{"title":"一种适用于镗孔式直线机的通用焦斑测量方法的开发。","authors":"Hidetoshi Shimizu, Kazuharu Nishitani, Tomoki Kitagawa, Koji Sasaki, Takahiro Aoyama, Takeshi Kodaira","doi":"10.1002/acm2.70077","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>In linear accelerators, deviations in the x-ray focal spot position significantly affect the accuracy of radiation therapy. However, as the focal spot position in bore-type linac systems such as the Radixact system, cannot be assessed using conventional methods, a new evaluation method is required. This study aimed to develop a novel method to measure the focal spot position of Radixact and evaluate any deviations from the ideal x-ray focal spot position.</p><p><strong>Methods: </strong>A structurally simplified measurement system was developed to evaluate the focal spot position of the Radixact system. This system consisted of a vertically aligned metal bar and an ionization chamber, which was moved stepwise to acquire the beam profiles. The focal spot position deviation was calculated based on the center differences of the profiles obtained from two different upstream and downstream locations of the metal bar.</p><p><strong>Results: </strong>The measurement results indicated that the focal spot position shift was 0.42 mm and -0.36 mm at the target height in the IEC-X and -Y directions, respectively. The measurement uncertainty was 0.187 mm, confirming a slight deviation from the ideal focal position.</p><p><strong>Conclusions: </strong>This study developed a novel method to accurately evaluate the x-ray focal spot position of the Radixact system, which can potentially be applied to other conventional linear accelerators and bore-type systems, such as Halcyon, to improve the accuracy of radiotherapy. However, its generalizability and applicability to different radiotherapy machines must be explored further.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":" ","pages":"e70077"},"PeriodicalIF":2.0000,"publicationDate":"2025-04-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Development of a generic focal spot measurement method suitable for bore-type linacs.\",\"authors\":\"Hidetoshi Shimizu, Kazuharu Nishitani, Tomoki Kitagawa, Koji Sasaki, Takahiro Aoyama, Takeshi Kodaira\",\"doi\":\"10.1002/acm2.70077\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>In linear accelerators, deviations in the x-ray focal spot position significantly affect the accuracy of radiation therapy. However, as the focal spot position in bore-type linac systems such as the Radixact system, cannot be assessed using conventional methods, a new evaluation method is required. This study aimed to develop a novel method to measure the focal spot position of Radixact and evaluate any deviations from the ideal x-ray focal spot position.</p><p><strong>Methods: </strong>A structurally simplified measurement system was developed to evaluate the focal spot position of the Radixact system. This system consisted of a vertically aligned metal bar and an ionization chamber, which was moved stepwise to acquire the beam profiles. The focal spot position deviation was calculated based on the center differences of the profiles obtained from two different upstream and downstream locations of the metal bar.</p><p><strong>Results: </strong>The measurement results indicated that the focal spot position shift was 0.42 mm and -0.36 mm at the target height in the IEC-X and -Y directions, respectively. The measurement uncertainty was 0.187 mm, confirming a slight deviation from the ideal focal position.</p><p><strong>Conclusions: </strong>This study developed a novel method to accurately evaluate the x-ray focal spot position of the Radixact system, which can potentially be applied to other conventional linear accelerators and bore-type systems, such as Halcyon, to improve the accuracy of radiotherapy. However, its generalizability and applicability to different radiotherapy machines must be explored further.</p>\",\"PeriodicalId\":14989,\"journal\":{\"name\":\"Journal of Applied Clinical Medical Physics\",\"volume\":\" \",\"pages\":\"e70077\"},\"PeriodicalIF\":2.0000,\"publicationDate\":\"2025-04-07\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Applied Clinical Medical Physics\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1002/acm2.70077\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/acm2.70077","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

目的:在直线加速器中,x线焦斑位置的偏差严重影响放射治疗的准确性。然而,由于径向直线系统(Radixact)等孔型直线系统的焦点光斑位置无法用常规方法进行评估,因此需要一种新的评估方法。本研究旨在建立一种测量Radixact焦斑位置的新方法,并评估其与理想x射线焦斑位置的偏差。方法:开发了一种结构简化的测量系统,用于评估Radixact系统的焦点光斑位置。该系统由一个垂直排列的金属棒和一个电离室组成,电离室逐步移动以获得光束的轮廓。根据从金属棒的上下游两个不同位置获得的轮廓的中心差计算焦点光斑位置偏差。结果:测量结果表明,在IEC-X和-Y方向上,焦斑在目标高度处的位置位移分别为0.42 mm和-0.36 mm。测量不确定度为0.187 mm,与理想焦点位置有轻微偏差。结论:本研究开发了一种新的方法来准确评估Radixact系统的x射线焦斑位置,该方法有可能应用于其他传统的线性加速器和孔型系统,如Halcyon,以提高放疗的准确性。然而,其通用性和对不同放疗机的适用性还有待进一步探讨。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Development of a generic focal spot measurement method suitable for bore-type linacs.

Purpose: In linear accelerators, deviations in the x-ray focal spot position significantly affect the accuracy of radiation therapy. However, as the focal spot position in bore-type linac systems such as the Radixact system, cannot be assessed using conventional methods, a new evaluation method is required. This study aimed to develop a novel method to measure the focal spot position of Radixact and evaluate any deviations from the ideal x-ray focal spot position.

Methods: A structurally simplified measurement system was developed to evaluate the focal spot position of the Radixact system. This system consisted of a vertically aligned metal bar and an ionization chamber, which was moved stepwise to acquire the beam profiles. The focal spot position deviation was calculated based on the center differences of the profiles obtained from two different upstream and downstream locations of the metal bar.

Results: The measurement results indicated that the focal spot position shift was 0.42 mm and -0.36 mm at the target height in the IEC-X and -Y directions, respectively. The measurement uncertainty was 0.187 mm, confirming a slight deviation from the ideal focal position.

Conclusions: This study developed a novel method to accurately evaluate the x-ray focal spot position of the Radixact system, which can potentially be applied to other conventional linear accelerators and bore-type systems, such as Halcyon, to improve the accuracy of radiotherapy. However, its generalizability and applicability to different radiotherapy machines must be explored further.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
3.60
自引率
19.00%
发文量
331
审稿时长
3 months
期刊介绍: Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission. JACMP will publish: -Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500. -Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed. -Technical Notes: These should be no longer than 3000 words, including key references. -Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents. -Book Reviews: The editorial office solicits Book Reviews. -Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics. -Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信