{"title":"用塑料闪烁体评估巨压光子光束质量保证的剂量学参数","authors":"Gregory Penoncello, Bernard L. Jones, David A. P. Dunkerley, Adam Mahl, Moyed Miften, Cem Altunbas, C-K Chris Wang, Daniel G. Robertson","doi":"10.1002/acm2.70276","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Introduction</h3>\n \n <p>Linear accelerators require a large amount of data to be collected daily, monthly and annually to verify safe deliveries for patients. Different detectors have been utilized to improve the simplicity, efficiency and accuracy of the various experimental setups required to collect the necessary data resulting in reduced data collection and evaluation time. Plastic scintillators are stable and energy independent radiation detectors with high spatial resolution that emit light proportional to the amount of radiation incident on them. With an appropriate photodetector and technique to measure and analyze the light that is emitted, scintillators can be utilized to measure dosimetric parameters necessary for various monthly quality assurance requirements.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>A uniform cylindrical plastic scintillator imaged by a complementary metal oxide semiconductor (CMOS) camera is designed to act as a radiation detector to measure 2D projections of linear accelerator beams. Beams were delivered to the detector to evaluate machine quality assurance (QA) parameters, including beam energy, output, profile consistency and reproducibility as described by TG-198. These measurements were compared to calculations from the Eclipse treatment planning system (TPS). Typical monthly quality assurance (QA) beams were delivered and 2D projections of the scintillation light were measured to validate the accuracy and reproducibility of this detector system for monthly QA.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The plastic scintillator was able to accurately characterize the radiation beam. Energy and profile measurements were reproducible and within 2%/2mm of calculations in the TPS. Output measurements had maximum variations of up to 1.3% and average differences of 0.5%.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>A simple cylindrical plastic scintillator and CMOS camera radiation detector setup was designed and tested for measuring monthly QA dosimetric parameters specified by TG-198 with accurate and reproducible output, energy and profiles measurements. This method reduces the number of measurements required, allowing multiple parameters to be evaluated in a single beam delivery.</p>\n </section>\n </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"26 10","pages":""},"PeriodicalIF":2.2000,"publicationDate":"2025-10-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://aapm.onlinelibrary.wiley.com/doi/epdf/10.1002/acm2.70276","citationCount":"0","resultStr":"{\"title\":\"Evaluating dosimetric parameters with a plastic scintillator for megavoltage photon beam quality assurance\",\"authors\":\"Gregory Penoncello, Bernard L. Jones, David A. P. Dunkerley, Adam Mahl, Moyed Miften, Cem Altunbas, C-K Chris Wang, Daniel G. Robertson\",\"doi\":\"10.1002/acm2.70276\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Introduction</h3>\\n \\n <p>Linear accelerators require a large amount of data to be collected daily, monthly and annually to verify safe deliveries for patients. Different detectors have been utilized to improve the simplicity, efficiency and accuracy of the various experimental setups required to collect the necessary data resulting in reduced data collection and evaluation time. Plastic scintillators are stable and energy independent radiation detectors with high spatial resolution that emit light proportional to the amount of radiation incident on them. With an appropriate photodetector and technique to measure and analyze the light that is emitted, scintillators can be utilized to measure dosimetric parameters necessary for various monthly quality assurance requirements.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Methods</h3>\\n \\n <p>A uniform cylindrical plastic scintillator imaged by a complementary metal oxide semiconductor (CMOS) camera is designed to act as a radiation detector to measure 2D projections of linear accelerator beams. Beams were delivered to the detector to evaluate machine quality assurance (QA) parameters, including beam energy, output, profile consistency and reproducibility as described by TG-198. These measurements were compared to calculations from the Eclipse treatment planning system (TPS). Typical monthly quality assurance (QA) beams were delivered and 2D projections of the scintillation light were measured to validate the accuracy and reproducibility of this detector system for monthly QA.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>The plastic scintillator was able to accurately characterize the radiation beam. Energy and profile measurements were reproducible and within 2%/2mm of calculations in the TPS. Output measurements had maximum variations of up to 1.3% and average differences of 0.5%.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusion</h3>\\n \\n <p>A simple cylindrical plastic scintillator and CMOS camera radiation detector setup was designed and tested for measuring monthly QA dosimetric parameters specified by TG-198 with accurate and reproducible output, energy and profiles measurements. This method reduces the number of measurements required, allowing multiple parameters to be evaluated in a single beam delivery.</p>\\n </section>\\n </div>\",\"PeriodicalId\":14989,\"journal\":{\"name\":\"Journal of Applied Clinical Medical Physics\",\"volume\":\"26 10\",\"pages\":\"\"},\"PeriodicalIF\":2.2000,\"publicationDate\":\"2025-10-10\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://aapm.onlinelibrary.wiley.com/doi/epdf/10.1002/acm2.70276\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Applied Clinical Medical Physics\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://aapm.onlinelibrary.wiley.com/doi/10.1002/acm2.70276\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://aapm.onlinelibrary.wiley.com/doi/10.1002/acm2.70276","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
Evaluating dosimetric parameters with a plastic scintillator for megavoltage photon beam quality assurance
Introduction
Linear accelerators require a large amount of data to be collected daily, monthly and annually to verify safe deliveries for patients. Different detectors have been utilized to improve the simplicity, efficiency and accuracy of the various experimental setups required to collect the necessary data resulting in reduced data collection and evaluation time. Plastic scintillators are stable and energy independent radiation detectors with high spatial resolution that emit light proportional to the amount of radiation incident on them. With an appropriate photodetector and technique to measure and analyze the light that is emitted, scintillators can be utilized to measure dosimetric parameters necessary for various monthly quality assurance requirements.
Methods
A uniform cylindrical plastic scintillator imaged by a complementary metal oxide semiconductor (CMOS) camera is designed to act as a radiation detector to measure 2D projections of linear accelerator beams. Beams were delivered to the detector to evaluate machine quality assurance (QA) parameters, including beam energy, output, profile consistency and reproducibility as described by TG-198. These measurements were compared to calculations from the Eclipse treatment planning system (TPS). Typical monthly quality assurance (QA) beams were delivered and 2D projections of the scintillation light were measured to validate the accuracy and reproducibility of this detector system for monthly QA.
Results
The plastic scintillator was able to accurately characterize the radiation beam. Energy and profile measurements were reproducible and within 2%/2mm of calculations in the TPS. Output measurements had maximum variations of up to 1.3% and average differences of 0.5%.
Conclusion
A simple cylindrical plastic scintillator and CMOS camera radiation detector setup was designed and tested for measuring monthly QA dosimetric parameters specified by TG-198 with accurate and reproducible output, energy and profiles measurements. This method reduces the number of measurements required, allowing multiple parameters to be evaluated in a single beam delivery.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic