在皮下植入式心律转复除颤器中出现罕见的高振幅b音信号:诊断挑战和处理。

IF 2.6 3区 医学 Q2 CARDIAC & CARDIOVASCULAR SYSTEMS
Yukihiro Uehara, Nobuhiko Ueda, Akinori Wakamiya, Kohei Ishibashi, Kengo Kusano
{"title":"在皮下植入式心律转复除颤器中出现罕见的高振幅b音信号:诊断挑战和处理。","authors":"Yukihiro Uehara, Nobuhiko Ueda, Akinori Wakamiya, Kohei Ishibashi, Kengo Kusano","doi":"10.1111/jce.70076","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Sense-B-noise is a rare cause of inappropriate shock (IAS) of subcutaneous implantable cardioverter defibrillator (S-ICD), which is an oversensing of the vectors using the Sense B and a factor of noncardiac signal oversensing. High-amplitude signals in Sense-B-noise have not been reported previously.</p><p><strong>Methods: </strong>We present a case of a 47-year-old man, who experienced IAS of S-ICD due to sense-B-noise followed by intense noise of primary and alternate vectors.</p><p><strong>Results: </strong>Analysis of the extracted lead revealed a slight tear in the connection of the Sense B electrode, but the non-physiological signal was not reproducible and no functional defects were identified in the device.</p><p><strong>Conclusions: </strong>High-amplitude signals after an inappropriate shock due to sense-B-noise are another presentation of Sense-B-noise.</p>","PeriodicalId":15178,"journal":{"name":"Journal of Cardiovascular Electrophysiology","volume":" ","pages":""},"PeriodicalIF":2.6000,"publicationDate":"2025-08-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A Rare Presentation of Sense-B-Noise With High-Amplitude Signals in a Subcutaneous Implantable Cardioverter Defibrillator: Diagnostic Challenges and Management.\",\"authors\":\"Yukihiro Uehara, Nobuhiko Ueda, Akinori Wakamiya, Kohei Ishibashi, Kengo Kusano\",\"doi\":\"10.1111/jce.70076\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>Sense-B-noise is a rare cause of inappropriate shock (IAS) of subcutaneous implantable cardioverter defibrillator (S-ICD), which is an oversensing of the vectors using the Sense B and a factor of noncardiac signal oversensing. High-amplitude signals in Sense-B-noise have not been reported previously.</p><p><strong>Methods: </strong>We present a case of a 47-year-old man, who experienced IAS of S-ICD due to sense-B-noise followed by intense noise of primary and alternate vectors.</p><p><strong>Results: </strong>Analysis of the extracted lead revealed a slight tear in the connection of the Sense B electrode, but the non-physiological signal was not reproducible and no functional defects were identified in the device.</p><p><strong>Conclusions: </strong>High-amplitude signals after an inappropriate shock due to sense-B-noise are another presentation of Sense-B-noise.</p>\",\"PeriodicalId\":15178,\"journal\":{\"name\":\"Journal of Cardiovascular Electrophysiology\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":2.6000,\"publicationDate\":\"2025-08-26\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Cardiovascular Electrophysiology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1111/jce.70076\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Cardiovascular Electrophysiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1111/jce.70076","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

摘要

B感噪声是皮下植入式心律转复除颤器(S-ICD)不适当休克(IAS)的罕见原因,它是利用B感对载体的过感和非心脏信号过感的一个因素。b -噪声中的高振幅信号以前未见报道。方法:我们报告了一位47岁男性患者,由于b感噪声,然后是主要和替代载体的强烈噪声,他经历了S-ICD的IAS。结果:对提取的铅进行分析,发现B电极连接处有轻微撕裂,但非生理信号不可重现,设备未发现功能缺陷。结论:b -噪声引起的不适当电击后的高振幅信号是b -噪声的另一种表现。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
A Rare Presentation of Sense-B-Noise With High-Amplitude Signals in a Subcutaneous Implantable Cardioverter Defibrillator: Diagnostic Challenges and Management.

Introduction: Sense-B-noise is a rare cause of inappropriate shock (IAS) of subcutaneous implantable cardioverter defibrillator (S-ICD), which is an oversensing of the vectors using the Sense B and a factor of noncardiac signal oversensing. High-amplitude signals in Sense-B-noise have not been reported previously.

Methods: We present a case of a 47-year-old man, who experienced IAS of S-ICD due to sense-B-noise followed by intense noise of primary and alternate vectors.

Results: Analysis of the extracted lead revealed a slight tear in the connection of the Sense B electrode, but the non-physiological signal was not reproducible and no functional defects were identified in the device.

Conclusions: High-amplitude signals after an inappropriate shock due to sense-B-noise are another presentation of Sense-B-noise.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
5.20
自引率
14.80%
发文量
433
审稿时长
3-6 weeks
期刊介绍: Journal of Cardiovascular Electrophysiology (JCE) keeps its readership well informed of the latest developments in the study and management of arrhythmic disorders. Edited by Bradley P. Knight, M.D., and a distinguished international editorial board, JCE is the leading journal devoted to the study of the electrophysiology of the heart.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信