{"title":"心室除颤的有效性取决于极性吗?","authors":"J C Schuder, H Stoeckle, W C McDaniel, M Dbeis","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Studies involving 240 fibrillation-defibrillation episodes via epicardial patch electrodes in 21-27-kg dogs and 480 transthoracic episodes in 100-kg calves are reported. In dogs, 120 episodes involving shocks by a 3.7-A, 5-ms unidirectional rectangular wave of one polarity were interlaced with 120 similar episodes of the reverse polarity. When the upper right ventricular patch was positive with respect to the left ventricular apex patch, 85% of the episodes yielded defibrillation; 78% defibrillated with the reverse polarity. In one series in calves, 120 episodes involving shocks by a 42-A, 4-ms unidirectional rectangular wave of one polarity were interlaced with 120 similar episodes involving the reverse polarity. In a second series, 59-A, 4-ms shocks were employed. In the 42-A series, 38% of the episodes were successful when the upper right electrode was positive with respect to the electrode over the apex, and 28% were successful with the reverse polarity. Corresponding results in the 59-A series were 80% and 68%, respectively. On an unpaired basis, the three p values were 0.19, 0.14, and 0.04, respectively. On a paired basis, all three comparisons yielded significant differences (p less than 0.05). We conclude that in these cases, at least, there is a moderate dependence upon electrode polarity and that our results appear to warrant clinical studies.</p>","PeriodicalId":76133,"journal":{"name":"Medical instrumentation","volume":"21 5","pages":"262-5"},"PeriodicalIF":0.0000,"publicationDate":"1987-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Is the effectiveness of cardiac ventricular defibrillation dependent upon polarity?\",\"authors\":\"J C Schuder, H Stoeckle, W C McDaniel, M Dbeis\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Studies involving 240 fibrillation-defibrillation episodes via epicardial patch electrodes in 21-27-kg dogs and 480 transthoracic episodes in 100-kg calves are reported. In dogs, 120 episodes involving shocks by a 3.7-A, 5-ms unidirectional rectangular wave of one polarity were interlaced with 120 similar episodes of the reverse polarity. When the upper right ventricular patch was positive with respect to the left ventricular apex patch, 85% of the episodes yielded defibrillation; 78% defibrillated with the reverse polarity. In one series in calves, 120 episodes involving shocks by a 42-A, 4-ms unidirectional rectangular wave of one polarity were interlaced with 120 similar episodes involving the reverse polarity. In a second series, 59-A, 4-ms shocks were employed. In the 42-A series, 38% of the episodes were successful when the upper right electrode was positive with respect to the electrode over the apex, and 28% were successful with the reverse polarity. Corresponding results in the 59-A series were 80% and 68%, respectively. On an unpaired basis, the three p values were 0.19, 0.14, and 0.04, respectively. On a paired basis, all three comparisons yielded significant differences (p less than 0.05). We conclude that in these cases, at least, there is a moderate dependence upon electrode polarity and that our results appear to warrant clinical studies.</p>\",\"PeriodicalId\":76133,\"journal\":{\"name\":\"Medical instrumentation\",\"volume\":\"21 5\",\"pages\":\"262-5\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1987-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Medical instrumentation\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical instrumentation","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
有研究报道,在体重21-27公斤的犬中通过心外膜贴片电极进行240次纤颤-除颤,在体重100公斤的小牛中进行480次经胸纤颤。在狗身上,120次3.7 a、5毫秒单向矩形波的单极性电击与120次类似的反极性电击交织在一起。当右上心室膜片相对于左心室心尖膜片呈阳性时,85%的发作产生除颤;78%的患者极性相反。在小牛的一个系列中,120个片段涉及42 a, 4 ms单向矩形波的单极性电击,与120个类似的反极性片段交错。在第二个系列中,使用了59-A, 4毫秒的冲击。在42-A系列中,当右上电极相对于顶端电极为正时,38%的发作成功,28%的发作成功。59-A系列的对应结果分别为80%和68%。在未配对的基础上,三个p值分别为0.19、0.14和0.04。在配对基础上,三个比较均产生显著差异(p < 0.05)。我们的结论是,在这些情况下,至少有一个适度的依赖于电极极性,我们的结果似乎保证临床研究。
Is the effectiveness of cardiac ventricular defibrillation dependent upon polarity?
Studies involving 240 fibrillation-defibrillation episodes via epicardial patch electrodes in 21-27-kg dogs and 480 transthoracic episodes in 100-kg calves are reported. In dogs, 120 episodes involving shocks by a 3.7-A, 5-ms unidirectional rectangular wave of one polarity were interlaced with 120 similar episodes of the reverse polarity. When the upper right ventricular patch was positive with respect to the left ventricular apex patch, 85% of the episodes yielded defibrillation; 78% defibrillated with the reverse polarity. In one series in calves, 120 episodes involving shocks by a 42-A, 4-ms unidirectional rectangular wave of one polarity were interlaced with 120 similar episodes involving the reverse polarity. In a second series, 59-A, 4-ms shocks were employed. In the 42-A series, 38% of the episodes were successful when the upper right electrode was positive with respect to the electrode over the apex, and 28% were successful with the reverse polarity. Corresponding results in the 59-A series were 80% and 68%, respectively. On an unpaired basis, the three p values were 0.19, 0.14, and 0.04, respectively. On a paired basis, all three comparisons yielded significant differences (p less than 0.05). We conclude that in these cases, at least, there is a moderate dependence upon electrode polarity and that our results appear to warrant clinical studies.