室内CT-on-rails成像引导质子治疗的地形图和3DCT几何校正

IF 3.4 Q2 ONCOLOGY
Giovanni Fattori , Riccardo Via , Antony J. Lomax , Sairos Safai
{"title":"室内CT-on-rails成像引导质子治疗的地形图和3DCT几何校正","authors":"Giovanni Fattori ,&nbsp;Riccardo Via ,&nbsp;Antony J. Lomax ,&nbsp;Sairos Safai","doi":"10.1016/j.phro.2025.100799","DOIUrl":null,"url":null,"abstract":"<div><div>In-room computer tomography (CT) on-rails scanners are available in proton therapy centers but often limited to control imaging, due to geometric distortion and lack of integration. We present a calibration method combining CT-on-rails with a robotic table to achieve sub-millimeter patient positioning accuracy. Geometric distortions in 3DCT and topogram images were corrected using affine models and validated with laser tracker data. Phantom experiments simulating clinical conditions showed errors below 0.15 mm for 3D/3D and 0.3–0.55 mm for 2D/3D registration, the latter using a novel algorithm for digitally reconstructed topograms. The workflow enables accurate positioning without isocenter verification, supporting high-precision radiotherapy.</div></div>","PeriodicalId":36850,"journal":{"name":"Physics and Imaging in Radiation Oncology","volume":"35 ","pages":"Article 100799"},"PeriodicalIF":3.4000,"publicationDate":"2025-06-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Topogram and 3DCT geometry calibration for image-guided proton therapy with in-room CT-on-rails\",\"authors\":\"Giovanni Fattori ,&nbsp;Riccardo Via ,&nbsp;Antony J. Lomax ,&nbsp;Sairos Safai\",\"doi\":\"10.1016/j.phro.2025.100799\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><div>In-room computer tomography (CT) on-rails scanners are available in proton therapy centers but often limited to control imaging, due to geometric distortion and lack of integration. We present a calibration method combining CT-on-rails with a robotic table to achieve sub-millimeter patient positioning accuracy. Geometric distortions in 3DCT and topogram images were corrected using affine models and validated with laser tracker data. Phantom experiments simulating clinical conditions showed errors below 0.15 mm for 3D/3D and 0.3–0.55 mm for 2D/3D registration, the latter using a novel algorithm for digitally reconstructed topograms. The workflow enables accurate positioning without isocenter verification, supporting high-precision radiotherapy.</div></div>\",\"PeriodicalId\":36850,\"journal\":{\"name\":\"Physics and Imaging in Radiation Oncology\",\"volume\":\"35 \",\"pages\":\"Article 100799\"},\"PeriodicalIF\":3.4000,\"publicationDate\":\"2025-06-24\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Physics and Imaging in Radiation Oncology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2405631625001046\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"ONCOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physics and Imaging in Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2405631625001046","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

室内计算机断层扫描(CT)轨上扫描仪可用于质子治疗中心,但由于几何畸变和缺乏集成,通常仅限于控制成像。我们提出了一种结合CT-on-rails和机器人工作台的校准方法,以达到亚毫米级的患者定位精度。利用仿射模型对三维ct和地形图图像的几何畸变进行校正,并用激光跟踪器数据进行验证。模拟临床条件的幻影实验显示,3D/3D配准误差在0.15 mm以下,2D/3D配准误差在0.3-0.55 mm之间,后者使用了一种新的数字重建地形算法。该工作流程无需等中心验证即可实现精确定位,支持高精度放疗。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Topogram and 3DCT geometry calibration for image-guided proton therapy with in-room CT-on-rails
In-room computer tomography (CT) on-rails scanners are available in proton therapy centers but often limited to control imaging, due to geometric distortion and lack of integration. We present a calibration method combining CT-on-rails with a robotic table to achieve sub-millimeter patient positioning accuracy. Geometric distortions in 3DCT and topogram images were corrected using affine models and validated with laser tracker data. Phantom experiments simulating clinical conditions showed errors below 0.15 mm for 3D/3D and 0.3–0.55 mm for 2D/3D registration, the latter using a novel algorithm for digitally reconstructed topograms. The workflow enables accurate positioning without isocenter verification, supporting high-precision radiotherapy.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Physics and Imaging in Radiation Oncology
Physics and Imaging in Radiation Oncology Physics and Astronomy-Radiation
CiteScore
5.30
自引率
18.90%
发文量
93
审稿时长
6 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信