular Surgery\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1583/1074-6218(1998)005<0261:AMFAAM>2.0.CO;2\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of endovascular surgery : the official journal of the International Society for Endovascular Surgery","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1583/1074-6218(1998)005<0261:AMFAAM>2.0.CO;2","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
A method for adjusting a malpositioned bifurcated aortic endograft.
Purpose: To report the successful application of a method to adjust a malpositioned bifurcated stent-graft after endovascular aortic aneurysm repair.
Method and results: A 62-year-old male patient underwent endovascular repair of a 5.1-cm abdominal aortic aneurysm (AAA) with a Vanguard bifurcated stent-graft. After complete deployment of the stent-graft, the intraoperative completion angiogram disclosed unexpected occlusion of the left renal artery. Intra-aortic adjustment of the bifurcated graft was possible with a crossover guidewire, which was pulled caudally. The method worked perfectly to restore blood flow to the left renal artery. The patient is well 16 months postoperatively without any evidence of endoleak or graft migration; the left renal artery remains open.
Conclusions: A technique is demonstrated for intra-aortic repositioning of a bifurcated stentgraft to correct insufficient deployment. If required, this technique should be attempted before conversion to an open procedure.