能量仪表板:术后对电外科设备使用的洞察。

IF 2.4 2区 医学 Q2 SURGERY
Simon C Baltus, Vincent J Ribbens, Arjen Wiersma, Renske M Hoeben, Can Ozan Tan, Ivo A M J Broeders
{"title":"能量仪表板:术后对电外科设备使用的洞察。","authors":"Simon C Baltus, Vincent J Ribbens, Arjen Wiersma, Renske M Hoeben, Can Ozan Tan, Ivo A M J Broeders","doi":"10.1007/s00464-025-11642-3","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>This study presents a post-operative energy dashboard to teach surgeons about electrosurgical device use. By analyzing the energy generator, we aim to add new information to the current assessment of surgical skills. This study evaluated how such a dashboard can provide insight into differences in electrosurgery application.</p><p><strong>Methods: </strong>A semi-automated methodology for the energy dashboard was developed by acquiring intra-operative energy generator and video data, and computing metrics to compare device use. The energy dashboard quantified the use of the electrosurgical device based on the number of activations (N), the duration of individual activations (s), the total use time (s), and the total applied energy (kJ). The methodology and differences in device use were assessed based on forty-eight fundoplication surgeries performed by three surgeons.</p><p><strong>Results: </strong>The methodology identified the device activations with an F1-score of 0.95. The comparison between the surgeons showed significant differences in total usage, turn-on count, and amount of applied energy. In addition, the dashboard showed a significant difference in total applied energy (kJ) over the dissections of the gastrohepatic and gastrosplenic ligament.</p><p><strong>Conclusion: </strong>The study showed that energy monitoring can provide insights into application differences. In addition, the pilot study showed that the use of electrosurgical devices can differ significantly between surgeons. A broader application of the energy dashboard can enable a new source of information for surgical skill assessment.</p>","PeriodicalId":22174,"journal":{"name":"Surgical Endoscopy And Other Interventional Techniques","volume":" ","pages":"2641-2647"},"PeriodicalIF":2.4000,"publicationDate":"2025-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11933152/pdf/","citationCount":"0","resultStr":"{\"title\":\"Energy dashboard: post-operative insights into electrosurgical device use.\",\"authors\":\"Simon C Baltus, Vincent J Ribbens, Arjen Wiersma, Renske M Hoeben, Can Ozan Tan, Ivo A M J Broeders\",\"doi\":\"10.1007/s00464-025-11642-3\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>This study presents a post-operative energy dashboard to teach surgeons about electrosurgical device use. By analyzing the energy generator, we aim to add new information to the current assessment of surgical skills. This study evaluated how such a dashboard can provide insight into differences in electrosurgery application.</p><p><strong>Methods: </strong>A semi-automated methodology for the energy dashboard was developed by acquiring intra-operative energy generator and video data, and computing metrics to compare device use. The energy dashboard quantified the use of the electrosurgical device based on the number of activations (N), the duration of individual activations (s), the total use time (s), and the total applied energy (kJ). The methodology and differences in device use were assessed based on forty-eight fundoplication surgeries performed by three surgeons.</p><p><strong>Results: </strong>The methodology identified the device activations with an F1-score of 0.95. The comparison between the surgeons showed significant differences in total usage, turn-on count, and amount of applied energy. In addition, the dashboard showed a significant difference in total applied energy (kJ) over the dissections of the gastrohepatic and gastrosplenic ligament.</p><p><strong>Conclusion: </strong>The study showed that energy monitoring can provide insights into application differences. In addition, the pilot study showed that the use of electrosurgical devices can differ significantly between surgeons. A broader application of the energy dashboard can enable a new source of information for surgical skill assessment.</p>\",\"PeriodicalId\":22174,\"journal\":{\"name\":\"Surgical Endoscopy And Other Interventional Techniques\",\"volume\":\" \",\"pages\":\"2641-2647\"},\"PeriodicalIF\":2.4000,\"publicationDate\":\"2025-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11933152/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Surgical Endoscopy And Other Interventional Techniques\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1007/s00464-025-11642-3\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2025/3/7 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"SURGERY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Surgical Endoscopy And Other Interventional Techniques","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00464-025-11642-3","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/3/7 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"SURGERY","Score":null,"Total":0}
引用次数: 0

摘要

背景:本研究提出了一个术后能量仪表板,用于指导外科医生如何使用电外科器械。通过分析能量发生器,我们的目标是为目前的手术技能评估增加新的信息。本研究评估了这样的仪表盘如何提供对电外科应用差异的见解。方法:通过获取术中能量发生器和视频数据,并计算指标来比较设备使用情况,开发了一种半自动化的能量仪表板方法。能量仪表板根据激活次数(N)、单个激活持续时间(s)、总使用时间(s)和总应用能量(kJ)对电手术装置的使用进行量化。我们根据三位外科医生进行的48例复底手术评估了方法和设备使用的差异。结果:该方法识别设备激活的f1得分为0.95。外科医生之间的比较显示,在总使用量、开启次数和应用能量方面存在显著差异。此外,仪表板显示在胃肝韧带和胃脾韧带夹层上的总应用能量(kJ)有显著差异。结论:该研究表明,能源监测可以提供应用差异的见解。此外,初步研究表明,外科医生对电外科设备的使用有很大不同。能量仪表板的广泛应用可以为外科技能评估提供新的信息来源。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Energy dashboard: post-operative insights into electrosurgical device use.

Background: This study presents a post-operative energy dashboard to teach surgeons about electrosurgical device use. By analyzing the energy generator, we aim to add new information to the current assessment of surgical skills. This study evaluated how such a dashboard can provide insight into differences in electrosurgery application.

Methods: A semi-automated methodology for the energy dashboard was developed by acquiring intra-operative energy generator and video data, and computing metrics to compare device use. The energy dashboard quantified the use of the electrosurgical device based on the number of activations (N), the duration of individual activations (s), the total use time (s), and the total applied energy (kJ). The methodology and differences in device use were assessed based on forty-eight fundoplication surgeries performed by three surgeons.

Results: The methodology identified the device activations with an F1-score of 0.95. The comparison between the surgeons showed significant differences in total usage, turn-on count, and amount of applied energy. In addition, the dashboard showed a significant difference in total applied energy (kJ) over the dissections of the gastrohepatic and gastrosplenic ligament.

Conclusion: The study showed that energy monitoring can provide insights into application differences. In addition, the pilot study showed that the use of electrosurgical devices can differ significantly between surgeons. A broader application of the energy dashboard can enable a new source of information for surgical skill assessment.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
6.10
自引率
12.90%
发文量
890
审稿时长
6 months
期刊介绍: Uniquely positioned at the interface between various medical and surgical disciplines, Surgical Endoscopy serves as a focal point for the international surgical community to exchange information on practice, theory, and research. Topics covered in the journal include: -Surgical aspects of: Interventional endoscopy, Ultrasound, Other techniques in the fields of gastroenterology, obstetrics, gynecology, and urology, -Gastroenterologic surgery -Thoracic surgery -Traumatic surgery -Orthopedic surgery -Pediatric surgery
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信