腹腔镜和机器人手术中用于器官操作的PAF导轨和手术器械的比较:一项随机交叉可用性研究。

IF 4.4 2区 医学 Q2 ENGINEERING, BIOMEDICAL
A McDonald-Bowyer, L Caciolli, S El-Sheikh, J Opie, R Barod, D Stoyanov, A Stilli
{"title":"腹腔镜和机器人手术中用于器官操作的PAF导轨和手术器械的比较:一项随机交叉可用性研究。","authors":"A McDonald-Bowyer, L Caciolli, S El-Sheikh, J Opie, R Barod, D Stoyanov, A Stilli","doi":"10.1109/TBME.2025.3552900","DOIUrl":null,"url":null,"abstract":"<p><p>This study evaluates the usability and workload associated with the pneumatically attachable flexible (PAF) rail, a soft robotic device designed for safer organ manipulation and retraction during robotic-assisted partial nephrectomy (RAPN) and other laparoscopic procedures. Fourteen expert robotic and laparoscopic surgeons performed a simulated surgical retraction task using the PAF rail and standard surgical instruments. Usability was assessed using the System Usability Scale (SUS), and workload was measured with the NASA-TLX. Qualitative feedback was also collected to explore surgeon perceptions, and analysed thematically. Histopathological analysis was conducted to assess tissue integrity following instrument interaction. The PAF rail achieved SUS scores exceeding the good usability threshold, particularly among urology surgeons. However, its use was associated with increased cognitive load and longer task completion times, especially for less experienced surgeons. Histopathological analysis showed no additional tissue damage from the PAF rail compared to existing instruments. This study demonstrates that the PAF rail has potential as a safe and effective method for organ manipulation and retraction, achieving good usability and showing no additional tissue damage compared to existing instruments.</p>","PeriodicalId":13245,"journal":{"name":"IEEE Transactions on Biomedical Engineering","volume":"PP ","pages":""},"PeriodicalIF":4.4000,"publicationDate":"2025-03-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Comparison Between the PAF Rail and Surgical Instruments for Organ Manipulation in Laparoscopic and Robotic Surgery: A Randomised Cross-Over Usability Study.\",\"authors\":\"A McDonald-Bowyer, L Caciolli, S El-Sheikh, J Opie, R Barod, D Stoyanov, A Stilli\",\"doi\":\"10.1109/TBME.2025.3552900\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>This study evaluates the usability and workload associated with the pneumatically attachable flexible (PAF) rail, a soft robotic device designed for safer organ manipulation and retraction during robotic-assisted partial nephrectomy (RAPN) and other laparoscopic procedures. Fourteen expert robotic and laparoscopic surgeons performed a simulated surgical retraction task using the PAF rail and standard surgical instruments. Usability was assessed using the System Usability Scale (SUS), and workload was measured with the NASA-TLX. Qualitative feedback was also collected to explore surgeon perceptions, and analysed thematically. Histopathological analysis was conducted to assess tissue integrity following instrument interaction. The PAF rail achieved SUS scores exceeding the good usability threshold, particularly among urology surgeons. However, its use was associated with increased cognitive load and longer task completion times, especially for less experienced surgeons. Histopathological analysis showed no additional tissue damage from the PAF rail compared to existing instruments. This study demonstrates that the PAF rail has potential as a safe and effective method for organ manipulation and retraction, achieving good usability and showing no additional tissue damage compared to existing instruments.</p>\",\"PeriodicalId\":13245,\"journal\":{\"name\":\"IEEE Transactions on Biomedical Engineering\",\"volume\":\"PP \",\"pages\":\"\"},\"PeriodicalIF\":4.4000,\"publicationDate\":\"2025-03-28\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"IEEE Transactions on Biomedical Engineering\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1109/TBME.2025.3552900\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE Transactions on Biomedical Engineering","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1109/TBME.2025.3552900","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

摘要

本研究评估了气动可附柔性导轨(PAF)的可用性和工作量,PAF是一种软机器人设备,设计用于机器人辅助部分肾切除术(RAPN)和其他腹腔镜手术中更安全的器官操作和收缩。14位专家机器人和腹腔镜外科医生使用PAF导轨和标准手术器械进行了模拟手术牵回任务。使用系统可用性量表(SUS)评估可用性,使用NASA-TLX测量工作量。定性反馈也被收集来探讨外科医生的看法,并进行主题分析。进行组织病理学分析以评估仪器相互作用后的组织完整性。PAF导轨的SUS评分超过了良好的可用性阈值,特别是在泌尿外科医生中。然而,它的使用与认知负荷增加和任务完成时间延长有关,特别是对于经验不足的外科医生。组织病理学分析显示,与现有仪器相比,PAF导轨没有额外的组织损伤。这项研究表明,PAF导轨具有作为一种安全有效的器官操作和收缩方法的潜力,与现有仪器相比,它具有良好的可用性,并且没有额外的组织损伤。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Comparison Between the PAF Rail and Surgical Instruments for Organ Manipulation in Laparoscopic and Robotic Surgery: A Randomised Cross-Over Usability Study.

This study evaluates the usability and workload associated with the pneumatically attachable flexible (PAF) rail, a soft robotic device designed for safer organ manipulation and retraction during robotic-assisted partial nephrectomy (RAPN) and other laparoscopic procedures. Fourteen expert robotic and laparoscopic surgeons performed a simulated surgical retraction task using the PAF rail and standard surgical instruments. Usability was assessed using the System Usability Scale (SUS), and workload was measured with the NASA-TLX. Qualitative feedback was also collected to explore surgeon perceptions, and analysed thematically. Histopathological analysis was conducted to assess tissue integrity following instrument interaction. The PAF rail achieved SUS scores exceeding the good usability threshold, particularly among urology surgeons. However, its use was associated with increased cognitive load and longer task completion times, especially for less experienced surgeons. Histopathological analysis showed no additional tissue damage from the PAF rail compared to existing instruments. This study demonstrates that the PAF rail has potential as a safe and effective method for organ manipulation and retraction, achieving good usability and showing no additional tissue damage compared to existing instruments.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
IEEE Transactions on Biomedical Engineering
IEEE Transactions on Biomedical Engineering 工程技术-工程:生物医学
CiteScore
9.40
自引率
4.30%
发文量
880
审稿时长
2.5 months
期刊介绍: IEEE Transactions on Biomedical Engineering contains basic and applied papers dealing with biomedical engineering. Papers range from engineering development in methods and techniques with biomedical applications to experimental and clinical investigations with engineering contributions.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信