L Cipriani, G Baldereschi, L Boncinelli, N Marchionni
{"title":"主动脉内球囊破裂血管夹持:一例未手术成功切除。","authors":"L Cipriani, G Baldereschi, L Boncinelli, N Marchionni","doi":"10.1002/(sici)1097-0304(199806)44:2<218::aid-ccd19>3.0.co;2-i","DOIUrl":null,"url":null,"abstract":"<p><p>Intra-aortic balloon pump entrapment is a rare complication that may necessitate major abdominal surgery that is potentially life threatening in the critically ill patients who require balloon counterpulsation. We report successful removal of a ruptured and entrapped intra-aortic balloon pump catheter after use of streptokinase solution to clear clots from the device. We suggest this procedure as a safer, nonsurgical method that may eliminate the need for abdominal surgery.</p>","PeriodicalId":9664,"journal":{"name":"Catheterization and cardiovascular diagnosis","volume":"44 2","pages":"218-9"},"PeriodicalIF":0.0000,"publicationDate":"1998-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"9","resultStr":"{\"title\":\"Vascular entrapment of a ruptured intra-aortic balloon: a case report of successful removal without surgery.\",\"authors\":\"L Cipriani, G Baldereschi, L Boncinelli, N Marchionni\",\"doi\":\"10.1002/(sici)1097-0304(199806)44:2<218::aid-ccd19>3.0.co;2-i\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Intra-aortic balloon pump entrapment is a rare complication that may necessitate major abdominal surgery that is potentially life threatening in the critically ill patients who require balloon counterpulsation. We report successful removal of a ruptured and entrapped intra-aortic balloon pump catheter after use of streptokinase solution to clear clots from the device. We suggest this procedure as a safer, nonsurgical method that may eliminate the need for abdominal surgery.</p>\",\"PeriodicalId\":9664,\"journal\":{\"name\":\"Catheterization and cardiovascular diagnosis\",\"volume\":\"44 2\",\"pages\":\"218-9\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1998-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"9\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Catheterization and cardiovascular diagnosis\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1002/(sici)1097-0304(199806)44:2<218::aid-ccd19>3.0.co;2-i\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Catheterization and cardiovascular diagnosis","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1002/(sici)1097-0304(199806)44:2<218::aid-ccd19>3.0.co;2-i","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Vascular entrapment of a ruptured intra-aortic balloon: a case report of successful removal without surgery.
Intra-aortic balloon pump entrapment is a rare complication that may necessitate major abdominal surgery that is potentially life threatening in the critically ill patients who require balloon counterpulsation. We report successful removal of a ruptured and entrapped intra-aortic balloon pump catheter after use of streptokinase solution to clear clots from the device. We suggest this procedure as a safer, nonsurgical method that may eliminate the need for abdominal surgery.