{"title":"一种用于磁共振成像图像质量评估的新型移动虚体插入","authors":"Frédérique P.D. van Gameren , Pim T.S. Borman , Cornelis A.T. van den Berg , Mike Cole , Grant R. Koenig , Martin F. Fast , Astrid L.H.M.W. van Lier","doi":"10.1016/j.phro.2025.100742","DOIUrl":null,"url":null,"abstract":"<div><div>Dedicated motion compensated Magnetic Resonance Imaging (MRI) for radiotherapy treatment planning promises to mitigate motion effects on imaging. We demonstrate a novel insert for an MRI safe motion phantom, which enables quality assurance of these image strategies. The capability to analyse apparent slice thickness, positional accuracy and motion blur is demonstrated for scenarios with and without motion. A respiratory-compensated scan with a 4 mm trigger-window and 16 mm peak-to-peak (p2p) motion showed a +5.0% deviation from the nominal 2 mm slice thickness. In contrast, a non-compensated scan with 4 mm p2p motion showed a +77.5% deviation, illustrating the effectiveness of motion compensation.</div></div>","PeriodicalId":36850,"journal":{"name":"Physics and Imaging in Radiation Oncology","volume":"34 ","pages":"Article 100742"},"PeriodicalIF":3.4000,"publicationDate":"2025-03-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A novel moving phantom insert for image quality assessment in magnetic resonance imaging\",\"authors\":\"Frédérique P.D. van Gameren , Pim T.S. Borman , Cornelis A.T. van den Berg , Mike Cole , Grant R. Koenig , Martin F. Fast , Astrid L.H.M.W. van Lier\",\"doi\":\"10.1016/j.phro.2025.100742\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><div>Dedicated motion compensated Magnetic Resonance Imaging (MRI) for radiotherapy treatment planning promises to mitigate motion effects on imaging. We demonstrate a novel insert for an MRI safe motion phantom, which enables quality assurance of these image strategies. The capability to analyse apparent slice thickness, positional accuracy and motion blur is demonstrated for scenarios with and without motion. A respiratory-compensated scan with a 4 mm trigger-window and 16 mm peak-to-peak (p2p) motion showed a +5.0% deviation from the nominal 2 mm slice thickness. In contrast, a non-compensated scan with 4 mm p2p motion showed a +77.5% deviation, illustrating the effectiveness of motion compensation.</div></div>\",\"PeriodicalId\":36850,\"journal\":{\"name\":\"Physics and Imaging in Radiation Oncology\",\"volume\":\"34 \",\"pages\":\"Article 100742\"},\"PeriodicalIF\":3.4000,\"publicationDate\":\"2025-03-07\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Physics and Imaging in Radiation Oncology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2405631625000478\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"ONCOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physics and Imaging in Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2405631625000478","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ONCOLOGY","Score":null,"Total":0}
A novel moving phantom insert for image quality assessment in magnetic resonance imaging
Dedicated motion compensated Magnetic Resonance Imaging (MRI) for radiotherapy treatment planning promises to mitigate motion effects on imaging. We demonstrate a novel insert for an MRI safe motion phantom, which enables quality assurance of these image strategies. The capability to analyse apparent slice thickness, positional accuracy and motion blur is demonstrated for scenarios with and without motion. A respiratory-compensated scan with a 4 mm trigger-window and 16 mm peak-to-peak (p2p) motion showed a +5.0% deviation from the nominal 2 mm slice thickness. In contrast, a non-compensated scan with 4 mm p2p motion showed a +77.5% deviation, illustrating the effectiveness of motion compensation.