{"title":"心脏手术中自动血压控制器的临床评价。","authors":"R H Meijers, D Schmartz, F R Cantraine, L Barvais, A A d'Hollander, J A Blom","doi":"10.1023/a:1007389631055","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>During surgery, computers can be of great use to support the anesthesiologist in providing task automation. In this paper we describe a closed loop blood pressure controller and show the results of its clinical evaluation.</p><p><strong>Methods: </strong>The controller is based on a simple and robust Proportional-Integral controller and a supervising, rule based, expert system. Adaptive control is necessary because the sensitivity of the patients to sodium nitroprusside varies over a wide range. Thirty-three clinical tests during cardiac surgery, including the cardiopulmonary bypass phase, were performed.</p><p><strong>Results: </strong>On average the controller was in automatic mode for 90.6 +/- 9.6% of the time. The performance during automatic control showed the mean arterial pressure to be within 10 mmHg of the setpoint for 71.4 +/- 15.5% of the time. The average absolute distance to the setpoint was 8.1 +/- 7.2 mmHg.</p><p><strong>Conclusions: </strong>The overall performance of the controller was noted as very satisfactory by the anesthesiologists.</p>","PeriodicalId":77199,"journal":{"name":"Journal of clinical monitoring","volume":"13 4","pages":"261-8"},"PeriodicalIF":0.0000,"publicationDate":"1997-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1023/a:1007389631055","citationCount":"10","resultStr":"{\"title\":\"Clinical evaluation of an automatic blood pressure controller during cardiac surgery.\",\"authors\":\"R H Meijers, D Schmartz, F R Cantraine, L Barvais, A A d'Hollander, J A Blom\",\"doi\":\"10.1023/a:1007389631055\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>During surgery, computers can be of great use to support the anesthesiologist in providing task automation. In this paper we describe a closed loop blood pressure controller and show the results of its clinical evaluation.</p><p><strong>Methods: </strong>The controller is based on a simple and robust Proportional-Integral controller and a supervising, rule based, expert system. Adaptive control is necessary because the sensitivity of the patients to sodium nitroprusside varies over a wide range. Thirty-three clinical tests during cardiac surgery, including the cardiopulmonary bypass phase, were performed.</p><p><strong>Results: </strong>On average the controller was in automatic mode for 90.6 +/- 9.6% of the time. The performance during automatic control showed the mean arterial pressure to be within 10 mmHg of the setpoint for 71.4 +/- 15.5% of the time. The average absolute distance to the setpoint was 8.1 +/- 7.2 mmHg.</p><p><strong>Conclusions: </strong>The overall performance of the controller was noted as very satisfactory by the anesthesiologists.</p>\",\"PeriodicalId\":77199,\"journal\":{\"name\":\"Journal of clinical monitoring\",\"volume\":\"13 4\",\"pages\":\"261-8\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1997-07-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1023/a:1007389631055\",\"citationCount\":\"10\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of clinical monitoring\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1023/a:1007389631055\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of clinical monitoring","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1023/a:1007389631055","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Clinical evaluation of an automatic blood pressure controller during cardiac surgery.
Objective: During surgery, computers can be of great use to support the anesthesiologist in providing task automation. In this paper we describe a closed loop blood pressure controller and show the results of its clinical evaluation.
Methods: The controller is based on a simple and robust Proportional-Integral controller and a supervising, rule based, expert system. Adaptive control is necessary because the sensitivity of the patients to sodium nitroprusside varies over a wide range. Thirty-three clinical tests during cardiac surgery, including the cardiopulmonary bypass phase, were performed.
Results: On average the controller was in automatic mode for 90.6 +/- 9.6% of the time. The performance during automatic control showed the mean arterial pressure to be within 10 mmHg of the setpoint for 71.4 +/- 15.5% of the time. The average absolute distance to the setpoint was 8.1 +/- 7.2 mmHg.
Conclusions: The overall performance of the controller was noted as very satisfactory by the anesthesiologists.