{"title":"一种新型外套导管系统的评价。","authors":"A J Gear,&nbsp;W D Nguyen,&nbsp;M A Bealle,&nbsp;R F Edlich","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The purpose of this study was to perform a biomechanical performance study and a clinical trial on a new external condom catheter system. The biomechanical performance study demonstrated that the vented connector failed following exposure to urine for 14 days. In addition, angulation of the inlet adapter of anti-reflux valve of the urinary leg bag to an angle of 35 degrees (+/-5 degrees) caused complete obstruction of urinary flow. A clinical trial of the new condom catheter with its inner flap revealed twisting of the catheter that interrupted urinary flow in four of ten patients. This catheter twisting was not observed in condom catheters without inner flaps. Furthermore, this new system is not easily applied or removed by patients with neurosensory deficits or musculoskeletal weakness. Design modifications of the vent within the connection tubing, the anti-reflux valve in the collection bag, the condom catheter with its inner flap, as well as the leg bag attachments should remarkably improve the performance of this new urinary drainage system.</p>","PeriodicalId":76136,"journal":{"name":"Medical progress through technology","volume":"21 4","pages":"181-5"},"PeriodicalIF":0.0000,"publicationDate":"1996-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Evaluation of a new external condom catheter system.\",\"authors\":\"A J Gear,&nbsp;W D Nguyen,&nbsp;M A Bealle,&nbsp;R F Edlich\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The purpose of this study was to perform a biomechanical performance study and a clinical trial on a new external condom catheter system. The biomechanical performance study demonstrated that the vented connector failed following exposure to urine for 14 days. In addition, angulation of the inlet adapter of anti-reflux valve of the urinary leg bag to an angle of 35 degrees (+/-5 degrees) caused complete obstruction of urinary flow. A clinical trial of the new condom catheter with its inner flap revealed twisting of the catheter that interrupted urinary flow in four of ten patients. This catheter twisting was not observed in condom catheters without inner flaps. Furthermore, this new system is not easily applied or removed by patients with neurosensory deficits or musculoskeletal weakness. Design modifications of the vent within the connection tubing, the anti-reflux valve in the collection bag, the condom catheter with its inner flap, as well as the leg bag attachments should remarkably improve the performance of this new urinary drainage system.</p>\",\"PeriodicalId\":76136,\"journal\":{\"name\":\"Medical progress through technology\",\"volume\":\"21 4\",\"pages\":\"181-5\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1996-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Medical progress through technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical progress through technology","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

本研究的目的是对一种新型外套导管系统进行生物力学性能研究和临床试验。生物力学性能研究表明,通气接头暴露于尿液中14天后失效。另外,尿腿袋防反流阀进口接头成角度为35度(+/-5度),造成尿流完全阻塞。一项对带有内瓣的新型避孕套导尿管的临床试验显示,在10名患者中,有4名患者的导尿管扭曲导致尿流中断。在没有内瓣的避孕套导管中没有观察到这种导管扭转。此外,这种新系统不容易被神经感觉缺陷或肌肉骨骼无力的患者应用或移除。对连接管内的通风口、收集袋内的抗反流阀、带内瓣的避孕套导管以及腿袋附件的设计修改应能显著改善这种新型泌尿引流系统的性能。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Evaluation of a new external condom catheter system.

The purpose of this study was to perform a biomechanical performance study and a clinical trial on a new external condom catheter system. The biomechanical performance study demonstrated that the vented connector failed following exposure to urine for 14 days. In addition, angulation of the inlet adapter of anti-reflux valve of the urinary leg bag to an angle of 35 degrees (+/-5 degrees) caused complete obstruction of urinary flow. A clinical trial of the new condom catheter with its inner flap revealed twisting of the catheter that interrupted urinary flow in four of ten patients. This catheter twisting was not observed in condom catheters without inner flaps. Furthermore, this new system is not easily applied or removed by patients with neurosensory deficits or musculoskeletal weakness. Design modifications of the vent within the connection tubing, the anti-reflux valve in the collection bag, the condom catheter with its inner flap, as well as the leg bag attachments should remarkably improve the performance of this new urinary drainage system.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信