G P Kotelnikov, A N Nikolaenko, V V Ivanov, S O Doroganov, I M Mikailov, A P Borisov
{"title":"[髋关节假体摩擦节点模拟器测试]。","authors":"G P Kotelnikov, A N Nikolaenko, V V Ivanov, S O Doroganov, I M Mikailov, A P Borisov","doi":"10.17116/hirurgia202505172","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To estimate advisability of prototypes of ultra-high molecular weight polyethylene (UHMWPE) in hip arthroplasty.</p><p><strong>Material and methods: </strong>We used 34 UHMWPE specimens for tests and developed a test program consisting of incoming control of specimen geometrical parameters, tribotechnical tests, assessment of liner surface condition and deviation from roundness.</p><p><strong>Results: </strong>Input control of geometrical parameters of specimens showed their heterogeneity. UHMWPE specimens with negative value <i>(D</i><sub>c</sub><i>-D</i><sub>h</sub><i>)</i> failed tribotechnical tests. Specimens with value <i>(D</i><sub>c</sub><i>-D</i><sub>h</sub><i>)</i> within 0.6-0.8 mm demonstrated lower value of <i>T</i> moment compared to other specimens.</p><p><strong>Conclusion: </strong>About 75% of UHMWPE prototype liners are suitable for arthroplasty.</p>","PeriodicalId":35986,"journal":{"name":"Khirurgiya","volume":" 5","pages":"72-78"},"PeriodicalIF":0.0000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Testing of friction node simulators for hip prostheses].\",\"authors\":\"G P Kotelnikov, A N Nikolaenko, V V Ivanov, S O Doroganov, I M Mikailov, A P Borisov\",\"doi\":\"10.17116/hirurgia202505172\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>To estimate advisability of prototypes of ultra-high molecular weight polyethylene (UHMWPE) in hip arthroplasty.</p><p><strong>Material and methods: </strong>We used 34 UHMWPE specimens for tests and developed a test program consisting of incoming control of specimen geometrical parameters, tribotechnical tests, assessment of liner surface condition and deviation from roundness.</p><p><strong>Results: </strong>Input control of geometrical parameters of specimens showed their heterogeneity. UHMWPE specimens with negative value <i>(D</i><sub>c</sub><i>-D</i><sub>h</sub><i>)</i> failed tribotechnical tests. Specimens with value <i>(D</i><sub>c</sub><i>-D</i><sub>h</sub><i>)</i> within 0.6-0.8 mm demonstrated lower value of <i>T</i> moment compared to other specimens.</p><p><strong>Conclusion: </strong>About 75% of UHMWPE prototype liners are suitable for arthroplasty.</p>\",\"PeriodicalId\":35986,\"journal\":{\"name\":\"Khirurgiya\",\"volume\":\" 5\",\"pages\":\"72-78\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2025-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Khirurgiya\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.17116/hirurgia202505172\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Khirurgiya","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.17116/hirurgia202505172","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
[Testing of friction node simulators for hip prostheses].
Objective: To estimate advisability of prototypes of ultra-high molecular weight polyethylene (UHMWPE) in hip arthroplasty.
Material and methods: We used 34 UHMWPE specimens for tests and developed a test program consisting of incoming control of specimen geometrical parameters, tribotechnical tests, assessment of liner surface condition and deviation from roundness.
Results: Input control of geometrical parameters of specimens showed their heterogeneity. UHMWPE specimens with negative value (Dc-Dh) failed tribotechnical tests. Specimens with value (Dc-Dh) within 0.6-0.8 mm demonstrated lower value of T moment compared to other specimens.
Conclusion: About 75% of UHMWPE prototype liners are suitable for arthroplasty.