Arnaud Deschavannes, Kazuma Terashi, Marie Piquemal, Catherine Rioufol, Anthony Clotagatide
{"title":"[68Ga]Ga-EDOTREOTIDE (Somakit-TOC®)放射化学纯度评估优化:缩短r-TLC方法改进PET放射制药工作流程","authors":"Arnaud Deschavannes, Kazuma Terashi, Marie Piquemal, Catherine Rioufol, Anthony Clotagatide","doi":"10.1186/s41181-025-00341-y","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>The increasing use of [⁶⁸Ga]Ga-based radiopharmaceuticals in PET imaging, requires efficient quality control procedures. The standard r-TLC method for verifying [⁶⁸Ga]Ga-EDOTREOTIDE (Somakit-TOC<sup>®</sup>) radiochemical purity (RCP) is time-consuming, creating workflow challenges in radiopharmacies. This study evaluates an optimized r-TLC method with a reduced migration distance (4 cm vs. 9 cm) to improve efficiency while maintaining analytical reliability. Tests for specificity, accuracy and robustness were performed using ITLC-SG – Acetate and ITLC-SG – Citrate systems. Additionally, migration time was analyzed to evaluate whether the alternative method could offer added benefits.</p><h3>Results</h3><p>The mean Rs for ITLC-SG – Acetate at 4 cm was 2.43 ± 0.28, while for ITLC-SG – Citrate with added [⁶⁸Ga]GaCl₃ was 5.58 ± 0.23, both exceeding the threshold of 1.5. The mean RCP at 4 cm was 98.90% ± 0.25%, and 99.21% ± 0.19% at 9 cm, with [⁶⁸Ga]Ga-uncomplexed remaining within acceptable limits. No [⁶⁸Ga]GaCl₃ was detected. The coefficient of variation (CV) for RCP between methods was < 2% (0.22%). Operator-based analysis yielded a mean Rs of 3.95 ± 0.06 (CV = 1.52%) and a mean [⁶⁸Ga]Ga-EDOTREOTIDE percentage of 99.60% ± 0.03% (CV = 0.03%). Migration times were significantly reduced with the alternative method (85% reduction).</p><h3>Conclusion</h3><p>Shortening the migration distance in r-TLC did not compromise specificity, accuracy or robustness while significantly reducing analysis time. The proposed method enhances PET radiopharmaceutical workflows, allowing faster patient dose preparation without quality loss. This approach could be investigated to other [<sup>68</sup>Ga]Ga-labeled compounds, supporting improved clinical and research applications in nuclear medicine.</p></div>","PeriodicalId":534,"journal":{"name":"EJNMMI Radiopharmacy and Chemistry","volume":"10 1","pages":""},"PeriodicalIF":4.4000,"publicationDate":"2025-05-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ejnmmipharmchem.springeropen.com/counter/pdf/10.1186/s41181-025-00341-y","citationCount":"0","resultStr":"{\"title\":\"Optimization of radiochemical purity assessment for [68Ga]Ga-EDOTREOTIDE (Somakit-TOC®): a shortened r-TLC method for improved PET radiopharmaceutical workflow\",\"authors\":\"Arnaud Deschavannes, Kazuma Terashi, Marie Piquemal, Catherine Rioufol, Anthony Clotagatide\",\"doi\":\"10.1186/s41181-025-00341-y\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Background</h3><p>The increasing use of [⁶⁸Ga]Ga-based radiopharmaceuticals in PET imaging, requires efficient quality control procedures. The standard r-TLC method for verifying [⁶⁸Ga]Ga-EDOTREOTIDE (Somakit-TOC<sup>®</sup>) radiochemical purity (RCP) is time-consuming, creating workflow challenges in radiopharmacies. This study evaluates an optimized r-TLC method with a reduced migration distance (4 cm vs. 9 cm) to improve efficiency while maintaining analytical reliability. Tests for specificity, accuracy and robustness were performed using ITLC-SG – Acetate and ITLC-SG – Citrate systems. Additionally, migration time was analyzed to evaluate whether the alternative method could offer added benefits.</p><h3>Results</h3><p>The mean Rs for ITLC-SG – Acetate at 4 cm was 2.43 ± 0.28, while for ITLC-SG – Citrate with added [⁶⁸Ga]GaCl₃ was 5.58 ± 0.23, both exceeding the threshold of 1.5. The mean RCP at 4 cm was 98.90% ± 0.25%, and 99.21% ± 0.19% at 9 cm, with [⁶⁸Ga]Ga-uncomplexed remaining within acceptable limits. No [⁶⁸Ga]GaCl₃ was detected. The coefficient of variation (CV) for RCP between methods was < 2% (0.22%). Operator-based analysis yielded a mean Rs of 3.95 ± 0.06 (CV = 1.52%) and a mean [⁶⁸Ga]Ga-EDOTREOTIDE percentage of 99.60% ± 0.03% (CV = 0.03%). Migration times were significantly reduced with the alternative method (85% reduction).</p><h3>Conclusion</h3><p>Shortening the migration distance in r-TLC did not compromise specificity, accuracy or robustness while significantly reducing analysis time. The proposed method enhances PET radiopharmaceutical workflows, allowing faster patient dose preparation without quality loss. This approach could be investigated to other [<sup>68</sup>Ga]Ga-labeled compounds, supporting improved clinical and research applications in nuclear medicine.</p></div>\",\"PeriodicalId\":534,\"journal\":{\"name\":\"EJNMMI Radiopharmacy and Chemistry\",\"volume\":\"10 1\",\"pages\":\"\"},\"PeriodicalIF\":4.4000,\"publicationDate\":\"2025-05-13\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://ejnmmipharmchem.springeropen.com/counter/pdf/10.1186/s41181-025-00341-y\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"EJNMMI Radiopharmacy and Chemistry\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://link.springer.com/article/10.1186/s41181-025-00341-y\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"CHEMISTRY, INORGANIC & NUCLEAR\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"EJNMMI Radiopharmacy and Chemistry","FirstCategoryId":"1085","ListUrlMain":"https://link.springer.com/article/10.1186/s41181-025-00341-y","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CHEMISTRY, INORGANIC & NUCLEAR","Score":null,"Total":0}
Optimization of radiochemical purity assessment for [68Ga]Ga-EDOTREOTIDE (Somakit-TOC®): a shortened r-TLC method for improved PET radiopharmaceutical workflow
Background
The increasing use of [⁶⁸Ga]Ga-based radiopharmaceuticals in PET imaging, requires efficient quality control procedures. The standard r-TLC method for verifying [⁶⁸Ga]Ga-EDOTREOTIDE (Somakit-TOC®) radiochemical purity (RCP) is time-consuming, creating workflow challenges in radiopharmacies. This study evaluates an optimized r-TLC method with a reduced migration distance (4 cm vs. 9 cm) to improve efficiency while maintaining analytical reliability. Tests for specificity, accuracy and robustness were performed using ITLC-SG – Acetate and ITLC-SG – Citrate systems. Additionally, migration time was analyzed to evaluate whether the alternative method could offer added benefits.
Results
The mean Rs for ITLC-SG – Acetate at 4 cm was 2.43 ± 0.28, while for ITLC-SG – Citrate with added [⁶⁸Ga]GaCl₃ was 5.58 ± 0.23, both exceeding the threshold of 1.5. The mean RCP at 4 cm was 98.90% ± 0.25%, and 99.21% ± 0.19% at 9 cm, with [⁶⁸Ga]Ga-uncomplexed remaining within acceptable limits. No [⁶⁸Ga]GaCl₃ was detected. The coefficient of variation (CV) for RCP between methods was < 2% (0.22%). Operator-based analysis yielded a mean Rs of 3.95 ± 0.06 (CV = 1.52%) and a mean [⁶⁸Ga]Ga-EDOTREOTIDE percentage of 99.60% ± 0.03% (CV = 0.03%). Migration times were significantly reduced with the alternative method (85% reduction).
Conclusion
Shortening the migration distance in r-TLC did not compromise specificity, accuracy or robustness while significantly reducing analysis time. The proposed method enhances PET radiopharmaceutical workflows, allowing faster patient dose preparation without quality loss. This approach could be investigated to other [68Ga]Ga-labeled compounds, supporting improved clinical and research applications in nuclear medicine.