{"title":"使用 TLD100H 剂量计评估数字断层合成系统诊断成像中的器官剂量。","authors":"Giuseppe Stella, Grazia Asero, Mariajessica Nicotra, Giuliana Candiano, Rosaria Galvagno, Anna Maria Gueli","doi":"10.3390/tomography11030032","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Digital tomosynthesis (DTS) is an advanced imaging modality that enhances diagnostic accuracy by offering three-dimensional visualization from two-dimensional projections, which is particularly beneficial in breast and lung imaging. However, this increased imaging capability raises concerns about patient exposure to ionizing radiation.</p><p><strong>Methods: </strong>This study explores the energy and angular dependence of thermoluminescent dosimeters (TLDs), specifically TLD100H, to improve the accuracy of organ dose assessment during DTS. Using a comprehensive experimental approach, organ doses were measured in both DTS and traditional RX modes.</p><p><strong>Results: </strong>The results showed lung doses of approximately 3.21 mGy for the left lung and 3.32 mGy for the right lung during DTS, aligning with the existing literature. In contrast, the RX mode yielded significantly lower lung doses of 0.33 mGy. The heart dose during DTS was measured at 2.81 mGy, corroborating findings from similar studies.</p><p><strong>Conclusions: </strong>These results reinforce the reliability of TLD100H dosimetry in assessing radiation exposure and highlight the need for optimizing imaging protocols to minimize doses. Overall, this study contributes to the ongoing dialogue on enhancing patient safety in diagnostic imaging and advocates for collaboration among medical physicists, radiologists, and technologists to establish best practices.</p>","PeriodicalId":51330,"journal":{"name":"Tomography","volume":"11 3","pages":""},"PeriodicalIF":2.2000,"publicationDate":"2025-03-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11945968/pdf/","citationCount":"0","resultStr":"{\"title\":\"Assessing the Organ Dose in Diagnostic Imaging with Digital Tomosynthesis System Using TLD100H Dosimeters.\",\"authors\":\"Giuseppe Stella, Grazia Asero, Mariajessica Nicotra, Giuliana Candiano, Rosaria Galvagno, Anna Maria Gueli\",\"doi\":\"10.3390/tomography11030032\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Digital tomosynthesis (DTS) is an advanced imaging modality that enhances diagnostic accuracy by offering three-dimensional visualization from two-dimensional projections, which is particularly beneficial in breast and lung imaging. However, this increased imaging capability raises concerns about patient exposure to ionizing radiation.</p><p><strong>Methods: </strong>This study explores the energy and angular dependence of thermoluminescent dosimeters (TLDs), specifically TLD100H, to improve the accuracy of organ dose assessment during DTS. Using a comprehensive experimental approach, organ doses were measured in both DTS and traditional RX modes.</p><p><strong>Results: </strong>The results showed lung doses of approximately 3.21 mGy for the left lung and 3.32 mGy for the right lung during DTS, aligning with the existing literature. In contrast, the RX mode yielded significantly lower lung doses of 0.33 mGy. The heart dose during DTS was measured at 2.81 mGy, corroborating findings from similar studies.</p><p><strong>Conclusions: </strong>These results reinforce the reliability of TLD100H dosimetry in assessing radiation exposure and highlight the need for optimizing imaging protocols to minimize doses. Overall, this study contributes to the ongoing dialogue on enhancing patient safety in diagnostic imaging and advocates for collaboration among medical physicists, radiologists, and technologists to establish best practices.</p>\",\"PeriodicalId\":51330,\"journal\":{\"name\":\"Tomography\",\"volume\":\"11 3\",\"pages\":\"\"},\"PeriodicalIF\":2.2000,\"publicationDate\":\"2025-03-11\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11945968/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Tomography\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.3390/tomography11030032\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Tomography","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3390/tomography11030032","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
Assessing the Organ Dose in Diagnostic Imaging with Digital Tomosynthesis System Using TLD100H Dosimeters.
Background: Digital tomosynthesis (DTS) is an advanced imaging modality that enhances diagnostic accuracy by offering three-dimensional visualization from two-dimensional projections, which is particularly beneficial in breast and lung imaging. However, this increased imaging capability raises concerns about patient exposure to ionizing radiation.
Methods: This study explores the energy and angular dependence of thermoluminescent dosimeters (TLDs), specifically TLD100H, to improve the accuracy of organ dose assessment during DTS. Using a comprehensive experimental approach, organ doses were measured in both DTS and traditional RX modes.
Results: The results showed lung doses of approximately 3.21 mGy for the left lung and 3.32 mGy for the right lung during DTS, aligning with the existing literature. In contrast, the RX mode yielded significantly lower lung doses of 0.33 mGy. The heart dose during DTS was measured at 2.81 mGy, corroborating findings from similar studies.
Conclusions: These results reinforce the reliability of TLD100H dosimetry in assessing radiation exposure and highlight the need for optimizing imaging protocols to minimize doses. Overall, this study contributes to the ongoing dialogue on enhancing patient safety in diagnostic imaging and advocates for collaboration among medical physicists, radiologists, and technologists to establish best practices.
TomographyMedicine-Radiology, Nuclear Medicine and Imaging
CiteScore
2.70
自引率
10.50%
发文量
222
期刊介绍:
TomographyTM publishes basic (technical and pre-clinical) and clinical scientific articles which involve the advancement of imaging technologies. Tomography encompasses studies that use single or multiple imaging modalities including for example CT, US, PET, SPECT, MR and hyperpolarization technologies, as well as optical modalities (i.e. bioluminescence, photoacoustic, endomicroscopy, fiber optic imaging and optical computed tomography) in basic sciences, engineering, preclinical and clinical medicine.
Tomography also welcomes studies involving exploration and refinement of contrast mechanisms and image-derived metrics within and across modalities toward the development of novel imaging probes for image-based feedback and intervention. The use of imaging in biology and medicine provides unparalleled opportunities to noninvasively interrogate tissues to obtain real-time dynamic and quantitative information required for diagnosis and response to interventions and to follow evolving pathological conditions. As multi-modal studies and the complexities of imaging technologies themselves are ever increasing to provide advanced information to scientists and clinicians.
Tomography provides a unique publication venue allowing investigators the opportunity to more precisely communicate integrated findings related to the diverse and heterogeneous features associated with underlying anatomical, physiological, functional, metabolic and molecular genetic activities of normal and diseased tissue. Thus Tomography publishes peer-reviewed articles which involve the broad use of imaging of any tissue and disease type including both preclinical and clinical investigations. In addition, hardware/software along with chemical and molecular probe advances are welcome as they are deemed to significantly contribute towards the long-term goal of improving the overall impact of imaging on scientific and clinical discovery.