{"title":"低剂量双场旋转TSET的调试与临床实施","authors":"Broderick Ivan McCallum-Hee, Mounir Ibrahim, Godfrey Mukwada, Pejman Rowshanfarzad, Joshua Dass, Jude Dewitt, Robert Parin, Gregory Withey, Zaid Alkhatib","doi":"10.1002/acm2.70180","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Total skin electron therapy (TSET) is a specialized technique for treating cutaneous T-cell lymphomas, such as mycosis fungoides. While traditional TSET regimens prescribe 30–36 Gy, low dose TSET at 12 Gy has gained attention due to its reduced toxicity. The dual-field rotational TSET technique, utilizing a standard linac, offers a practical solution for improved dose distribution. However, limited studies exist on its commissioning and clinical implementation.</p>\n </section>\n \n <section>\n \n <h3> Purpose</h3>\n \n <p>This study aimed to commission and clinically implement a low dose dual-field rotational TSET technique. The primary objectives were to optimize beam geometry, characterize dosimetry, and ensure treatment uniformity while maintaining compliance with ACPSEM, EORTC, and AAPM guidance.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>A Varian TrueBeam linac equipped with a 6 MeV high-dose-rate total skin electron beam was commissioned for TSET. Dosimetric measurements included beam output calibration, relative dosimetry, and dose uniformity assessment using an ion chamber and Gafchromic film, with verification using an anthropomorphic phantom. A treatment dose calculation methodology was developed. Shielding, ozone generation, and quality assurance were assessed. In vivo dosimetry was performed for treatment validation.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The optimal dual-field beam geometry was determined to be ± 18° from the horizontal, achieving vertical uniformity within ± 2.9% and an overall treatment plane uniformity of ± 8.6%. Under rotation, the depth dose is delivered 100% at the surface, with bremsstrahlung contamination of 0.2 Gy. Custom eye shielding was developed, and ozone concentrations remained below the NCRP 0.1 parts per million (ppm) safety thresholds. In vivo dosimetry confirmed treatment uniformity within EORTC guidelines and identified three regions requiring dose boosts: the shoulders, palms, and inner thighs.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>The commissioned dual-field rotational TSET technique provides a viable treatment option, achieving clinically acceptable dose distribution.</p>\n </section>\n </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"26 7","pages":""},"PeriodicalIF":2.0000,"publicationDate":"2025-07-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/acm2.70180","citationCount":"0","resultStr":"{\"title\":\"Commissioning and clinical implementation of low dose dual-field rotational TSET\",\"authors\":\"Broderick Ivan McCallum-Hee, Mounir Ibrahim, Godfrey Mukwada, Pejman Rowshanfarzad, Joshua Dass, Jude Dewitt, Robert Parin, Gregory Withey, Zaid Alkhatib\",\"doi\":\"10.1002/acm2.70180\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Background</h3>\\n \\n <p>Total skin electron therapy (TSET) is a specialized technique for treating cutaneous T-cell lymphomas, such as mycosis fungoides. While traditional TSET regimens prescribe 30–36 Gy, low dose TSET at 12 Gy has gained attention due to its reduced toxicity. The dual-field rotational TSET technique, utilizing a standard linac, offers a practical solution for improved dose distribution. However, limited studies exist on its commissioning and clinical implementation.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Purpose</h3>\\n \\n <p>This study aimed to commission and clinically implement a low dose dual-field rotational TSET technique. The primary objectives were to optimize beam geometry, characterize dosimetry, and ensure treatment uniformity while maintaining compliance with ACPSEM, EORTC, and AAPM guidance.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Methods</h3>\\n \\n <p>A Varian TrueBeam linac equipped with a 6 MeV high-dose-rate total skin electron beam was commissioned for TSET. Dosimetric measurements included beam output calibration, relative dosimetry, and dose uniformity assessment using an ion chamber and Gafchromic film, with verification using an anthropomorphic phantom. A treatment dose calculation methodology was developed. Shielding, ozone generation, and quality assurance were assessed. In vivo dosimetry was performed for treatment validation.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>The optimal dual-field beam geometry was determined to be ± 18° from the horizontal, achieving vertical uniformity within ± 2.9% and an overall treatment plane uniformity of ± 8.6%. Under rotation, the depth dose is delivered 100% at the surface, with bremsstrahlung contamination of 0.2 Gy. Custom eye shielding was developed, and ozone concentrations remained below the NCRP 0.1 parts per million (ppm) safety thresholds. In vivo dosimetry confirmed treatment uniformity within EORTC guidelines and identified three regions requiring dose boosts: the shoulders, palms, and inner thighs.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusions</h3>\\n \\n <p>The commissioned dual-field rotational TSET technique provides a viable treatment option, achieving clinically acceptable dose distribution.</p>\\n </section>\\n </div>\",\"PeriodicalId\":14989,\"journal\":{\"name\":\"Journal of Applied Clinical Medical Physics\",\"volume\":\"26 7\",\"pages\":\"\"},\"PeriodicalIF\":2.0000,\"publicationDate\":\"2025-07-15\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://onlinelibrary.wiley.com/doi/epdf/10.1002/acm2.70180\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Applied Clinical Medical Physics\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1002/acm2.70180\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/acm2.70180","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
Commissioning and clinical implementation of low dose dual-field rotational TSET
Background
Total skin electron therapy (TSET) is a specialized technique for treating cutaneous T-cell lymphomas, such as mycosis fungoides. While traditional TSET regimens prescribe 30–36 Gy, low dose TSET at 12 Gy has gained attention due to its reduced toxicity. The dual-field rotational TSET technique, utilizing a standard linac, offers a practical solution for improved dose distribution. However, limited studies exist on its commissioning and clinical implementation.
Purpose
This study aimed to commission and clinically implement a low dose dual-field rotational TSET technique. The primary objectives were to optimize beam geometry, characterize dosimetry, and ensure treatment uniformity while maintaining compliance with ACPSEM, EORTC, and AAPM guidance.
Methods
A Varian TrueBeam linac equipped with a 6 MeV high-dose-rate total skin electron beam was commissioned for TSET. Dosimetric measurements included beam output calibration, relative dosimetry, and dose uniformity assessment using an ion chamber and Gafchromic film, with verification using an anthropomorphic phantom. A treatment dose calculation methodology was developed. Shielding, ozone generation, and quality assurance were assessed. In vivo dosimetry was performed for treatment validation.
Results
The optimal dual-field beam geometry was determined to be ± 18° from the horizontal, achieving vertical uniformity within ± 2.9% and an overall treatment plane uniformity of ± 8.6%. Under rotation, the depth dose is delivered 100% at the surface, with bremsstrahlung contamination of 0.2 Gy. Custom eye shielding was developed, and ozone concentrations remained below the NCRP 0.1 parts per million (ppm) safety thresholds. In vivo dosimetry confirmed treatment uniformity within EORTC guidelines and identified three regions requiring dose boosts: the shoulders, palms, and inner thighs.
Conclusions
The commissioned dual-field rotational TSET technique provides a viable treatment option, achieving clinically acceptable dose distribution.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic