{"title":"钛合金颈部断裂引起的模块化髋柄机械故障:单中心案例系列。","authors":"N Papapietro, B Zampogna, C Caria, S Giusti, A Marinozzi, A Di Martino, R Papalia, V Denaro","doi":"10.1007/s12306-025-00935-4","DOIUrl":null,"url":null,"abstract":"<p><strong>Backround: </strong>Modular neck implants in total hip arthroplasty (THA) allow control of length, offset, and version. However, these implants showed problems such as implant fracture and adverse reactions due to metal debris. This study aims to analyze our modular titanium-titanium neck THA stems, focus on mechanical complications related to its breakage, conduct a metallurgical analysis on these broken implants, and report revision surgery outcomes.</p><p><strong>Methods: </strong>We performed a retrospective review to identify patients who had undergone THA with a specific modular (H2M-C2F Implants) implant at our institute (Fondazione Policlinico Universitario Campus Bio-Medico, Rome). Patient demographics, implant characteristics, clinical follow-up, and revision outcomes were obtained. A metallurgical examination was conducted on the broken implants.</p><p><strong>Results: </strong>Out of thirty-eight patients eligible patients, seventeen had a titanium alloy neck, whereas the remaining twenty-one had CrCo necks. Thirteen of seventeen patients with titanium necks reported spontaneous ruptures of the neck due to fretting corrosion phenomena. Revision surgery was offered to all patients, including the ones who did not experience spontaneous rupture (four patients). Neither intraoperative nor long-term complications were reported, except for one patient who showed gluteal insufficiency. No clinical differences were reported between the patients who underwent revision surgery and those who were revised preventively.</p><p><strong>Conclusion: </strong>Using modular implants should be limited to strict indications. Patients' characteristics are vital in selecting the ideal candidate for a modular THA. Modular implants should answer all the questions regarding durability and fracture risk to become a valid alternative to fixed-neck stems.</p>","PeriodicalId":18875,"journal":{"name":"MUSCULOSKELETAL SURGERY","volume":" ","pages":"335-342"},"PeriodicalIF":0.0000,"publicationDate":"2026-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Mechanical failure of modular hip stems due to titanium alloy neck breakage: a single-center case series.\",\"authors\":\"N Papapietro, B Zampogna, C Caria, S Giusti, A Marinozzi, A Di Martino, R Papalia, V Denaro\",\"doi\":\"10.1007/s12306-025-00935-4\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Backround: </strong>Modular neck implants in total hip arthroplasty (THA) allow control of length, offset, and version. However, these implants showed problems such as implant fracture and adverse reactions due to metal debris. This study aims to analyze our modular titanium-titanium neck THA stems, focus on mechanical complications related to its breakage, conduct a metallurgical analysis on these broken implants, and report revision surgery outcomes.</p><p><strong>Methods: </strong>We performed a retrospective review to identify patients who had undergone THA with a specific modular (H2M-C2F Implants) implant at our institute (Fondazione Policlinico Universitario Campus Bio-Medico, Rome). Patient demographics, implant characteristics, clinical follow-up, and revision outcomes were obtained. A metallurgical examination was conducted on the broken implants.</p><p><strong>Results: </strong>Out of thirty-eight patients eligible patients, seventeen had a titanium alloy neck, whereas the remaining twenty-one had CrCo necks. Thirteen of seventeen patients with titanium necks reported spontaneous ruptures of the neck due to fretting corrosion phenomena. Revision surgery was offered to all patients, including the ones who did not experience spontaneous rupture (four patients). Neither intraoperative nor long-term complications were reported, except for one patient who showed gluteal insufficiency. No clinical differences were reported between the patients who underwent revision surgery and those who were revised preventively.</p><p><strong>Conclusion: </strong>Using modular implants should be limited to strict indications. Patients' characteristics are vital in selecting the ideal candidate for a modular THA. Modular implants should answer all the questions regarding durability and fracture risk to become a valid alternative to fixed-neck stems.</p>\",\"PeriodicalId\":18875,\"journal\":{\"name\":\"MUSCULOSKELETAL SURGERY\",\"volume\":\" \",\"pages\":\"335-342\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2026-09-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"MUSCULOSKELETAL SURGERY\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1007/s12306-025-00935-4\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2025/11/9 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q1\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"MUSCULOSKELETAL SURGERY","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s12306-025-00935-4","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/11/9 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

背景:全髋关节置换术(THA)中的模块化颈部植入物可以控制长度、偏移量和版本。然而,这些种植体存在诸如种植体骨折和金属碎片引起的不良反应等问题。本研究旨在分析我们的模块化钛-钛颈部THA干,重点关注其断裂相关的机械并发症,对这些断裂的植入物进行冶金分析,并报告翻修手术结果。方法:我们进行了一项回顾性研究,以确定在我们研究所(Fondazione Policlinico Universitario Campus Bio-Medico,罗马)接受了特定模块(H2M-C2F implant)种植体THA的患者。获得了患者人口统计、种植体特征、临床随访和翻修结果。对破损的种植体进行了金相检查。结果:在38例符合条件的患者中,17例采用钛合金颈部,其余21例采用CrCo颈部。17例钛颈患者中有13例报告由于微动腐蚀现象导致颈部自发破裂。所有患者均接受翻修手术,包括未发生自发性破裂的患者(4例)。除1例患者出现臀肌功能不全外,术中及长期并发症均未见报道。在接受翻修手术的患者和预防性翻修的患者之间没有临床差异。结论:使用模块化种植体应严格限制适应症。患者的特点是至关重要的选择理想的候选模块THA。模块化植入物应该回答所有关于耐久性和骨折风险的问题,从而成为固定颈假体的有效替代品。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Mechanical failure of modular hip stems due to titanium alloy neck breakage: a single-center case series.

Backround: Modular neck implants in total hip arthroplasty (THA) allow control of length, offset, and version. However, these implants showed problems such as implant fracture and adverse reactions due to metal debris. This study aims to analyze our modular titanium-titanium neck THA stems, focus on mechanical complications related to its breakage, conduct a metallurgical analysis on these broken implants, and report revision surgery outcomes.

Methods: We performed a retrospective review to identify patients who had undergone THA with a specific modular (H2M-C2F Implants) implant at our institute (Fondazione Policlinico Universitario Campus Bio-Medico, Rome). Patient demographics, implant characteristics, clinical follow-up, and revision outcomes were obtained. A metallurgical examination was conducted on the broken implants.

Results: Out of thirty-eight patients eligible patients, seventeen had a titanium alloy neck, whereas the remaining twenty-one had CrCo necks. Thirteen of seventeen patients with titanium necks reported spontaneous ruptures of the neck due to fretting corrosion phenomena. Revision surgery was offered to all patients, including the ones who did not experience spontaneous rupture (four patients). Neither intraoperative nor long-term complications were reported, except for one patient who showed gluteal insufficiency. No clinical differences were reported between the patients who underwent revision surgery and those who were revised preventively.

Conclusion: Using modular implants should be limited to strict indications. Patients' characteristics are vital in selecting the ideal candidate for a modular THA. Modular implants should answer all the questions regarding durability and fracture risk to become a valid alternative to fixed-neck stems.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
MUSCULOSKELETAL SURGERY
MUSCULOSKELETAL SURGERY Medicine-Surgery
CiteScore
4.50
自引率
0.00%
发文量
35
期刊介绍: Musculoskeletal Surgery – Formerly La Chirurgia degli Organi di Movimento, founded in 1917 at the Istituto Ortopedico Rizzoli, is a peer-reviewed journal published three times a year. The journal provides up-to-date information to clinicians and scientists through the publication of original papers, reviews, case reports, and brief communications dealing with the pathogenesis and treatment of orthopaedic conditions.An electronic version is also available at http://www.springerlink.com.The journal is open for publication of supplements and for publishing abstracts of scientific meetings; conditions can be obtained from the Editors-in-Chief or the Publisher.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书