基于同位素稀释-液相色谱-串联质谱的候选参考测量方法定量人血清和血浆中17β-雌二醇。

IF 3.8 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY
Myriam Ott, Tobias Santner, Neeraj Singh, Friederike Bauland, Daniel Köppl, Alexander Gaudl, Andrea Geistanger, Uta Ceglarek, Manfred Rauh, Christian Geletneky, Judith Taibon
{"title":"基于同位素稀释-液相色谱-串联质谱的候选参考测量方法定量人血清和血浆中17β-雌二醇。","authors":"Myriam Ott, Tobias Santner, Neeraj Singh, Friederike Bauland, Daniel Köppl, Alexander Gaudl, Andrea Geistanger, Uta Ceglarek, Manfred Rauh, Christian Geletneky, Judith Taibon","doi":"10.1515/cclm-2024-1255","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>A new candidate isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based reference measurement procedure (RMP) has been developed for the accurate and precise quantification of 17β-estradiol (E2) in human serum and plasma covering a measurement range from 0.400 to 5,000 pg/mL (1.47-18,357 pmol/L). To address this broad range, two separate methods were created: a high sensitivity (HS) method for concentrations between 0.400 and 5.00 pg/mL (1.47-18.4 pmol/L) and a standard range (SR) method for concentrations between 5.00 and 5,000 pg/mL (18.4-18,357 pmol/L).</p><p><strong>Methods: </strong>As the primary reference material, E2 (CRM 6004-a) from the National Metrology Institute of Japan was used to ensure traceability to the international system (SI). A two-dimensional heart-cut LC approach was utilized for LC-MS/MS analysis, employing a supported liquid extraction sample preparation protocol for the SR method and a liquid-liquid extraction protocol followed by derivatization for the HS method. Assay validation was conducted following current guidelines. Selectivity and specificity were assessed using spiked serum samples, while potential matrix effects were evaluated through a post-column infusion experiment and comparison of standard line slopes. Precision, accuracy, and trueness were determined using an extensive 5-day protocol. Standard measurement uncertainty was evaluated according to the Guide to the Expression of Uncertainty in Measurement (GUM), with three individual sample preparations performed on at least two different days. Equivalence with higher-order RMPs was demonstrated through participation in the CDC Steroid Hormones Standardization (HoSt) program.</p><p><strong>Results: </strong>The RMP enabled the quantification of E2 within the range of 0.400-5,000 pg/mL (1.47-18,357 pmol/L), demonstrating no interference from structurally related compounds and no evidence of matrix effects. The relative mean bias of the SR method ranged from -2.4 to 1.9 % across all levels, including secondary reference materials and spiked samples, whereas the HS method exhibited a mean bias ranging from -3.0 to 2.9 %. Expanded measurement uncertainties (k=2) for target value assignment ranged from 1.7 to 4.4 % for the SR method and were found to be ≤6.1 % for the HS method. The method's transferability was demonstrated in a comparison study at a second laboratory. Additionally, the candidate RMP exhibited excellent correlation and equivalence to JCTLM-listed RMPs through the CDC HoSt program.</p><p><strong>Conclusions: </strong>In summary, the ID-LC-MS/MS-based RMP accurately quantifies E2. Its robust performance makes it suitable for standardizing routine assays and measuring individual patient samples, ensuring traceability.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":""},"PeriodicalIF":3.8000,"publicationDate":"2025-05-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"An isotope dilution-liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedure for the quantification of 17β-estradiol in human serum and plasma.\",\"authors\":\"Myriam Ott, Tobias Santner, Neeraj Singh, Friederike Bauland, Daniel Köppl, Alexander Gaudl, Andrea Geistanger, Uta Ceglarek, Manfred Rauh, Christian Geletneky, Judith Taibon\",\"doi\":\"10.1515/cclm-2024-1255\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objectives: </strong>A new candidate isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based reference measurement procedure (RMP) has been developed for the accurate and precise quantification of 17β-estradiol (E2) in human serum and plasma covering a measurement range from 0.400 to 5,000 pg/mL (1.47-18,357 pmol/L). To address this broad range, two separate methods were created: a high sensitivity (HS) method for concentrations between 0.400 and 5.00 pg/mL (1.47-18.4 pmol/L) and a standard range (SR) method for concentrations between 5.00 and 5,000 pg/mL (18.4-18,357 pmol/L).</p><p><strong>Methods: </strong>As the primary reference material, E2 (CRM 6004-a) from the National Metrology Institute of Japan was used to ensure traceability to the international system (SI). A two-dimensional heart-cut LC approach was utilized for LC-MS/MS analysis, employing a supported liquid extraction sample preparation protocol for the SR method and a liquid-liquid extraction protocol followed by derivatization for the HS method. Assay validation was conducted following current guidelines. Selectivity and specificity were assessed using spiked serum samples, while potential matrix effects were evaluated through a post-column infusion experiment and comparison of standard line slopes. Precision, accuracy, and trueness were determined using an extensive 5-day protocol. Standard measurement uncertainty was evaluated according to the Guide to the Expression of Uncertainty in Measurement (GUM), with three individual sample preparations performed on at least two different days. Equivalence with higher-order RMPs was demonstrated through participation in the CDC Steroid Hormones Standardization (HoSt) program.</p><p><strong>Results: </strong>The RMP enabled the quantification of E2 within the range of 0.400-5,000 pg/mL (1.47-18,357 pmol/L), demonstrating no interference from structurally related compounds and no evidence of matrix effects. The relative mean bias of the SR method ranged from -2.4 to 1.9 % across all levels, including secondary reference materials and spiked samples, whereas the HS method exhibited a mean bias ranging from -3.0 to 2.9 %. Expanded measurement uncertainties (k=2) for target value assignment ranged from 1.7 to 4.4 % for the SR method and were found to be ≤6.1 % for the HS method. The method's transferability was demonstrated in a comparison study at a second laboratory. Additionally, the candidate RMP exhibited excellent correlation and equivalence to JCTLM-listed RMPs through the CDC HoSt program.</p><p><strong>Conclusions: </strong>In summary, the ID-LC-MS/MS-based RMP accurately quantifies E2. Its robust performance makes it suitable for standardizing routine assays and measuring individual patient samples, ensuring traceability.</p>\",\"PeriodicalId\":10390,\"journal\":{\"name\":\"Clinical chemistry and laboratory medicine\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":3.8000,\"publicationDate\":\"2025-05-26\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Clinical chemistry and laboratory medicine\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1515/cclm-2024-1255\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical chemistry and laboratory medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1515/cclm-2024-1255","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的:建立了一种新的候选同位素稀释-液相色谱-串联质谱(ID-LC-MS/MS)参考测量方法(RMP),用于准确定量人血清和血浆中17β-雌二醇(E2)的含量,测量范围为0.400 ~ 5000 pg/mL(1.47 ~ 18357 pmol/L)。为了解决这个广泛的范围,创建了两种不同的方法:浓度在0.400至5.00 pg/mL(1.47至18.4 pmol/L)之间的高灵敏度(HS)方法和浓度在5.00至5000 pg/mL(18.4至18357 pmol/L)之间的标准范围(SR)方法。方法:采用日本国家计量所的E2 (CRM 6004-a)作为主要标准物质,确保其可溯源至国际单位制(SI)。LC-MS/MS分析采用二维心形切割LC方法,SR法采用支撑液萃取制样方案,HS法采用液液萃取衍生化方案。试验验证按照现行指南进行。使用加标血清样品评估选择性和特异性,而通过柱后输注实验和标准线斜率比较评估潜在的基质效应。精密度、准确度和真实性采用广泛的5天方案来确定。根据测量不确定度表达指南(GUM)评估标准测量不确定度,在至少两个不同的日子进行三个单独的样品制备。通过参与CDC类固醇激素标准化(HoSt)项目,证明了与高阶RMPs的等效性。结果:RMP可在0.400- 5000 pg/mL(1.47- 18357 pmol/L)范围内定量E2,无结构相关化合物干扰,无基质效应。SR方法的相对平均偏差范围为-2.4 ~ 1.9 %,而HS方法的平均偏差范围为-3.0 ~ 2.9 %。SR法的目标值分配的扩展测量不确定度(k=2)范围为1.7 ~ 4.4 %,HS法的不确定度≤6.1 %。该方法的可移植性在第二个实验室的比较研究中得到了证明。此外,通过CDC HoSt程序,候选RMP与jctlm列出的RMP表现出良好的相关性和等效性。结论:综上所述,基于ID-LC-MS/ ms的RMP能准确定量E2。其强大的性能使其适用于标准化常规分析和测量个体患者样本,确保可追溯性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
An isotope dilution-liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedure for the quantification of 17β-estradiol in human serum and plasma.

Objectives: A new candidate isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based reference measurement procedure (RMP) has been developed for the accurate and precise quantification of 17β-estradiol (E2) in human serum and plasma covering a measurement range from 0.400 to 5,000 pg/mL (1.47-18,357 pmol/L). To address this broad range, two separate methods were created: a high sensitivity (HS) method for concentrations between 0.400 and 5.00 pg/mL (1.47-18.4 pmol/L) and a standard range (SR) method for concentrations between 5.00 and 5,000 pg/mL (18.4-18,357 pmol/L).

Methods: As the primary reference material, E2 (CRM 6004-a) from the National Metrology Institute of Japan was used to ensure traceability to the international system (SI). A two-dimensional heart-cut LC approach was utilized for LC-MS/MS analysis, employing a supported liquid extraction sample preparation protocol for the SR method and a liquid-liquid extraction protocol followed by derivatization for the HS method. Assay validation was conducted following current guidelines. Selectivity and specificity were assessed using spiked serum samples, while potential matrix effects were evaluated through a post-column infusion experiment and comparison of standard line slopes. Precision, accuracy, and trueness were determined using an extensive 5-day protocol. Standard measurement uncertainty was evaluated according to the Guide to the Expression of Uncertainty in Measurement (GUM), with three individual sample preparations performed on at least two different days. Equivalence with higher-order RMPs was demonstrated through participation in the CDC Steroid Hormones Standardization (HoSt) program.

Results: The RMP enabled the quantification of E2 within the range of 0.400-5,000 pg/mL (1.47-18,357 pmol/L), demonstrating no interference from structurally related compounds and no evidence of matrix effects. The relative mean bias of the SR method ranged from -2.4 to 1.9 % across all levels, including secondary reference materials and spiked samples, whereas the HS method exhibited a mean bias ranging from -3.0 to 2.9 %. Expanded measurement uncertainties (k=2) for target value assignment ranged from 1.7 to 4.4 % for the SR method and were found to be ≤6.1 % for the HS method. The method's transferability was demonstrated in a comparison study at a second laboratory. Additionally, the candidate RMP exhibited excellent correlation and equivalence to JCTLM-listed RMPs through the CDC HoSt program.

Conclusions: In summary, the ID-LC-MS/MS-based RMP accurately quantifies E2. Its robust performance makes it suitable for standardizing routine assays and measuring individual patient samples, ensuring traceability.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Clinical chemistry and laboratory medicine
Clinical chemistry and laboratory medicine 医学-医学实验技术
CiteScore
11.30
自引率
16.20%
发文量
306
审稿时长
3 months
期刊介绍: Clinical Chemistry and Laboratory Medicine (CCLM) publishes articles on novel teaching and training methods applicable to laboratory medicine. CCLM welcomes contributions on the progress in fundamental and applied research and cutting-edge clinical laboratory medicine. It is one of the leading journals in the field, with an impact factor over 3. CCLM is issued monthly, and it is published in print and electronically. CCLM is the official journal of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) and publishes regularly EFLM recommendations and news. CCLM is the official journal of the National Societies from Austria (ÖGLMKC); Belgium (RBSLM); Germany (DGKL); Hungary (MLDT); Ireland (ACBI); Italy (SIBioC); Portugal (SPML); and Slovenia (SZKK); and it is affiliated to AACB (Australia) and SFBC (France). Topics: - clinical biochemistry - clinical genomics and molecular biology - clinical haematology and coagulation - clinical immunology and autoimmunity - clinical microbiology - drug monitoring and analysis - evaluation of diagnostic biomarkers - disease-oriented topics (cardiovascular disease, cancer diagnostics, diabetes) - new reagents, instrumentation and technologies - new methodologies - reference materials and methods - reference values and decision limits - quality and safety in laboratory medicine - translational laboratory medicine - clinical metrology Follow @cclm_degruyter on Twitter!
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信