{"title":"颅压遥测系统。一、硬件开发。","authors":"W H Ko, A Leung, E Cheng, R J Lorig","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Results in the development of intracranial pressure(ICP) telemetry systems are reported. Included are a single-channel ICP system and a two-channel ICP and temperature system. The unit used hybrid-integrated circuits housed in a Kovar flatpack and packaged with polyurethane or Hysol epoxy. Battery and radio frequency (RF) induction power supplies were tested. The two-channel systems used RF power at 3.5 MHz and transmitted signals around 120 MHz. The package measures 3 x 2 x 0.8 cm and weighs 9 g. The pressure range is -20 to +100 mm Hg with accuracy to 1.0 mm Hg. The implant measures absolute pressure and has a baseline stability of better than +/- 2 mm Hg/month. 17 dogs and 4 goats were used for in vivo evaluation. A summary of results is presented. Detailed evaluation is given in section II of this paper.</p>","PeriodicalId":75603,"journal":{"name":"Biotelemetry and patient monitoring","volume":"8 3","pages":"131-50"},"PeriodicalIF":0.0000,"publicationDate":"1981-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Intracranial pressure telemetry system. I. Hardware development.\",\"authors\":\"W H Ko, A Leung, E Cheng, R J Lorig\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Results in the development of intracranial pressure(ICP) telemetry systems are reported. Included are a single-channel ICP system and a two-channel ICP and temperature system. The unit used hybrid-integrated circuits housed in a Kovar flatpack and packaged with polyurethane or Hysol epoxy. Battery and radio frequency (RF) induction power supplies were tested. The two-channel systems used RF power at 3.5 MHz and transmitted signals around 120 MHz. The package measures 3 x 2 x 0.8 cm and weighs 9 g. The pressure range is -20 to +100 mm Hg with accuracy to 1.0 mm Hg. The implant measures absolute pressure and has a baseline stability of better than +/- 2 mm Hg/month. 17 dogs and 4 goats were used for in vivo evaluation. A summary of results is presented. Detailed evaluation is given in section II of this paper.</p>\",\"PeriodicalId\":75603,\"journal\":{\"name\":\"Biotelemetry and patient monitoring\",\"volume\":\"8 3\",\"pages\":\"131-50\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1981-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Biotelemetry and patient monitoring\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biotelemetry and patient monitoring","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Intracranial pressure telemetry system. I. Hardware development.
Results in the development of intracranial pressure(ICP) telemetry systems are reported. Included are a single-channel ICP system and a two-channel ICP and temperature system. The unit used hybrid-integrated circuits housed in a Kovar flatpack and packaged with polyurethane or Hysol epoxy. Battery and radio frequency (RF) induction power supplies were tested. The two-channel systems used RF power at 3.5 MHz and transmitted signals around 120 MHz. The package measures 3 x 2 x 0.8 cm and weighs 9 g. The pressure range is -20 to +100 mm Hg with accuracy to 1.0 mm Hg. The implant measures absolute pressure and has a baseline stability of better than +/- 2 mm Hg/month. 17 dogs and 4 goats were used for in vivo evaluation. A summary of results is presented. Detailed evaluation is given in section II of this paper.