{"title":"深硅基光子计数CT的碘定量性能:虚拟成像试验研究","authors":"Raj Kumar Panta ,&nbsp;Zhye Yin ,&nbsp;Fredrik Grönberg ,&nbsp;Mridul Bhattarai ,&nbsp;Ehsan Abadi ,&nbsp;Paul Segars ,&nbsp;Ehsan Samei","doi":"10.1016/j.ejmp.2025.105003","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><div>This study investigates the imaging performance of a deep silicon-based photon-counting CT (Si-PCCT) in quantifying iodine contrast through a virtual imaging trial (VIT).</div></div><div><h3>Methods</h3><div>We developed a VIT framework using Si-PCCT simulator and benchmarked it against a prototype using an ACR phantom for assessing spatial resolution and noise characteristics, and a geometric phantom for iodine quantification. We imaged geometrical phantoms (20 – 40 cm) with iodine concentrations ranging from 1 to 19.7 mg/ml and XCAT human models with iodine contrast at BMI of 19 to 38 kg/m<sup>2</sup> across different radiation dose levels (13.9, 27.8, and 41.7 mGy of CTDI<sub>vol</sub>). We performed material decomposition, reconstructed iodine CT images, and evaluated iodine quantification accuracy.</div></div><div><h3>Results</h3><div>The Si-PCCT simulator closely matched with the prototype, with differences within 3 % in MTF (f<sub>50</sub> and f<sub>10</sub>) and 3.7 % (f<sub>peak</sub>) in NNPS, and Root-Mean-Square Error of 0.12 mg/ml in iodine quantification. The mean absolute errors (MAE) between the estimated and ground-truth iodine concentration were 0.10, 0.25, and 1.80 mg/ml for 20, 30, and 40 cm phantoms, and 0.31, 0.37, and 0.70 mg/ml for XCAT human models with BMIs of 19, 28, and 38 kg/m<sup>2</sup>, respectively. Similarly, the MAEs were 0.88, 0.45, and 0.31 mg/ml for the geometrical phantoms, and 0.66, 0.5, and 0.46 mg/ml for human models at CTDI<sub>vol</sub> of 13.9, 27.8, and 41.7 mGy respectively. These results demonstrate accurate iodine quantification performance, influenced by object size and radiation dose.</div></div><div><h3>Conclusion</h3><div>This study shows the promising clinical utility of Si-PCCT for accurate iodine quantification under clinically relevant imaging conditions.</div></div>","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":"135 ","pages":"Article 105003"},"PeriodicalIF":2.7000,"publicationDate":"2025-06-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Iodine quantification performance with deep silicon-based Photon-Counting CT: A virtual imaging trial study\",\"authors\":\"Raj Kumar Panta ,&nbsp;Zhye Yin ,&nbsp;Fredrik Grönberg ,&nbsp;Mridul Bhattarai ,&nbsp;Ehsan Abadi ,&nbsp;Paul Segars ,&nbsp;Ehsan Samei\",\"doi\":\"10.1016/j.ejmp.2025.105003\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Purpose</h3><div>This study investigates the imaging performance of a deep silicon-based photon-counting CT (Si-PCCT) in quantifying iodine contrast through a virtual imaging trial (VIT).</div></div><div><h3>Methods</h3><div>We developed a VIT framework using Si-PCCT simulator and benchmarked it against a prototype using an ACR phantom for assessing spatial resolution and noise characteristics, and a geometric phantom for iodine quantification. We imaged geometrical phantoms (20 – 40 cm) with iodine concentrations ranging from 1 to 19.7 mg/ml and XCAT human models with iodine contrast at BMI of 19 to 38 kg/m<sup>2</sup> across different radiation dose levels (13.9, 27.8, and 41.7 mGy of CTDI<sub>vol</sub>). We performed material decomposition, reconstructed iodine CT images, and evaluated iodine quantification accuracy.</div></div><div><h3>Results</h3><div>The Si-PCCT simulator closely matched with the prototype, with differences within 3 % in MTF (f<sub>50</sub> and f<sub>10</sub>) and 3.7 % (f<sub>peak</sub>) in NNPS, and Root-Mean-Square Error of 0.12 mg/ml in iodine quantification. The mean absolute errors (MAE) between the estimated and ground-truth iodine concentration were 0.10, 0.25, and 1.80 mg/ml for 20, 30, and 40 cm phantoms, and 0.31, 0.37, and 0.70 mg/ml for XCAT human models with BMIs of 19, 28, and 38 kg/m<sup>2</sup>, respectively. Similarly, the MAEs were 0.88, 0.45, and 0.31 mg/ml for the geometrical phantoms, and 0.66, 0.5, and 0.46 mg/ml for human models at CTDI<sub>vol</sub> of 13.9, 27.8, and 41.7 mGy respectively. These results demonstrate accurate iodine quantification performance, influenced by object size and radiation dose.</div></div><div><h3>Conclusion</h3><div>This study shows the promising clinical utility of Si-PCCT for accurate iodine quantification under clinically relevant imaging conditions.</div></div>\",\"PeriodicalId\":56092,\"journal\":{\"name\":\"Physica Medica-European Journal of Medical Physics\",\"volume\":\"135 \",\"pages\":\"Article 105003\"},\"PeriodicalIF\":2.7000,\"publicationDate\":\"2025-06-06\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Physica Medica-European Journal of Medical Physics\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1120179725001139\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physica Medica-European Journal of Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1120179725001139","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

目的研究深硅基光子计数CT (Si-PCCT)通过虚拟成像试验(VIT)定量碘造影剂的成像性能。方法使用Si-PCCT模拟器开发了一个VIT框架,并使用ACR模体来评估空间分辨率和噪声特性,使用几何模体来评估碘定量。在不同的辐射剂量水平(13.9、27.8和41.7 mGy的CTDIvol)下,我们对碘浓度范围为1至19.7 mg/ml的几何模型(20 - 40 cm)和BMI为19至38 kg/m2的XCAT人体模型进行了成像。我们进行了材料分解,重建了碘CT图像,并评估了碘定量的准确性。结果Si-PCCT模拟器与样品吻合较好,MTF (f50和f10)和NNPS (f峰)差异在3%以内,碘定量均方根误差为0.12 mg/ml。对于20、30和40 cm的模型,估计的碘浓度和真实碘浓度之间的平均绝对误差(MAE)分别为0.10、0.25和1.80 mg/ml,对于bmi为19、28和38 kg/m2的XCAT模型,估计的碘浓度和真实碘浓度之间的平均绝对误差(MAE)分别为0.31、0.37和0.70 mg/ml。同样,在CTDIvol为13.9、27.8和41.7 mGy时,几何模型的MAEs分别为0.88、0.45和0.31 mg/ml,人体模型的MAEs分别为0.66、0.5和0.46 mg/ml。这些结果证明了碘定量的准确性,受物体大小和辐射剂量的影响。结论本研究显示Si-PCCT在临床相关成像条件下用于准确碘定量的临床应用前景广阔。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Iodine quantification performance with deep silicon-based Photon-Counting CT: A virtual imaging trial study

Purpose

This study investigates the imaging performance of a deep silicon-based photon-counting CT (Si-PCCT) in quantifying iodine contrast through a virtual imaging trial (VIT).

Methods

We developed a VIT framework using Si-PCCT simulator and benchmarked it against a prototype using an ACR phantom for assessing spatial resolution and noise characteristics, and a geometric phantom for iodine quantification. We imaged geometrical phantoms (20 – 40 cm) with iodine concentrations ranging from 1 to 19.7 mg/ml and XCAT human models with iodine contrast at BMI of 19 to 38 kg/m2 across different radiation dose levels (13.9, 27.8, and 41.7 mGy of CTDIvol). We performed material decomposition, reconstructed iodine CT images, and evaluated iodine quantification accuracy.

Results

The Si-PCCT simulator closely matched with the prototype, with differences within 3 % in MTF (f50 and f10) and 3.7 % (fpeak) in NNPS, and Root-Mean-Square Error of 0.12 mg/ml in iodine quantification. The mean absolute errors (MAE) between the estimated and ground-truth iodine concentration were 0.10, 0.25, and 1.80 mg/ml for 20, 30, and 40 cm phantoms, and 0.31, 0.37, and 0.70 mg/ml for XCAT human models with BMIs of 19, 28, and 38 kg/m2, respectively. Similarly, the MAEs were 0.88, 0.45, and 0.31 mg/ml for the geometrical phantoms, and 0.66, 0.5, and 0.46 mg/ml for human models at CTDIvol of 13.9, 27.8, and 41.7 mGy respectively. These results demonstrate accurate iodine quantification performance, influenced by object size and radiation dose.

Conclusion

This study shows the promising clinical utility of Si-PCCT for accurate iodine quantification under clinically relevant imaging conditions.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
6.80
自引率
14.70%
发文量
493
审稿时长
78 days
期刊介绍: Physica Medica, European Journal of Medical Physics, publishing with Elsevier from 2007, provides an international forum for research and reviews on the following main topics: Medical Imaging Radiation Therapy Radiation Protection Measuring Systems and Signal Processing Education and training in Medical Physics Professional issues in Medical Physics.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信