{"title":"[髋关节假体摩擦节点模拟器测试]。","authors":"G P Kotelnikov, A N Nikolaenko, V V Ivanov, S O Doroganov, I M Mikailov, A P Borisov","doi":"10.17116/hirurgia202505172","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To estimate advisability of prototypes of ultra-high molecular weight polyethylene (UHMWPE) in hip arthroplasty.</p><p><strong>Material and methods: </strong>We used 34 UHMWPE specimens for tests and developed a test program consisting of incoming control of specimen geometrical parameters, tribotechnical tests, assessment of liner surface condition and deviation from roundness.</p><p><strong>Results: </strong>Input control of geometrical parameters of specimens showed their heterogeneity. UHMWPE specimens with negative value <i>(D</i><sub>c</sub><i>-D</i><sub>h</sub><i>)</i> failed tribotechnical tests. Specimens with value <i>(D</i><sub>c</sub><i>-D</i><sub>h</sub><i>)</i> within 0.6-0.8 mm demonstrated lower value of <i>T</i> moment compared to other specimens.</p><p><strong>Conclusion: </strong>About 75% of UHMWPE prototype liners are suitable for arthroplasty.</p>","PeriodicalId":35986,"journal":{"name":"Khirurgiya","volume":" 5","pages":"72-78"},"PeriodicalIF":0.0000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Testing of friction node simulators for hip prostheses].\",\"authors\":\"G P Kotelnikov, A N Nikolaenko, V V Ivanov, S O Doroganov, I M Mikailov, A P Borisov\",\"doi\":\"10.17116/hirurgia202505172\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>To estimate advisability of prototypes of ultra-high molecular weight polyethylene (UHMWPE) in hip arthroplasty.</p><p><strong>Material and methods: </strong>We used 34 UHMWPE specimens for tests and developed a test program consisting of incoming control of specimen geometrical parameters, tribotechnical tests, assessment of liner surface condition and deviation from roundness.</p><p><strong>Results: </strong>Input control of geometrical parameters of specimens showed their heterogeneity. UHMWPE specimens with negative value <i>(D</i><sub>c</sub><i>-D</i><sub>h</sub><i>)</i> failed tribotechnical tests. Specimens with value <i>(D</i><sub>c</sub><i>-D</i><sub>h</sub><i>)</i> within 0.6-0.8 mm demonstrated lower value of <i>T</i> moment compared to other specimens.</p><p><strong>Conclusion: </strong>About 75% of UHMWPE prototype liners are suitable for arthroplasty.</p>\",\"PeriodicalId\":35986,\"journal\":{\"name\":\"Khirurgiya\",\"volume\":\" 5\",\"pages\":\"72-78\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2025-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Khirurgiya\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.17116/hirurgia202505172\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Khirurgiya","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.17116/hirurgia202505172","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

目的:评价超高分子量聚乙烯(UHMWPE)原型材料在髋关节置换术中的可行性。材料和方法:我们使用了34个超高分子量聚乙烯试样进行测试,并制定了一个测试程序,包括试样几何参数的输入控制、摩擦技术测试、衬垫表面状况和圆度偏差的评估。结果:标本几何参数输入控制呈现异质性。具有负值(Dc-Dh)的超高分子量聚乙烯试样摩擦学试验失败。值(Dc-Dh)在0.6 ~ 0.8 mm范围内的试件T矩值较其他试件低。结论:75%的UHMWPE原型衬垫适合于关节置换术。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
[Testing of friction node simulators for hip prostheses].

Objective: To estimate advisability of prototypes of ultra-high molecular weight polyethylene (UHMWPE) in hip arthroplasty.

Material and methods: We used 34 UHMWPE specimens for tests and developed a test program consisting of incoming control of specimen geometrical parameters, tribotechnical tests, assessment of liner surface condition and deviation from roundness.

Results: Input control of geometrical parameters of specimens showed their heterogeneity. UHMWPE specimens with negative value (Dc-Dh) failed tribotechnical tests. Specimens with value (Dc-Dh) within 0.6-0.8 mm demonstrated lower value of T moment compared to other specimens.

Conclusion: About 75% of UHMWPE prototype liners are suitable for arthroplasty.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Khirurgiya
Khirurgiya Medicine-Medicine (all)
CiteScore
0.70
自引率
0.00%
发文量
161
期刊介绍: Хирургия отдельных областей сердце, сосуды легкие пищевод молочная железа желудок и двенадцатиперстная кишка кишечник желчевыводящие пути печень
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信