{"title":"动脉血压动态监测。在实验室条件下和日常活动中,用Remler M 2000便携式记录仪和无线电遥测获得的血压测量值的比较。","authors":"K Bachmann, G Bäuerlein","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Simultaneous indirect and direct arterial blood pressure measurements were performed with a Remler M 2000 (RM) portable recorder and by radiotelemetry (RT) in 21 subjects under various conditions such as rest in the supine position and in an 80 degree upright position on a tilting table, during bicycle ergometry, physical exercise on the terrain and under the emotional stress of driving a car. There is a good correlation between semiautomatic pressure measurements carried out with the RM portable recorder and radiotelemetric data. The differences observed are basically due to the auscultation method. But the RM has an advantage over the auscultation method in that it objectively registers the Korotkov sounds and records them on a tape, not only under laboratory conditions but also during everyday activities.</p>","PeriodicalId":75603,"journal":{"name":"Biotelemetry and patient monitoring","volume":"8 1-2","pages":"47-55"},"PeriodicalIF":0.0000,"publicationDate":"1981-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Ambulatory monitoring of arterial blood pressure. Comparison between blood pressure measurements obtained with the Remler M 2000 portable recorder and by radiotelemetry under laboratory conditions and during everyday activities.\",\"authors\":\"K Bachmann, G Bäuerlein\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Simultaneous indirect and direct arterial blood pressure measurements were performed with a Remler M 2000 (RM) portable recorder and by radiotelemetry (RT) in 21 subjects under various conditions such as rest in the supine position and in an 80 degree upright position on a tilting table, during bicycle ergometry, physical exercise on the terrain and under the emotional stress of driving a car. There is a good correlation between semiautomatic pressure measurements carried out with the RM portable recorder and radiotelemetric data. The differences observed are basically due to the auscultation method. But the RM has an advantage over the auscultation method in that it objectively registers the Korotkov sounds and records them on a tape, not only under laboratory conditions but also during everyday activities.</p>\",\"PeriodicalId\":75603,\"journal\":{\"name\":\"Biotelemetry and patient monitoring\",\"volume\":\"8 1-2\",\"pages\":\"47-55\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1981-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Biotelemetry and patient monitoring\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biotelemetry and patient monitoring","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
用Remler M 2000 (RM)便携式记录仪和无线电遥测(RT)同时测量了21名受试者在不同条件下的间接和直接动脉血压,如仰卧位和倾斜桌上80度直立位置休息,自行车几何测量,在地形上进行体育锻炼和驾驶汽车时的情绪压力。用RM便携式记录仪进行的半自动压力测量与无线电遥测数据之间有很好的相关性。所观察到的差异基本上是听诊方法造成的。但是RM比听诊法有一个优势,因为它客观地记录了科罗特科夫的声音,并将它们记录在磁带上,不仅在实验室条件下,而且在日常活动中。
Ambulatory monitoring of arterial blood pressure. Comparison between blood pressure measurements obtained with the Remler M 2000 portable recorder and by radiotelemetry under laboratory conditions and during everyday activities.
Simultaneous indirect and direct arterial blood pressure measurements were performed with a Remler M 2000 (RM) portable recorder and by radiotelemetry (RT) in 21 subjects under various conditions such as rest in the supine position and in an 80 degree upright position on a tilting table, during bicycle ergometry, physical exercise on the terrain and under the emotional stress of driving a car. There is a good correlation between semiautomatic pressure measurements carried out with the RM portable recorder and radiotelemetric data. The differences observed are basically due to the auscultation method. But the RM has an advantage over the auscultation method in that it objectively registers the Korotkov sounds and records them on a tape, not only under laboratory conditions but also during everyday activities.