{"title":"[各种icbct对肺部适应性放疗图像质量影响的定量分析]。","authors":"Ruifeng Zhao, Bin Su, Xiaofei Jiang","doi":"10.12455/j.issn.1671-7104.250014","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To investigate the impact of different iterative cone beam CT (iCBCT) scanning beam currents from a ring-mounted linac on synthetic CT image quality for lung adaptive radiotherapy under lung scanning protocol.</p><p><strong>Methods: </strong>The CIRS lung motion phantom was configured to simulate conventional respiratory motion pattern, followed by 4D-CT simulation. After transferring the radiotherapy plan to the ring-mounted Halcyon 3.0 linac, three groups of typical iCBCT scans with different beam currents [ <i>I</i> <sub>low</sub> (160 mA), <i>I</i> <sub>middle</sub> (282 mA), and <i>I</i> <sub>high</sub> (491 mA)] were performed and corresponding image reconstructions were completed. Synthetic CT (sCT) images were subsequently obtained based on the deformable registration algorithm.</p><p><strong>Results: </strong>Compared to the corresponding CBCT images, the sCT images exhibited a significant reduction in artifacts. The fine structure of the planning CT (pCT) image was preserved for sCT images corresponding to different scanning beam currents, with Dice similarity coefficients exceeding 0.90 for all cases.</p><p><strong>Conclusion: </strong>The image quality of sCT corresponding to different iCBCTs is comparable to that of pCT, and changes in iCBCT beam parameters have a negligible impact on sCT image quality. Taking into account both image quality and imaging dose factors associated with the beam currents, iCBCT with a lower beam current on the ring-mounted Halcyon linac offers greater clinical value in lung adaptive radiotherapy.</p>","PeriodicalId":52535,"journal":{"name":"中国医疗器械杂志","volume":"49 4","pages":"423-428"},"PeriodicalIF":0.0000,"publicationDate":"2025-07-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Quantitative Analysis of the Impact of Various iCBCTs on the Image Quality of Lung Adaptive Radiotherapy].\",\"authors\":\"Ruifeng Zhao, Bin Su, Xiaofei Jiang\",\"doi\":\"10.12455/j.issn.1671-7104.250014\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>To investigate the impact of different iterative cone beam CT (iCBCT) scanning beam currents from a ring-mounted linac on synthetic CT image quality for lung adaptive radiotherapy under lung scanning protocol.</p><p><strong>Methods: </strong>The CIRS lung motion phantom was configured to simulate conventional respiratory motion pattern, followed by 4D-CT simulation. After transferring the radiotherapy plan to the ring-mounted Halcyon 3.0 linac, three groups of typical iCBCT scans with different beam currents [ <i>I</i> <sub>low</sub> (160 mA), <i>I</i> <sub>middle</sub> (282 mA), and <i>I</i> <sub>high</sub> (491 mA)] were performed and corresponding image reconstructions were completed. Synthetic CT (sCT) images were subsequently obtained based on the deformable registration algorithm.</p><p><strong>Results: </strong>Compared to the corresponding CBCT images, the sCT images exhibited a significant reduction in artifacts. The fine structure of the planning CT (pCT) image was preserved for sCT images corresponding to different scanning beam currents, with Dice similarity coefficients exceeding 0.90 for all cases.</p><p><strong>Conclusion: </strong>The image quality of sCT corresponding to different iCBCTs is comparable to that of pCT, and changes in iCBCT beam parameters have a negligible impact on sCT image quality. Taking into account both image quality and imaging dose factors associated with the beam currents, iCBCT with a lower beam current on the ring-mounted Halcyon linac offers greater clinical value in lung adaptive radiotherapy.</p>\",\"PeriodicalId\":52535,\"journal\":{\"name\":\"中国医疗器械杂志\",\"volume\":\"49 4\",\"pages\":\"423-428\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2025-07-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"中国医疗器械杂志\",\"FirstCategoryId\":\"1087\",\"ListUrlMain\":\"https://doi.org/10.12455/j.issn.1671-7104.250014\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国医疗器械杂志","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.12455/j.issn.1671-7104.250014","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
[Quantitative Analysis of the Impact of Various iCBCTs on the Image Quality of Lung Adaptive Radiotherapy].
Objective: To investigate the impact of different iterative cone beam CT (iCBCT) scanning beam currents from a ring-mounted linac on synthetic CT image quality for lung adaptive radiotherapy under lung scanning protocol.
Methods: The CIRS lung motion phantom was configured to simulate conventional respiratory motion pattern, followed by 4D-CT simulation. After transferring the radiotherapy plan to the ring-mounted Halcyon 3.0 linac, three groups of typical iCBCT scans with different beam currents [ Ilow (160 mA), Imiddle (282 mA), and Ihigh (491 mA)] were performed and corresponding image reconstructions were completed. Synthetic CT (sCT) images were subsequently obtained based on the deformable registration algorithm.
Results: Compared to the corresponding CBCT images, the sCT images exhibited a significant reduction in artifacts. The fine structure of the planning CT (pCT) image was preserved for sCT images corresponding to different scanning beam currents, with Dice similarity coefficients exceeding 0.90 for all cases.
Conclusion: The image quality of sCT corresponding to different iCBCTs is comparable to that of pCT, and changes in iCBCT beam parameters have a negligible impact on sCT image quality. Taking into account both image quality and imaging dose factors associated with the beam currents, iCBCT with a lower beam current on the ring-mounted Halcyon linac offers greater clinical value in lung adaptive radiotherapy.
期刊介绍:
Chinese Journal of Medical Instrumentation mainly reports on the development, progress, research and development, production, clinical application, management, and maintenance of medical devices and biomedical engineering. Its aim is to promote the exchange of information on medical devices and biomedical engineering in China and turn the journal into a high-quality academic journal that leads academic directions and advocates academic debates.