[用于肿瘤消融边界监测的电阻抗断层成像系统设计与实验研究]。

Q4 Medicine
Wei Wei, Lidong Xing, Xiaofei Jin, Zhiyu Qian, Jingqi Song, Kairan Wan, Haotian Wang
{"title":"[用于肿瘤消融边界监测的电阻抗断层成像系统设计与实验研究]。","authors":"Wei Wei, Lidong Xing, Xiaofei Jin, Zhiyu Qian, Jingqi Song, Kairan Wan, Haotian Wang","doi":"10.12455/j.issn.1671-7104.250010","DOIUrl":null,"url":null,"abstract":"<p><p>The minimally invasive thermal ablation technology differs from traditional surgical operations, which requires auxiliary equipment to evaluate ablation results. However, the ultrasound and CT currently used in clinical practice have shortcomings such as artifacts and radiation. Therefore, this paper proposes a design for a minimally invasive thermal ablation evaluation system based on the principle of electrical impedance tomography technology to monitor the ablation range. At the same time, the innovative introduction of a programmable gain feedforward signal as the parameter signal of the multiplier demodulator in the electrical impedance tomography system design can effectively solve the problem of weak signals being submerged in noise and improve imaging accuracy. The system controls the amplitude of the excitation current signal and the acquisition / processing of boundary voltages <i>via</i> an STM32, uploads the collected data to an upper computer, and reconstructs the conductivity distribution using the Newton-Raphson algorithm to map the size of the ablation area. Experimental results show that the system can effectively reflect the size of the microwave ablation area. Under the same minimally invasive ablation parameters, the average imaging errors are 0.6 mm for the long diameter, 0.8 mm for the short diameter, and 1.75% for the axial ratio (long diameter / short diameter), demonstrating high consistency. This verifies the technical potential of electrical impedance tomography in minimally invasive thermal ablation.</p>","PeriodicalId":52535,"journal":{"name":"中国医疗器械杂志","volume":"49 4","pages":"444-452"},"PeriodicalIF":0.0000,"publicationDate":"2025-07-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Design and Experimental Study of Electrical Impedance Tomography System for Tumor Ablation Boundary Monitoring].\",\"authors\":\"Wei Wei, Lidong Xing, Xiaofei Jin, Zhiyu Qian, Jingqi Song, Kairan Wan, Haotian Wang\",\"doi\":\"10.12455/j.issn.1671-7104.250010\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The minimally invasive thermal ablation technology differs from traditional surgical operations, which requires auxiliary equipment to evaluate ablation results. However, the ultrasound and CT currently used in clinical practice have shortcomings such as artifacts and radiation. Therefore, this paper proposes a design for a minimally invasive thermal ablation evaluation system based on the principle of electrical impedance tomography technology to monitor the ablation range. At the same time, the innovative introduction of a programmable gain feedforward signal as the parameter signal of the multiplier demodulator in the electrical impedance tomography system design can effectively solve the problem of weak signals being submerged in noise and improve imaging accuracy. The system controls the amplitude of the excitation current signal and the acquisition / processing of boundary voltages <i>via</i> an STM32, uploads the collected data to an upper computer, and reconstructs the conductivity distribution using the Newton-Raphson algorithm to map the size of the ablation area. Experimental results show that the system can effectively reflect the size of the microwave ablation area. Under the same minimally invasive ablation parameters, the average imaging errors are 0.6 mm for the long diameter, 0.8 mm for the short diameter, and 1.75% for the axial ratio (long diameter / short diameter), demonstrating high consistency. This verifies the technical potential of electrical impedance tomography in minimally invasive thermal ablation.</p>\",\"PeriodicalId\":52535,\"journal\":{\"name\":\"中国医疗器械杂志\",\"volume\":\"49 4\",\"pages\":\"444-452\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2025-07-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"中国医疗器械杂志\",\"FirstCategoryId\":\"1087\",\"ListUrlMain\":\"https://doi.org/10.12455/j.issn.1671-7104.250010\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国医疗器械杂志","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.12455/j.issn.1671-7104.250010","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

微创热消融技术不同于传统的外科手术,需要辅助设备来评估消融效果。但是目前临床上使用的超声和CT存在伪影、辐射等缺点。因此,本文提出了一种基于电阻抗断层成像技术原理的微创热消融评估系统的设计,以监测消融范围。同时,在电阻抗层析成像系统设计中创新性地引入可编程增益前馈信号作为乘法器的参数信号,可以有效地解决弱信号被噪声淹没的问题,提高成像精度。该系统通过STM32控制励磁电流信号的幅值和边界电压的采集/处理,将采集到的数据上传到上位机,利用Newton-Raphson算法重建电导率分布,绘制烧蚀区域的大小。实验结果表明,该系统能有效地反映微波烧蚀面积的大小。在相同的微创消融参数下,长径平均成像误差为0.6 mm,短径平均成像误差为0.8 mm,轴比(长径/短径)平均成像误差为1.75%,一致性较高。这证实了电阻抗断层扫描在微创热消融中的技术潜力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
[Design and Experimental Study of Electrical Impedance Tomography System for Tumor Ablation Boundary Monitoring].

The minimally invasive thermal ablation technology differs from traditional surgical operations, which requires auxiliary equipment to evaluate ablation results. However, the ultrasound and CT currently used in clinical practice have shortcomings such as artifacts and radiation. Therefore, this paper proposes a design for a minimally invasive thermal ablation evaluation system based on the principle of electrical impedance tomography technology to monitor the ablation range. At the same time, the innovative introduction of a programmable gain feedforward signal as the parameter signal of the multiplier demodulator in the electrical impedance tomography system design can effectively solve the problem of weak signals being submerged in noise and improve imaging accuracy. The system controls the amplitude of the excitation current signal and the acquisition / processing of boundary voltages via an STM32, uploads the collected data to an upper computer, and reconstructs the conductivity distribution using the Newton-Raphson algorithm to map the size of the ablation area. Experimental results show that the system can effectively reflect the size of the microwave ablation area. Under the same minimally invasive ablation parameters, the average imaging errors are 0.6 mm for the long diameter, 0.8 mm for the short diameter, and 1.75% for the axial ratio (long diameter / short diameter), demonstrating high consistency. This verifies the technical potential of electrical impedance tomography in minimally invasive thermal ablation.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
中国医疗器械杂志
中国医疗器械杂志 Medicine-Medicine (all)
CiteScore
0.40
自引率
0.00%
发文量
8086
期刊介绍: Chinese Journal of Medical Instrumentation mainly reports on the development, progress, research and development, production, clinical application, management, and maintenance of medical devices and biomedical engineering. Its aim is to promote the exchange of information on medical devices and biomedical engineering in China and turn the journal into a high-quality academic journal that leads academic directions and advocates academic debates.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信