Bao Li, Yongzhong Chen, Jun Jin, Longjun Yan, Xiaoyong Wang
{"title":"锥形束CT图像在骨盆肿瘤剂量计算中的准确性评估。","authors":"Bao Li, Yongzhong Chen, Jun Jin, Longjun Yan, Xiaoyong Wang","doi":"10.12455/j.issn.1671-7104.240655","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To evaluate the feasibility and accuracy of cone-beam CT (CBCT) images for radiotherapy dose calculation in pelvic tumors.</p><p><strong>Methods: </strong>An improved volumetric density coverage method was used to establish CT value-relative electron density (RED) curves for CBCT images. The planning CT plans were transferred to the CBCT images, and the constructed density curves were applied to calculate doses for CBCT plans while maintaining the optimization parameters unchanged. Dose calculation deviations between the two plans were analyzed.</p><p><strong>Results: </strong>The mean differences in dosimetric parameters for the target volume and organs at risk (OAR) between the two plans were less than 1% and 1.5%, respectively. The target conformity index (CI), homogeneity index (HI), and gamma passing rates were highly consistent, with no statistically significant differences.</p><p><strong>Conclusion: </strong>CBCT images corrected by this method can be used for dose calculation in pelvic tumor radiotherapy planning.</p>","PeriodicalId":52535,"journal":{"name":"中国医疗器械杂志","volume":"49 3","pages":"302-307"},"PeriodicalIF":0.0000,"publicationDate":"2025-05-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Accuracy Assessment of Cone-Beam CT Images for Pelvic Tumor Dose Calculation].\",\"authors\":\"Bao Li, Yongzhong Chen, Jun Jin, Longjun Yan, Xiaoyong Wang\",\"doi\":\"10.12455/j.issn.1671-7104.240655\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>To evaluate the feasibility and accuracy of cone-beam CT (CBCT) images for radiotherapy dose calculation in pelvic tumors.</p><p><strong>Methods: </strong>An improved volumetric density coverage method was used to establish CT value-relative electron density (RED) curves for CBCT images. The planning CT plans were transferred to the CBCT images, and the constructed density curves were applied to calculate doses for CBCT plans while maintaining the optimization parameters unchanged. Dose calculation deviations between the two plans were analyzed.</p><p><strong>Results: </strong>The mean differences in dosimetric parameters for the target volume and organs at risk (OAR) between the two plans were less than 1% and 1.5%, respectively. The target conformity index (CI), homogeneity index (HI), and gamma passing rates were highly consistent, with no statistically significant differences.</p><p><strong>Conclusion: </strong>CBCT images corrected by this method can be used for dose calculation in pelvic tumor radiotherapy planning.</p>\",\"PeriodicalId\":52535,\"journal\":{\"name\":\"中国医疗器械杂志\",\"volume\":\"49 3\",\"pages\":\"302-307\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2025-05-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"中国医疗器械杂志\",\"FirstCategoryId\":\"1087\",\"ListUrlMain\":\"https://doi.org/10.12455/j.issn.1671-7104.240655\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国医疗器械杂志","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.12455/j.issn.1671-7104.240655","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
[Accuracy Assessment of Cone-Beam CT Images for Pelvic Tumor Dose Calculation].
Objective: To evaluate the feasibility and accuracy of cone-beam CT (CBCT) images for radiotherapy dose calculation in pelvic tumors.
Methods: An improved volumetric density coverage method was used to establish CT value-relative electron density (RED) curves for CBCT images. The planning CT plans were transferred to the CBCT images, and the constructed density curves were applied to calculate doses for CBCT plans while maintaining the optimization parameters unchanged. Dose calculation deviations between the two plans were analyzed.
Results: The mean differences in dosimetric parameters for the target volume and organs at risk (OAR) between the two plans were less than 1% and 1.5%, respectively. The target conformity index (CI), homogeneity index (HI), and gamma passing rates were highly consistent, with no statistically significant differences.
Conclusion: CBCT images corrected by this method can be used for dose calculation in pelvic tumor radiotherapy planning.
期刊介绍:
Chinese Journal of Medical Instrumentation mainly reports on the development, progress, research and development, production, clinical application, management, and maintenance of medical devices and biomedical engineering. Its aim is to promote the exchange of information on medical devices and biomedical engineering in China and turn the journal into a high-quality academic journal that leads academic directions and advocates academic debates.