{"title":"近距离放射治疗中使用日内瓦和威尼斯涂抹器的x线透视评估源位置:涂抹器间差异的评估。","authors":"Sayaka Kihara, Yoshihiro Ueda, Masahiro Morimoto, Setsuo Tamenaga, Shoki Inui, Akira Masaoka, Yuya Nitta, Masaru Isono, Teiji Nishio, Koji Konishi","doi":"10.1016/j.brachy.2025.07.004","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study aimed to evaluate the inter-applicator differences in source positions of the Geneva and Venezia applicators by using X-ray fluoroscopy to perform direct measurements of iridium-192 (<sup>192</sup>Ir) source positions.</p><p><strong>Materials and methods: </strong>Offset measurements of two to six sets of Geneva tandem and ovoid applicators and Venezia tandem applicators were performed. Each applicator was irradiated with <sup>192</sup>Ir at the most distal position and X-ray fluoroscopy images were acquired. The offset values were measured as the distance from the applicator tip to the center of <sup>192</sup>Ir source. For two sets of the Venezia ovoid applicators (set 1 and set 2), X-ray fluoroscopy images were acquired at a 2-mm step from the most distal position. The <sup>192</sup>Ir source positions were compared with the expected source positions using an applicator modeling module.</p><p><strong>Results: </strong>The differences in offset values of the same types of applicators were up to 0.6 mm for Geneva tandem and ovoid applicators and 0.3 mm for Venezia tandem applicators. The mean standard deviation calculated for the same applicator types was 0.1 mm across all applicators. For Venezia ovoid applicators, the mean difference between <sup>192</sup>Ir source positions and expected source positions were 0.1 mm for the set 1 applicator (maximum: 0.6 mm) and -0.2 mm for the set 2 applicator (maximum: 0.6 mm).</p><p><strong>Conclusion: </strong>This study demonstrated the feasibility of direct source position measurements using X-ray fluoroscopy. These results emphasize the need for commissioning all applicator, but individual characterization is unnecessary in clinical practice if differences fall within tolerance.</p>","PeriodicalId":93914,"journal":{"name":"Brachytherapy","volume":" ","pages":""},"PeriodicalIF":1.8000,"publicationDate":"2025-08-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Source position evaluation using X-ray fluoroscopy of the Geneva and Venezia applicators for brachytherapy: An assessment of inter-applicator differences.\",\"authors\":\"Sayaka Kihara, Yoshihiro Ueda, Masahiro Morimoto, Setsuo Tamenaga, Shoki Inui, Akira Masaoka, Yuya Nitta, Masaru Isono, Teiji Nishio, Koji Konishi\",\"doi\":\"10.1016/j.brachy.2025.07.004\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>This study aimed to evaluate the inter-applicator differences in source positions of the Geneva and Venezia applicators by using X-ray fluoroscopy to perform direct measurements of iridium-192 (<sup>192</sup>Ir) source positions.</p><p><strong>Materials and methods: </strong>Offset measurements of two to six sets of Geneva tandem and ovoid applicators and Venezia tandem applicators were performed. Each applicator was irradiated with <sup>192</sup>Ir at the most distal position and X-ray fluoroscopy images were acquired. The offset values were measured as the distance from the applicator tip to the center of <sup>192</sup>Ir source. For two sets of the Venezia ovoid applicators (set 1 and set 2), X-ray fluoroscopy images were acquired at a 2-mm step from the most distal position. The <sup>192</sup>Ir source positions were compared with the expected source positions using an applicator modeling module.</p><p><strong>Results: </strong>The differences in offset values of the same types of applicators were up to 0.6 mm for Geneva tandem and ovoid applicators and 0.3 mm for Venezia tandem applicators. The mean standard deviation calculated for the same applicator types was 0.1 mm across all applicators. For Venezia ovoid applicators, the mean difference between <sup>192</sup>Ir source positions and expected source positions were 0.1 mm for the set 1 applicator (maximum: 0.6 mm) and -0.2 mm for the set 2 applicator (maximum: 0.6 mm).</p><p><strong>Conclusion: </strong>This study demonstrated the feasibility of direct source position measurements using X-ray fluoroscopy. These results emphasize the need for commissioning all applicator, but individual characterization is unnecessary in clinical practice if differences fall within tolerance.</p>\",\"PeriodicalId\":93914,\"journal\":{\"name\":\"Brachytherapy\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":1.8000,\"publicationDate\":\"2025-08-28\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Brachytherapy\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1016/j.brachy.2025.07.004\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Brachytherapy","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1016/j.brachy.2025.07.004","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:本研究旨在通过x射线透视直接测量铱-192 (192Ir)源位置,评估日内瓦和威尼斯涂抹器间源位置的差异。材料和方法:进行了2至6套Geneva串联和卵形涂抹器和Venezia串联涂抹器的偏移测量。每个涂抹器在最远端位置用192Ir照射,并获得x线透视图像。偏移值测量为从涂抹器尖端到192Ir光源中心的距离。对于两套Venezia卵形涂药器(set 1和set 2), x线透视图像从最远端位置2毫米处获取。使用涂抹器建模模块将192Ir源位置与预期源位置进行比较。结果:相同类型的施药器中,Geneva串联施药器和卵形施药器的偏移值差异可达0.6 mm, Venezia串联施药器的偏移值差异可达0.3 mm。在所有涂布器中,相同涂布器类型的平均标准差为0.1 mm。对于Venezia卵形涂抹器,第1组涂抹器的192Ir源位置与预期源位置之间的平均差异为0.1 mm(最大值:0.6 mm),第2组涂抹器的-0.2 mm(最大值:0.6 mm)。结论:本研究证明了x线透视直接测量源位置的可行性。这些结果强调需要调试所有涂抹器,但在临床实践中,如果差异在容忍范围内,则无需进行个体表征。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Source position evaluation using X-ray fluoroscopy of the Geneva and Venezia applicators for brachytherapy: An assessment of inter-applicator differences.

Purpose: This study aimed to evaluate the inter-applicator differences in source positions of the Geneva and Venezia applicators by using X-ray fluoroscopy to perform direct measurements of iridium-192 (192Ir) source positions.

Materials and methods: Offset measurements of two to six sets of Geneva tandem and ovoid applicators and Venezia tandem applicators were performed. Each applicator was irradiated with 192Ir at the most distal position and X-ray fluoroscopy images were acquired. The offset values were measured as the distance from the applicator tip to the center of 192Ir source. For two sets of the Venezia ovoid applicators (set 1 and set 2), X-ray fluoroscopy images were acquired at a 2-mm step from the most distal position. The 192Ir source positions were compared with the expected source positions using an applicator modeling module.

Results: The differences in offset values of the same types of applicators were up to 0.6 mm for Geneva tandem and ovoid applicators and 0.3 mm for Venezia tandem applicators. The mean standard deviation calculated for the same applicator types was 0.1 mm across all applicators. For Venezia ovoid applicators, the mean difference between 192Ir source positions and expected source positions were 0.1 mm for the set 1 applicator (maximum: 0.6 mm) and -0.2 mm for the set 2 applicator (maximum: 0.6 mm).

Conclusion: This study demonstrated the feasibility of direct source position measurements using X-ray fluoroscopy. These results emphasize the need for commissioning all applicator, but individual characterization is unnecessary in clinical practice if differences fall within tolerance.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信