{"title":"201Tl心肌灌注SPECT非平滑图像掩模方法的有效性评价。","authors":"Ryuichi Miyajima, Ryo Ueno, Ryosuke Ichino, Satomi Teraoka, Ishikawa Yasushi, Masahiro Sonoda","doi":"10.1007/s12194-025-00959-2","DOIUrl":null,"url":null,"abstract":"<p><p>In <sup>201</sup>Tl myocardial perfusion single-photon emission computed tomography (SPECT), gastric wall uptake can impact the inferior wall. This study aimed to evaluate the effectiveness and usefulness of the masking on un-smoothed image (MUS) method for <sup>201</sup>Tl myocardial perfusion SPECT. A hemispherical gastric wall phantom was created to simulate the gastric fundus located closest to the myocardium, and the activity was enclosed to achieve an SPECT count ratio against the myocardium equivalent to that observed in clinical practice. The minimum values of the defect chip in the circumferential profile curve were compared for six SPECT count ratios and seven gap distances. In the conventional method, increasing SPECT count ratios or gap distances interfered with myocardial perfusion SPECT evaluation. Artifacts were less apparent when the MUS method was applied. The MUS method effectively suppressed the gastric wall uptake on <sup>201</sup>Tl myocardial perfusion SPECT.</p>","PeriodicalId":46252,"journal":{"name":"Radiological Physics and Technology","volume":" ","pages":""},"PeriodicalIF":1.5000,"publicationDate":"2025-09-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Evaluation of the usefulness of the masking on un-smoothed image method in <sup>201</sup>Tl myocardial perfusion SPECT.\",\"authors\":\"Ryuichi Miyajima, Ryo Ueno, Ryosuke Ichino, Satomi Teraoka, Ishikawa Yasushi, Masahiro Sonoda\",\"doi\":\"10.1007/s12194-025-00959-2\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>In <sup>201</sup>Tl myocardial perfusion single-photon emission computed tomography (SPECT), gastric wall uptake can impact the inferior wall. This study aimed to evaluate the effectiveness and usefulness of the masking on un-smoothed image (MUS) method for <sup>201</sup>Tl myocardial perfusion SPECT. A hemispherical gastric wall phantom was created to simulate the gastric fundus located closest to the myocardium, and the activity was enclosed to achieve an SPECT count ratio against the myocardium equivalent to that observed in clinical practice. The minimum values of the defect chip in the circumferential profile curve were compared for six SPECT count ratios and seven gap distances. In the conventional method, increasing SPECT count ratios or gap distances interfered with myocardial perfusion SPECT evaluation. Artifacts were less apparent when the MUS method was applied. The MUS method effectively suppressed the gastric wall uptake on <sup>201</sup>Tl myocardial perfusion SPECT.</p>\",\"PeriodicalId\":46252,\"journal\":{\"name\":\"Radiological Physics and Technology\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":1.5000,\"publicationDate\":\"2025-09-10\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Radiological Physics and Technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1007/s12194-025-00959-2\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiological Physics and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s12194-025-00959-2","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

在201Tl心肌灌注单光子发射计算机断层扫描(SPECT)中,胃壁摄取可影响下壁。本研究旨在评价非平滑图像(MUS)方法在201Tl心肌灌注SPECT中的有效性和实用性。创建一个半球形胃壁幻象来模拟位于最靠近心肌的胃底,并将其活动封闭以达到与临床实践中观察到的心肌相当的SPECT计数比。比较了6种SPECT计数比和7种间隙距离下缺陷芯片在周向轮廓曲线上的最小值。在常规方法中,增加SPECT计数比或间隙距离会干扰心肌灌注SPECT评价。当应用MUS方法时,工件不太明显。在201Tl心肌灌注SPECT上,MUS法能有效抑制胃壁摄取。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Evaluation of the usefulness of the masking on un-smoothed image method in 201Tl myocardial perfusion SPECT.

In 201Tl myocardial perfusion single-photon emission computed tomography (SPECT), gastric wall uptake can impact the inferior wall. This study aimed to evaluate the effectiveness and usefulness of the masking on un-smoothed image (MUS) method for 201Tl myocardial perfusion SPECT. A hemispherical gastric wall phantom was created to simulate the gastric fundus located closest to the myocardium, and the activity was enclosed to achieve an SPECT count ratio against the myocardium equivalent to that observed in clinical practice. The minimum values of the defect chip in the circumferential profile curve were compared for six SPECT count ratios and seven gap distances. In the conventional method, increasing SPECT count ratios or gap distances interfered with myocardial perfusion SPECT evaluation. Artifacts were less apparent when the MUS method was applied. The MUS method effectively suppressed the gastric wall uptake on 201Tl myocardial perfusion SPECT.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Radiological Physics and Technology
Radiological Physics and Technology RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
3.00
自引率
12.50%
发文量
40
期刊介绍: The purpose of the journal Radiological Physics and Technology is to provide a forum for sharing new knowledge related to research and development in radiological science and technology, including medical physics and radiological technology in diagnostic radiology, nuclear medicine, and radiation therapy among many other radiological disciplines, as well as to contribute to progress and improvement in medical practice and patient health care.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信