{"title":"一种用于磁共振成像图像质量评估的新型移动虚体插入","authors":"Frédérique P.D. van Gameren ,&nbsp;Pim T.S. Borman ,&nbsp;Cornelis A.T. van den Berg ,&nbsp;Mike Cole ,&nbsp;Grant R. Koenig ,&nbsp;Martin F. Fast ,&nbsp;Astrid L.H.M.W. van Lier","doi":"10.1016/j.phro.2025.100742","DOIUrl":null,"url":null,"abstract":"<div><div>Dedicated motion compensated Magnetic Resonance Imaging (MRI) for radiotherapy treatment planning promises to mitigate motion effects on imaging. We demonstrate a novel insert for an MRI safe motion phantom, which enables quality assurance of these image strategies. The capability to analyse apparent slice thickness, positional accuracy and motion blur is demonstrated for scenarios with and without motion. A respiratory-compensated scan with a 4 mm trigger-window and 16 mm peak-to-peak (p2p) motion showed a +5.0% deviation from the nominal 2 mm slice thickness. In contrast, a non-compensated scan with 4 mm p2p motion showed a +77.5% deviation, illustrating the effectiveness of motion compensation.</div></div>","PeriodicalId":36850,"journal":{"name":"Physics and Imaging in Radiation Oncology","volume":"34 ","pages":"Article 100742"},"PeriodicalIF":3.4000,"publicationDate":"2025-03-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A novel moving phantom insert for image quality assessment in magnetic resonance imaging\",\"authors\":\"Frédérique P.D. van Gameren ,&nbsp;Pim T.S. Borman ,&nbsp;Cornelis A.T. van den Berg ,&nbsp;Mike Cole ,&nbsp;Grant R. Koenig ,&nbsp;Martin F. Fast ,&nbsp;Astrid L.H.M.W. van Lier\",\"doi\":\"10.1016/j.phro.2025.100742\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><div>Dedicated motion compensated Magnetic Resonance Imaging (MRI) for radiotherapy treatment planning promises to mitigate motion effects on imaging. We demonstrate a novel insert for an MRI safe motion phantom, which enables quality assurance of these image strategies. The capability to analyse apparent slice thickness, positional accuracy and motion blur is demonstrated for scenarios with and without motion. A respiratory-compensated scan with a 4 mm trigger-window and 16 mm peak-to-peak (p2p) motion showed a +5.0% deviation from the nominal 2 mm slice thickness. In contrast, a non-compensated scan with 4 mm p2p motion showed a +77.5% deviation, illustrating the effectiveness of motion compensation.</div></div>\",\"PeriodicalId\":36850,\"journal\":{\"name\":\"Physics and Imaging in Radiation Oncology\",\"volume\":\"34 \",\"pages\":\"Article 100742\"},\"PeriodicalIF\":3.4000,\"publicationDate\":\"2025-03-07\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Physics and Imaging in Radiation Oncology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2405631625000478\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"ONCOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physics and Imaging in Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2405631625000478","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

专用的运动补偿磁共振成像(MRI)用于放射治疗计划有望减轻运动对成像的影响。我们展示了一种用于MRI安全运动幻影的新型插入物,它可以保证这些图像策略的质量。在有和没有运动的情况下,分析表观切片厚度、位置精度和运动模糊的能力得到了证明。采用4毫米触发窗和16毫米峰对峰(p2p)运动的呼吸补偿扫描显示,与标称的2mm切片厚度有+5.0%的偏差。相比之下,4毫米p2p运动的非补偿扫描显示+77.5%的偏差,说明运动补偿的有效性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
A novel moving phantom insert for image quality assessment in magnetic resonance imaging
Dedicated motion compensated Magnetic Resonance Imaging (MRI) for radiotherapy treatment planning promises to mitigate motion effects on imaging. We demonstrate a novel insert for an MRI safe motion phantom, which enables quality assurance of these image strategies. The capability to analyse apparent slice thickness, positional accuracy and motion blur is demonstrated for scenarios with and without motion. A respiratory-compensated scan with a 4 mm trigger-window and 16 mm peak-to-peak (p2p) motion showed a +5.0% deviation from the nominal 2 mm slice thickness. In contrast, a non-compensated scan with 4 mm p2p motion showed a +77.5% deviation, illustrating the effectiveness of motion compensation.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Physics and Imaging in Radiation Oncology
Physics and Imaging in Radiation Oncology Physics and Astronomy-Radiation
CiteScore
5.30
自引率
18.90%
发文量
93
审稿时长
6 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信