{"title":"同步任意门- pw多普勒超声微血管成像的研制及初步验证","authors":"Chao Li, Lanxi Xiang, Yigang Du, Zhilan Zheng, Shuangshuang Li","doi":"10.12455/j.issn.1671-7104.240675","DOIUrl":null,"url":null,"abstract":"<p><p>Ultra micro angiography (UMA) can effectively improve the sensitivity and spatial resolution of blood flow imaging. To further meet the clinical requirements of quantitative measurements for microvascular imaging, this article demonstrated technical features and performance verification results of the microflow quantitative analysis function, which was developed based on the UMA function. This article provided preliminary validation of the measurement of Synchronous Arbitrary Gate-PW (SAG-PW) using a Doppler flow phantom and a moving string phantom. The Doppler flow phantom results demonstrated that the SAG-PW measurement values were close to the traditional PW results, and the average error of multiple sets of measurement results under different conditions was 3.9%. The results of SAG-PW and PW were strongly linearly correlated, with the slope and correlation coefficient approaching 1. The results of the moving string phantom demonstrated that the relative errors of TAMean and SAG-PW with different phantom set values were within 10%.</p>","PeriodicalId":52535,"journal":{"name":"中国医疗器械杂志","volume":"49 4","pages":"355-362"},"PeriodicalIF":0.0000,"publicationDate":"2025-07-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Development of Synchronous Arbitrary Gate-PW Doppler for Ultrasound Microvascular Imaging and Preliminary Validation].\",\"authors\":\"Chao Li, Lanxi Xiang, Yigang Du, Zhilan Zheng, Shuangshuang Li\",\"doi\":\"10.12455/j.issn.1671-7104.240675\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Ultra micro angiography (UMA) can effectively improve the sensitivity and spatial resolution of blood flow imaging. To further meet the clinical requirements of quantitative measurements for microvascular imaging, this article demonstrated technical features and performance verification results of the microflow quantitative analysis function, which was developed based on the UMA function. This article provided preliminary validation of the measurement of Synchronous Arbitrary Gate-PW (SAG-PW) using a Doppler flow phantom and a moving string phantom. The Doppler flow phantom results demonstrated that the SAG-PW measurement values were close to the traditional PW results, and the average error of multiple sets of measurement results under different conditions was 3.9%. The results of SAG-PW and PW were strongly linearly correlated, with the slope and correlation coefficient approaching 1. The results of the moving string phantom demonstrated that the relative errors of TAMean and SAG-PW with different phantom set values were within 10%.</p>\",\"PeriodicalId\":52535,\"journal\":{\"name\":\"中国医疗器械杂志\",\"volume\":\"49 4\",\"pages\":\"355-362\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2025-07-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"中国医疗器械杂志\",\"FirstCategoryId\":\"1087\",\"ListUrlMain\":\"https://doi.org/10.12455/j.issn.1671-7104.240675\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国医疗器械杂志","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.12455/j.issn.1671-7104.240675","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

超微血管造影(UMA)可以有效提高血流成像的灵敏度和空间分辨率。为进一步满足临床对微血管成像定量测量的需求,本文展示了在UMA功能基础上开发的微流量定量分析功能的技术特点及性能验证结果。本文对同步任意门- pw (SAG-PW)的多普勒流模和动弦模测量进行了初步验证。多普勒流模结果表明,SAG-PW测量值与传统PW结果接近,不同工况下多组测量结果的平均误差为3.9%。SAG-PW与PW结果呈强线性相关,斜率和相关系数接近1。运动弦幻体实验结果表明,不同幻体设定值的TAMean和SAG-PW的相对误差在10%以内。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
[Development of Synchronous Arbitrary Gate-PW Doppler for Ultrasound Microvascular Imaging and Preliminary Validation].

Ultra micro angiography (UMA) can effectively improve the sensitivity and spatial resolution of blood flow imaging. To further meet the clinical requirements of quantitative measurements for microvascular imaging, this article demonstrated technical features and performance verification results of the microflow quantitative analysis function, which was developed based on the UMA function. This article provided preliminary validation of the measurement of Synchronous Arbitrary Gate-PW (SAG-PW) using a Doppler flow phantom and a moving string phantom. The Doppler flow phantom results demonstrated that the SAG-PW measurement values were close to the traditional PW results, and the average error of multiple sets of measurement results under different conditions was 3.9%. The results of SAG-PW and PW were strongly linearly correlated, with the slope and correlation coefficient approaching 1. The results of the moving string phantom demonstrated that the relative errors of TAMean and SAG-PW with different phantom set values were within 10%.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
中国医疗器械杂志
中国医疗器械杂志 Medicine-Medicine (all)
CiteScore
0.40
自引率
0.00%
发文量
8086
期刊介绍: Chinese Journal of Medical Instrumentation mainly reports on the development, progress, research and development, production, clinical application, management, and maintenance of medical devices and biomedical engineering. Its aim is to promote the exchange of information on medical devices and biomedical engineering in China and turn the journal into a high-quality academic journal that leads academic directions and advocates academic debates.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信