{"title":"一种清除下气道分泌物的新装置的评价。","authors":"Luigi Vivona, Fabiana Madotto, Eleonora Carlesso, Gaetano Florio, Stephen E Rees, Dan S Karbing, Federico Sodi, Irene Zainaghi, Veronica Rossi, Mauro Panigada, Alberto Zanella, Giacomo Grasselli","doi":"10.1089/respcare.12637","DOIUrl":null,"url":null,"abstract":"<p><p><b>Background:</b> Endotracheal intubation promotes the retention of secretions in the lower airways, potentially leading to infections and obstructions. A novel endotracheal cuff controller, synchronized with the ventilator, is designed to remove secretions from both above and below the endotracheal tube (ETT) cuff by generating a semi-automatic \"artificial cough\" through brief ETT cuff deflation during the inspiratory phase of a ventilator breath. This study assessed, first, in vitro, the accuracy of a computation for estimating the cough flow and, subsequently, in subjects, the device's effectiveness in moving secretions from lower airways toward the mouth during invasive ventilation. <b>Methods:</b> In the in vitro study, the mean cough flow estimated by the ratio between the expiratory volume lost and the duration of the cuff deflation was compared with the actual cough flow measured using flow meters. In the clinical study, we enrolled subjects admitted to the ICU. Three artificial cough maneuvers were performed within 3 min during a sigh (inspiratory pressure of 35 cm H<sub>2</sub>O for ∼1.8 s). Aspiration of the oral cavity was conducted before the first maneuver and after the third maneuver. <b>Results:</b> A strong linear association was found between actual and computed cough flow (Pearson's correlation coefficient 0.991, 95% CI: 0.987-0.994). Twenty-eight subjects were enrolled. After device activation, oral secretions were detected in 78.6% of subjects (responders). Cough flow was similar in both responders and nonresponders, with a mean highest value of 23.3 ± 11.5 L/min. Demographic, physiological, and ventilatory parameters before and after the maneuvers were also similar. No relevant adverse events were recorded, except for one transient episode of hypotension. <b>Conclusions:</b> The cough flow can be accurately estimated from the expiratory volume lost. The device successfully generated effective cough flows in all subjects, clearing secretions from the trachea to the mouth in nearly 80% of cases.</p>","PeriodicalId":21125,"journal":{"name":"Respiratory care","volume":" ","pages":"1110-1119"},"PeriodicalIF":2.1000,"publicationDate":"2025-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Evaluation of a New Device for Clearing Secretions From Lower Airways.\",\"authors\":\"Luigi Vivona, Fabiana Madotto, Eleonora Carlesso, Gaetano Florio, Stephen E Rees, Dan S Karbing, Federico Sodi, Irene Zainaghi, Veronica Rossi, Mauro Panigada, Alberto Zanella, Giacomo Grasselli\",\"doi\":\"10.1089/respcare.12637\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p><b>Background:</b> Endotracheal intubation promotes the retention of secretions in the lower airways, potentially leading to infections and obstructions. A novel endotracheal cuff controller, synchronized with the ventilator, is designed to remove secretions from both above and below the endotracheal tube (ETT) cuff by generating a semi-automatic \\\"artificial cough\\\" through brief ETT cuff deflation during the inspiratory phase of a ventilator breath. This study assessed, first, in vitro, the accuracy of a computation for estimating the cough flow and, subsequently, in subjects, the device's effectiveness in moving secretions from lower airways toward the mouth during invasive ventilation. <b>Methods:</b> In the in vitro study, the mean cough flow estimated by the ratio between the expiratory volume lost and the duration of the cuff deflation was compared with the actual cough flow measured using flow meters. In the clinical study, we enrolled subjects admitted to the ICU. Three artificial cough maneuvers were performed within 3 min during a sigh (inspiratory pressure of 35 cm H<sub>2</sub>O for ∼1.8 s). Aspiration of the oral cavity was conducted before the first maneuver and after the third maneuver. <b>Results:</b> A strong linear association was found between actual and computed cough flow (Pearson's correlation coefficient 0.991, 95% CI: 0.987-0.994). Twenty-eight subjects were enrolled. After device activation, oral secretions were detected in 78.6% of subjects (responders). Cough flow was similar in both responders and nonresponders, with a mean highest value of 23.3 ± 11.5 L/min. Demographic, physiological, and ventilatory parameters before and after the maneuvers were also similar. No relevant adverse events were recorded, except for one transient episode of hypotension. <b>Conclusions:</b> The cough flow can be accurately estimated from the expiratory volume lost. The device successfully generated effective cough flows in all subjects, clearing secretions from the trachea to the mouth in nearly 80% of cases.</p>\",\"PeriodicalId\":21125,\"journal\":{\"name\":\"Respiratory care\",\"volume\":\" \",\"pages\":\"1110-1119\"},\"PeriodicalIF\":2.1000,\"publicationDate\":\"2025-09-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Respiratory care\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1089/respcare.12637\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2025/4/28 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"CRITICAL CARE MEDICINE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Respiratory care","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1089/respcare.12637","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/4/28 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"CRITICAL CARE MEDICINE","Score":null,"Total":0}
引用次数: 0
摘要
背景:气管插管促进下气道分泌物潴留,可能导致感染和阻塞。一种新型的气管内袖带控制器,与呼吸机同步,通过在呼吸机呼吸的吸气阶段短暂的气管内袖带放气产生半自动的“人工咳嗽”,以去除气管内管(ETT)袖带上下的分泌物。本研究首先在体外评估了估算咳嗽流量计算的准确性,随后在受试者中评估了该装置在有创通气期间将分泌物从下气道移向口腔的有效性。方法:在体外研究中,通过呼气量损失与袖带收缩时间之比估计的平均咳嗽流量与用流量计测量的实际咳嗽流量进行比较。在临床研究中,我们招募了入住ICU的受试者。在叹气期间的3分钟内进行3次人工咳嗽动作(吸气压力为35 cm H2O,持续约1.8 s)。在第一次手术前和第三次手术后分别进行口腔吸痰。结果:实际咳嗽流量与计算咳嗽流量之间存在较强的线性关联(Pearson相关系数0.991,95% CI: 0.987 ~ 0.994)。共纳入28名受试者。在装置激活后,78.6%的受试者(应答者)检测到口腔分泌物。应答者和无应答者的咳嗽流相似,平均最高值为23.3±11.5 L/min。机动前后的人口学、生理和通气参数也相似。除一次短暂性低血压发作外,无相关不良事件记录。结论:通过呼气量的减少可以准确地估计咳嗽流量。该装置成功地在所有受试者中产生了有效的咳嗽流,在近80%的病例中清除了从气管到口腔的分泌物。
Evaluation of a New Device for Clearing Secretions From Lower Airways.
Background: Endotracheal intubation promotes the retention of secretions in the lower airways, potentially leading to infections and obstructions. A novel endotracheal cuff controller, synchronized with the ventilator, is designed to remove secretions from both above and below the endotracheal tube (ETT) cuff by generating a semi-automatic "artificial cough" through brief ETT cuff deflation during the inspiratory phase of a ventilator breath. This study assessed, first, in vitro, the accuracy of a computation for estimating the cough flow and, subsequently, in subjects, the device's effectiveness in moving secretions from lower airways toward the mouth during invasive ventilation. Methods: In the in vitro study, the mean cough flow estimated by the ratio between the expiratory volume lost and the duration of the cuff deflation was compared with the actual cough flow measured using flow meters. In the clinical study, we enrolled subjects admitted to the ICU. Three artificial cough maneuvers were performed within 3 min during a sigh (inspiratory pressure of 35 cm H2O for ∼1.8 s). Aspiration of the oral cavity was conducted before the first maneuver and after the third maneuver. Results: A strong linear association was found between actual and computed cough flow (Pearson's correlation coefficient 0.991, 95% CI: 0.987-0.994). Twenty-eight subjects were enrolled. After device activation, oral secretions were detected in 78.6% of subjects (responders). Cough flow was similar in both responders and nonresponders, with a mean highest value of 23.3 ± 11.5 L/min. Demographic, physiological, and ventilatory parameters before and after the maneuvers were also similar. No relevant adverse events were recorded, except for one transient episode of hypotension. Conclusions: The cough flow can be accurately estimated from the expiratory volume lost. The device successfully generated effective cough flows in all subjects, clearing secretions from the trachea to the mouth in nearly 80% of cases.
期刊介绍:
RESPIRATORY CARE is the official monthly science journal of the American Association for Respiratory Care. It is indexed in PubMed and included in ISI''s Web of Science.