{"title":"Maglumi免疫测定抗gad65在糖尿病和神经系统疾病中的临床意义:分析挑战和诊断见解","authors":"Gonzalo Verdú ,&nbsp;Antonio Jesús Blanco-Carrasco ,&nbsp;Raquel Ruiz-García ,&nbsp;Albert Saiz ,&nbsp;Manuel Morales-Ruiz ,&nbsp;Gregori Casals","doi":"10.1016/j.clinbiochem.2025.111024","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><div>Anti-GAD65 is an established diagnostic biomarker for autoimmune diabetes and neurological syndromes associated with GAD antibody-spectrum disorders. Automated chemiluminescence-based immunoassays such as Maglumi are increasingly used, but their behavior at high antibody concentrations remains poorly characterized. We report the analytical concordance between Maglumi and radioimmunoassay (RIA), the cut-off titer that differentiates neurological from diabetic patients, and the hook effect of the immunoassay.</div></div><div><h3>Materials and methods</h3><div>Anti-GAD65 levels were analyzed in samples from 90 patients with suspected autoimmune diabetes or neurological syndromes, using Maglumi 2000® and RIA (Medipan®). Analytical concordance was assessed by Passing–Bablok regression, Bland–Altman plots, and Cohen’s kappa. ROC curve analysis identified diagnostic thresholds, and dilution studies were performed in samples with unexpectedly low values to assess a potential hook effect.</div></div><div><h3>Results</h3><div>Maglumi results were approximately five times higher than RIA results (slope = 5.364; R<sup>2</sup> = 0.908); agreement was higher in endocrine (R<sup>2</sup> = 0.998) than in neurological patients (R<sup>2</sup> = 0.891). A cutoff of 6778 IU/mL discriminated neurological patients from diabetic patients (AUC = 0.96). Sixteen samples showed a hook effect, with direct (undiluted) values ≤273 IU/mL, while the corrected titers after dilution exceeded 20,000 IU/mL. A linear regression confirmed an inverse relationship between direct and true titers (r = –0.79, p &lt; 0.05).</div></div><div><h3>Discussion</h3><div>Maglumi anti-GAD65 immunoassay provides reliable discrimination between endocrine and neurological patients but requires method-specific cutoffs to avoid misclassification. Hook effect was frequent in high-titer neurological samples, potentially leading to misdiagnosis, underscoring the need for routine dilution protocols to improve diagnostic reliability.</div></div><div><h3>Conclusions</h3><div>Standardized dilution protocols and appropriate cutoffs are essential to ensure reliable anti-GAD65 testing with Maglumi chemiluminescent immunoassay, particularly in neurological indications.</div></div>","PeriodicalId":10172,"journal":{"name":"Clinical biochemistry","volume":"140 ","pages":"Article 111024"},"PeriodicalIF":2.1000,"publicationDate":"2025-10-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Clinical relevance of the Maglumi immunoassay for anti-GAD65 in diabetes and neurological disorders: Analytical challenges and diagnostic insights\",\"authors\":\"Gonzalo Verdú ,&nbsp;Antonio Jesús Blanco-Carrasco ,&nbsp;Raquel Ruiz-García ,&nbsp;Albert Saiz ,&nbsp;Manuel Morales-Ruiz ,&nbsp;Gregori Casals\",\"doi\":\"10.1016/j.clinbiochem.2025.111024\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Introduction</h3><div>Anti-GAD65 is an established diagnostic biomarker for autoimmune diabetes and neurological syndromes associated with GAD antibody-spectrum disorders. Automated chemiluminescence-based immunoassays such as Maglumi are increasingly used, but their behavior at high antibody concentrations remains poorly characterized. We report the analytical concordance between Maglumi and radioimmunoassay (RIA), the cut-off titer that differentiates neurological from diabetic patients, and the hook effect of the immunoassay.</div></div><div><h3>Materials and methods</h3><div>Anti-GAD65 levels were analyzed in samples from 90 patients with suspected autoimmune diabetes or neurological syndromes, using Maglumi 2000® and RIA (Medipan®). Analytical concordance was assessed by Passing–Bablok regression, Bland–Altman plots, and Cohen’s kappa. ROC curve analysis identified diagnostic thresholds, and dilution studies were performed in samples with unexpectedly low values to assess a potential hook effect.</div></div><div><h3>Results</h3><div>Maglumi results were approximately five times higher than RIA results (slope = 5.364; R<sup>2</sup> = 0.908); agreement was higher in endocrine (R<sup>2</sup> = 0.998) than in neurological patients (R<sup>2</sup> = 0.891). A cutoff of 6778 IU/mL discriminated neurological patients from diabetic patients (AUC = 0.96). Sixteen samples showed a hook effect, with direct (undiluted) values ≤273 IU/mL, while the corrected titers after dilution exceeded 20,000 IU/mL. A linear regression confirmed an inverse relationship between direct and true titers (r = –0.79, p &lt; 0.05).</div></div><div><h3>Discussion</h3><div>Maglumi anti-GAD65 immunoassay provides reliable discrimination between endocrine and neurological patients but requires method-specific cutoffs to avoid misclassification. Hook effect was frequent in high-titer neurological samples, potentially leading to misdiagnosis, underscoring the need for routine dilution protocols to improve diagnostic reliability.</div></div><div><h3>Conclusions</h3><div>Standardized dilution protocols and appropriate cutoffs are essential to ensure reliable anti-GAD65 testing with Maglumi chemiluminescent immunoassay, particularly in neurological indications.</div></div>\",\"PeriodicalId\":10172,\"journal\":{\"name\":\"Clinical biochemistry\",\"volume\":\"140 \",\"pages\":\"Article 111024\"},\"PeriodicalIF\":2.1000,\"publicationDate\":\"2025-10-10\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Clinical biochemistry\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0009912025001535\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical biochemistry","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0009912025001535","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

anti - gad65是与GAD抗体谱障碍相关的自身免疫性糖尿病和神经综合征的诊断性生物标志物。基于自动化学发光的免疫测定(如Maglumi)越来越多地使用,但它们在高抗体浓度下的行为特征仍然很差。我们报告了Maglumi和放射免疫测定(RIA)之间的分析一致性,区分神经性和糖尿病患者的截止效价,以及免疫测定的钩效应。材料和方法使用Maglumi 2000®和RIA (Medipan®)分析90例疑似自身免疫性糖尿病或神经综合征患者样本中的anti - gad65水平。分析一致性采用passingbablok回归、Bland-Altman图和Cohen’s kappa进行评估。ROC曲线分析确定了诊断阈值,并在出乎意料的低值样本中进行稀释研究,以评估潜在的钩效应。结果smaglumi结果比RIA结果高约5倍(斜率= 5.364,R2 = 0.908);内分泌组的一致性(R2 = 0.998)高于神经系统组(R2 = 0.891)。6778 IU/mL的临界值将神经系统患者与糖尿病患者区分开来(AUC = 0.96)。16个样品呈钩效应,直接(未稀释)值≤273 IU/mL,而稀释后的校正滴度超过20,000 IU/mL。线性回归证实了直接滴度和真实滴度之间的反比关系(r = -0.79, p < 0.05)。maglumi anti-GAD65免疫测定提供了内分泌和神经系统患者之间可靠的区分,但需要方法特异性的截止以避免误分类。Hook效应在高滴度神经学样本中很常见,可能导致误诊,强调需要常规稀释方案来提高诊断可靠性。结论:标准化的稀释方案和适当的截止点对于确保Maglumi化学发光免疫分析法可靠的抗gad65检测至关重要,特别是在神经系统适应症中。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Clinical relevance of the Maglumi immunoassay for anti-GAD65 in diabetes and neurological disorders: Analytical challenges and diagnostic insights

Introduction

Anti-GAD65 is an established diagnostic biomarker for autoimmune diabetes and neurological syndromes associated with GAD antibody-spectrum disorders. Automated chemiluminescence-based immunoassays such as Maglumi are increasingly used, but their behavior at high antibody concentrations remains poorly characterized. We report the analytical concordance between Maglumi and radioimmunoassay (RIA), the cut-off titer that differentiates neurological from diabetic patients, and the hook effect of the immunoassay.

Materials and methods

Anti-GAD65 levels were analyzed in samples from 90 patients with suspected autoimmune diabetes or neurological syndromes, using Maglumi 2000® and RIA (Medipan®). Analytical concordance was assessed by Passing–Bablok regression, Bland–Altman plots, and Cohen’s kappa. ROC curve analysis identified diagnostic thresholds, and dilution studies were performed in samples with unexpectedly low values to assess a potential hook effect.

Results

Maglumi results were approximately five times higher than RIA results (slope = 5.364; R2 = 0.908); agreement was higher in endocrine (R2 = 0.998) than in neurological patients (R2 = 0.891). A cutoff of 6778 IU/mL discriminated neurological patients from diabetic patients (AUC = 0.96). Sixteen samples showed a hook effect, with direct (undiluted) values ≤273 IU/mL, while the corrected titers after dilution exceeded 20,000 IU/mL. A linear regression confirmed an inverse relationship between direct and true titers (r = –0.79, p < 0.05).

Discussion

Maglumi anti-GAD65 immunoassay provides reliable discrimination between endocrine and neurological patients but requires method-specific cutoffs to avoid misclassification. Hook effect was frequent in high-titer neurological samples, potentially leading to misdiagnosis, underscoring the need for routine dilution protocols to improve diagnostic reliability.

Conclusions

Standardized dilution protocols and appropriate cutoffs are essential to ensure reliable anti-GAD65 testing with Maglumi chemiluminescent immunoassay, particularly in neurological indications.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Clinical biochemistry
Clinical biochemistry 医学-医学实验技术
CiteScore
5.10
自引率
0.00%
发文量
151
审稿时长
25 days
期刊介绍: Clinical Biochemistry publishes articles relating to clinical chemistry, molecular biology and genetics, therapeutic drug monitoring and toxicology, laboratory immunology and laboratory medicine in general, with the focus on analytical and clinical investigation of laboratory tests in humans used for diagnosis, prognosis, treatment and therapy, and monitoring of disease.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信